R 333.5036
R 333.5036 Shielding plan review.
Cite as Mich. Admin. Code R 333.5036
Rule 36. (1) An applicant, before registration, shall submit the floor plans, shielding
specifications, and equipment arrangement of all new installations, or modifications of
existing installations, using a radiation machine, to the department for review and
approval. Application for a radiation shielding plan review shall be completed on an
application form supplied by the department according to the instructions contained in
that form. Radiation shielding plans are not required to be submitted for any of the
following:
(a) Dental intraoral or panoramic machines used in dental clinical facilities.
(b) Cabinet x-ray systems, analytical systems, process or control gauges, or cold-
cathode gas discharge tubes.
(c) Bone densitometers.
(d) Mobile or portable radiographic machines unless routinely used in 1 location.
(e) C-arm fluoroscopic systems having a maximum source-image receptor distance
of less than 45 centimeters that are used for extremity use only.
(2) The department may require the applicant to use the services of a health physicist
or medical physicist to determine the shielding requirements before the department’s plan
review and approval.
(3) The department’s approval of a plan shall not preclude the requirement of
additional modifications if a subsequent analysis shows an individual could receive a
dose exceeding the limits in R 333.5057 to R 333.5060.
(4) Shielding design goals of 0.1 millisievert (10 mrem) per week for controlled
areas and 0.02 millisievert (2 mrem) per week and 0.02 millisievert (2 mrem) in any 1
hour for uncontrolled areas shall be applied to new facilities and to new construction in
existing facilities.
(a) For the purpose of this subrule, “controlled area” means a limited access area
where the occupational exposure of personnel to radiation is under the supervision of an
individual in charge of radiation protection. In a controlled area, the access, occupancy,
and working conditions are controlled for the purpose of radiation protection.
(b) For the purpose of this subrule, “uncontrolled area” means all areas of the facility
and the surrounding environs that are not controlled for the purpose of radiation
protection.
(5) For machines installed after the effective date of this part, the registrant shall
maintain for inspection by the department a scale drawing of the room where a stationary
radiation machine system is located. The drawing or accompanying attachments shall
indicate the use of areas adjacent to the room and include an estimate of the occupancy in
each area. In addition, the drawing or attachment shall include at least 1 of the following:
(a) The type and thickness of materials, or lead equivalency, of each protective
barrier.
(b) The results of a survey for radiation levels at the operator's position and at
pertinent points outside the room under specified test conditions.
(6) The department may withhold initial registration of a radiation machine pending
receipt of either of the following:
(a) Plans and specifications for room design and shielding and approval of those
plans and specifications.
(b) Documentation from the applicant that a certificate of need (CON) has been
issued if the use of that machine is a covered clinical service as defined in the act.