105 CMR 120.575
Dosimetry Equipment
(A) Except for low dose-rate remote afterloader sources where the source output or activity is
determined by the manufacturer, a licensee shall have a calibrated dosimetry system available
for use. To satisfy this requirement, one of the following two conditions must be met.
(1) The system must have been calibrated using a system or source traceable to the National
Institute of Standards and Technology (NIST) and published protocols accepted by nationally
recognized bodies; or by a calibration laboratory accredited by the American Association of
Physicists in Medicine (AAPM). The calibration must have been performed within the
previous two years and after any servicing that may have affected system calibration; or
(2) The system must have been calibrated within the previous four years; 18 to 30 months
after that calibration, the system must have been intercompared with another dosimetry
system that was calibrated within the past 24 months by NIST or by a calibration laboratory
accredited by the AAPM. The results of the intercomparison must have indicated that the
calibration factor of the licensee's system had not changed by more than 2%. The licensee
may not use the intercomparison result to change the calibration factor. When
intercomparing dosimetry systems to be used for calibrating sealed sources for therapeutic
units, the licensee shall use a comparable unit with beam attenuators or collimators, as
applicable, and sources of the same radionuclide as the source used at the licensee's facility.
(B) The licensee shall have available for use a dosimetry system for spot-check output
measurements, if applicable. To satisfy this requirement, the system may be compared with a
system that has been calibrated in accordance with 105 CMR 120.575(A). This comparison must
have been performed within the previous year and after each servicing that may have affected
system calibration. The spot-check system may be the same system used to meet the requirement
in 105 CMR 120.575(A).
(C) The licensee shall retain a record of each calibration, intercomparison, and comparison in
accordance with 105 CMR 120.592(E).