105 CMR 120.579
Periodic Spot-checks for Teletherapy Units
(A) A licensee authorized to use teletherapy units for medical use shall perform output
spot-checks on each teletherapy unit once in each calendar month that include determination of:
(1) Timer accuracy, and timer linearity over the range of use;
(2) "On-off" error;
(3) The coincidence of the radiation field and the field indicated by the light beam localizing
device;
(4) The accuracy of all distance measuring and localization devices used for medical use;
(5) The output for one typical set of operating conditions measured with the dosimetry
system described in 105 CMR 120.575(B); and
(6) The difference between the measurement made in 105 CMR 120.579(A)(5) and the
anticipated output, expressed as a percentage of the anticipated output (i.e., the value
obtained at last full calibration corrected mathematically for physical decay).
(B) A licensee shall perform measurements required by 105 CMR 120.579(A) in accordance
with procedures established by the authorized medical physicist. That individual need not
actually perform the output spot-check measurements.
(C) A licensee shall have the authorized medical physicist review the results of each output
spot-check within 15 days. The authorized medical physicist shall promptly notify the licensee
in writing of the results of each output spot-check.
(D) A licensee authorized to use a teletherapy unit for medical use shall perform safety
spot-checks of each teletherapy facility once in each calendar month and after each source
installation to assure proper operation of:
(1) Electrical interlocks at each teletherapy room entrance;
(2) Electrical or mechanical stops installed for the purpose of limiting use of the primary
beam of radiation restriction of source housing angulation or elevation, carriage or stand
travel, and operation of the beam "on-off" mechanism;
(3) Source exposure indicator lights on the teletherapy unit, on the control console, and in
the facility;
(4) Viewing and intercom systems;
(5) Treatment room doors from inside and outside the treatment room; and
(6) Electrically assisted treatment room doors with the teletherapy unit electrical power
turned "off".
(E) If the results of the checks required in 105 CMR 120.579(D) indicate the malfunction of any
system, a licensee shall lock the control console in the off position and not use the unit except
as may be necessary to repair, replace, or check the malfunctioning system.
(F) A licensee shall retain a record of each spot-check required by 105 CMR 120.579(A) and
(D), in accordance with 105 CMR 120.592(G).