105 CMR 120.605
Operating Requirements
(A) Procedures. Normal operating procedures shall be written and available to all analytical
x-ray equipment workers. No individual shall be permitted to operate analytical x-ray equipment
in any manner other than that specified in the procedures unless such individual has obtained
written approval of the radiation safety officer.
(B) Bypassing. No individual shall bypass a safety device or interlock unless such individual
has obtained the approval of the radiation safety officer. Such approval shall be for a specified
period of time. When a safety device or interlock has been bypassed, a readily discernible sign
bearing the words "SAFETY DEVICE NOT WORKING", or words having a similar intent, shall
be placed on the radiation source housing.
(C) Repair or Modification of X-ray Tube Systems. Except as specified in 105 CMR
120.605(B), no operation involving removal of covers, shielding materials or tube housings or
modifications to shutters, collimators, or beam stops shall be performed without ascertaining that
the tube is off and will remain off until safe conditions have been restored. The main switch,
rather than interlocks, shall be used for routine shutdown in preparation for repairs.
(D) Radioactive Source Replacement, Testing, or Repair. Radioactive source housings shall
be opened for source replacement, leak testing, or other maintenance or repair procedures only
by individuals authorized to specifically conduct such procedures under a license issued by the
U.S. Nuclear Regulatory Commission, an Agreement State, or a Licensing State.