105 CMR 180.395
Standard - Microbiology: Quality Control
Chemical and biological solutions, reagents, and antisera shall be tested and inspected as specified
below for reactivity and deterioration using positive and negative control organisms. The following
requirements apply to the specialties of Bacteriology, Mycology, Parasitology, and Virology:
(A) Materials which shall be tested and inspected on each day of use include:
(1) Media and discs used for antibiotic susceptibility testing;
(2) Reagents and solutions used in the performance of conventional biochemical testing;
(3) Stains (exception - Gram Stain).
(B) Materials which shall be tested and inspected each week of use and when a new vial or container
is opened include:
(1) Antibiotic discs (exception: antibiotic susceptibility discs);
(2) Chemical strips;
(3) Gram Stain materials.
(C) Materials which shall be tested and inspected each month of use and when a new vial or container
is opened include:
Antisera.
(D) Materials which shall be tested and inspected when each new batch or shipment is received
include:
(1) Tube and plate media;
(2) Gram negative identification sets.
(E)
Bacteriology and Mycology. Staining materials shall be tested for intended reactivity by
concurrent application to smears of micro-organisms with predictable staining characteristics. Each
batch of medium shall be tested before or concurrently with use with selected organisms to confirm
required growth characteristics, selectivity, enrichment, and biochemical response.
(F) Parasitology. A reference collection of slides, photographs, or gross specimens of identified
parasites shall be available and used in the laboratory for appropriate comparison with diagnostic
specimens. A calibrated ocular micrometer shall be used for determining the size of ova and parasites,
if size is a critical factor.
(G) Virology. Systems for the isolation of viruses and reagents for the identification of viruses shall
be available to cover the viruses for which services are offered. Records shall be maintained which
reflect the systems used and the reactions observed. In tests for the identification of viruses, controls
shall be employed which will identify erroneous results. If serodiagnostic tests for virus diseases are
performed, requirements for quality control as specified for serology shall apply.