105 CMR 127.019
Equipment Quality Assurance
(A) Quality Assurance Records for Mammography X-ray Equipment.
(1) Records of all quality assurance tests shall be maintained for mammographic x-ray
equipment and shall include at least the following:
(a) The mammographic x-ray equipment performance evaluation and optimization,
including initial acceptance testing (to be completed before patients are exposed with the
machine) and subsequent evaluations and testing;
(b) Verification that mammographic x-ray equipment is in safe operating order;
(c) Subsequent quality assurance test results;
(d) Notes regarding changes in operating conditions shall be recorded on the appropriate
control charts;
(e) Operating levels and control limits for all quality assurance tests shall be indicated
on the appropriate control chart;
(f) All records shall bear the date the procedure was done and the initials or name of the
individual performing the test;
(g) Quality assurance tests on the Radiologist workstation monitors shall be done in
accordance with manufacturer's requirements;
(h) Quality assurance tests on laser hardcopy printers shall be done in accordance with
manufacturer's requirements.
Records or written logs of maintenance and/or repairs of mammographic x-ray
equipment shall be kept for at least three years. All records for a machine shall be kept in
a single file at the machine use location. Repair records shall be referenced to quality
assurance test results, if applicable.
(3) All quality assurance records shall be maintained for a minimum of 24 months and
readily available at the use location for review by representatives of the Department.
(B) Quality Assurance Records and Requirements for Test Equipment and Mammography
Systems.
(1) The quality assurance test equipment shall be in good operating order.
(2) Records of each repair and/or calibration of quality assurance test equipment shall
include the date of repair and/or calibration.
(3) Records of each repair and/or calibration of quality assurance test equipment shall be
kept for at least three years.
(4) Quality assurance tests shall be performed in accordance with manufacturers’
recommendations after the repair and/or replacement of any component of the
mammographic x-ray system, prior to using the equipment.
(C) Quality Assurance Tests for Equipment.
Laser Hardcopy Printer. Quality assurance tests for laser printers, if used by the
mammography facility for clinical diagnosis, shall be performed in accordance with the
manufacturer's recommendations.
(2) Mammography Phantom.
(a) Quality assurance tests for image quality in fixed mammography facilities shall be
performed weekly.
(b) Quality assurance tests for image quality in mobile mammography facilities shall be
performed each time the machine has been moved, before beginning examinations of
patient at least weekly.
(c) Each facility shall utilize a mammography phantom approved or accepted by ACR
for performing quality assurance tests for image quality. The following image quality
requirements shall be met, for different mammography modalities:
1. For Digital Mammography, manufacturer's specifications or specifications stated
in the 2016 ACR Digital Mammography Quality Control Manual, or its revisions.
2. For Digital Breast Tomosynthesis, manufacturer's specifications.
3. For stereotactic breast biopsy, manufacturer's specifications or specifications
stated in the 1999 ACR Sterotactic Breast Biopsy Quality Control Manual, or its
revisions.
(d) Each facility shall take corrective action in the event that quality assurance tests for
image quality do not meet the requirements in 105 CMR 127.019(C)(2)(c).
(3) Retake Analysis.
(a) Quality assurance tests for retake analysis shall be performed as specified by the
mammography system manufacturer or at least quarterly.
(b) The test tools for performing quality assurance tests on retake analysis are records
of rejected images.
(c) When performing quality assurance tests for retake analysis, the total number of
images and the total number of rejected images used during the test period shall be
evaluated for ascertaining the reason for rejection. The rejected images shall be
identified in at least the following categories:
1. Technique;
2. Motion;
3. Positioning error;
4. Digital imaging processing; and
5. Clinically significant artifacts (dead pixel clusters on digital detector arrays, shifts
of masked spots, etc.).
(4) Compression in the Manual and/or Power Modes
(a) Quality assurance tests for compression in the manual and/or power modes in fixed
mammography facilities shall be performed semi-annually.
(b) Quality assurance tests for compression in the manual and/or power modes in
mobile mammography facilities shall be performed each time the machine has been
moved, but not less than semi-annually.
(c) Each facility shall use a push-pull force gauge or a flat bathroom type scale method
with accuracy of + 2% of full scale for performing quality assurance tests.
(d)
When performing the quality assurance test for compression, the maximum
compression force shall range from a minimum of 25 to a maximum of 45 pounds in both
the manual and power drive modes and shall remain consistent throughout the period of
exposure. The compression release mechanism shall be tested at the same interval.
(5) Automatic Exposure Control.
(a) Quality assurance tests for automatic exposure control (phototiming) at fixed
mammography facilities shall be performed at least annually.
(b) Quality assurance tests for automatic exposure control (phototiming) at mobile
mammography facilities shall be performed at least semi-annually.
(6) Kilovolt Peak (kVp) Accuracy.
(a) Quality assurance tests for kVp accuracy at fixed mammography facilities shall be
performed at least annually.
(b) Quality assurance tests for kVp accuracy at mobile mammography facilities shall be
performed at least semi-annually.
(c) Each facility shall use a non-invasive kVp meter with appropriate calibration
procedures or an oscilloscope and voltage divider network to perform quality assurance
tests.
(d) When performing quality assurance tests for kVp accuracy, the actual (measured)
versus indicated kVp shall be maintained as specified by the x-ray equipment
manufacturer, but shall not vary from the indicated kVp by greater than 5% from the
indicated.
(7) Glandular Tissue Dose.
(a) Quality assurance tests for glandular tissue dose at fixed mammography facilities
shall be performed at least annually.
(b) Quality assurance tests for glandular tissue dose at mobile mammography facilities
shall be performed at least semi-annually.
(c) Each facility shall use an ion chamber or solid state detector calibrated at the energy
ranges used in mammography for performing quality assurance tests.
(d) When performing quality assurance tests for glandular tissue dose, the technique
setting shall be those used for imaging a 4.5 cm compressed breast in the cranio-caudal
projection and the measurement point shall be 4.5 cm from the surface of the breast
support as appropriate.
(e) For mammographic examination of a 4.5 centimeter (cm) thick, compressed breast,
consisting of 50% glandular and 50% adipose tissue, the average glandular dose shall be
no more than 3 mGy per view.
(8) Spatial Resolution.
(a) Quality assurance tests for spatial resolution at fixed mammography facilities shall
be performed at least annually.
(b) Quality assurance tests for spatial resolution at mobile mammography facilities shall
be performed at least semi-annually.
(9) Half-value Layer (HVL)
(a) Quality assurance tests for half-value layer at fixed mammography facilities shall be
performed at least annually.
(b) Quality assurance tests for half-value layer at mobile mammography facilities shall
be performed at least semi-annually.
(c) When performing quality assurance tests for half-value layer, the HVL shall be
measured with the mammography x-ray unit's compression paddle in place over the
ionization chamber or solid state detector.
(10) All mammography facilities shall follow the manufacturer's recommendations or the
ACR Digital Mammography QC Manual for quality control of all components used for
digital image acquisition, digital image processing and digital image display.