105 CMR 120.603
Equipment Requirements
(A) Safety Device. A device which prevents the entry of any portion of an individual's body
into the primary x-ray beam path or which causes the beam to be shut off upon entry into its path
shall be provided on all open-beam configurations. A registrant [or licensee] may apply to the
Agency for an exemption from the requirement of a safety device. Such application shall
include:
(1) A description of the various safety devices that have been evaluated;
(2) The reason each of these devices cannot be used; and,
(3) A description of the alternative methods that will be employed to minimize the
possibility of an accidental exposure, including procedures to assure that operators and others
in the area will be informed of the absence of safety devices.
(B) Warning Devices.
(1) Open-beam configurations shall be provided with a readily discernible indication of:
(a) X-ray tube "on-off" status located near the radiation source housing, if the primary
beam is controlled in this manner; and/or,
(b) Shutter "open-closed" status located near each port on the radiation source housing,
if the primary beam is controlled in this manner.
(2) Warning devices shall be labeled so that their purpose is easily identified. On
equipment installed after the effective date of 105 CMR 120.600, warning devices shall have
fail-safe characteristics.
(C) Ports. Unused ports on radiation source housings shall be secured in the closed position
in a manner which will prevent casual opening.
(D) Labeling. All analytical x-ray equipment shall be labeled with a readily discernible sign
or signs bearing the radiation symbol and the words:
(1) "CAUTION - HIGH INTENSITY X-RAY BEAM", or words having a similar intent,
on the x-ray source housing; and,
(2) "CAUTION RADIATION - THIS EQUIPMENT PRODUCES RADIATION WHEN
ENERGIZED", or words having a similar intent, near any switch that energizes an x-ray tube
if the radiation source is an x-ray tube; or,
(3) "CAUTION - RADIOACTIVE MATERIAL", or words having a similar intent, on the
source housing in accordance with 105 CMR 120.237 and 120.238 if the radiation source is
a radionuclide.
(E) Shutters. On open-beam configurations installed after the effective date of 105 CMR
120.600, each port on the radiation source housing shall be equipped with a shutter that cannot
be opened unless a collimator or a coupling has been connected to the port.
(F) Warning Lights.
(1) An easily visible warning light labeled with the words "X-RAY ON", or words having
a similar intent, shall be located:
(a) Near any switch that energizes an x-ray tube and shall be illuminated only when the
tube is energized; or,
(b) In the case of a radioactive source, near any switch that opens a housing shutter and
shall be illuminated only when the shutter is open.
(2) On equipment installed after the effective date of 105 CMR 120.600, warning lights
shall have fail-safe characteristics.
(G) Radiation Source Housing. Each radiation source housing shall be subject to the following
requirements:
(1) Each x-ray tube housing shall be equipped with an interlock that shuts off the tube if it
is removed from the radiation source housing or if the housing is disassembled.
(2) Each radioactive source housing or port cover or each x-ray tube housing shall be so
constructed that, with all shutters closed, the radiation measured at a distance of five
centimeters from its surface is not capable of producing a dose in excess of 2.5 millirems
(0.025 mSv) in one hour. For systems utilizing x-ray tubes, this limit shall be met at any
specified tube rating.
(H) Generator Cabinet. Each x-ray generator shall be supplied with a protective cabinet which
limits leakage radiation measured at a distance of five centimeters from its surface such that it
is not capable of producing a dose in excess of 0.25 millirem (2.5 µSv) in one hour.