R 333.5325 X-ray equipment.
Cite as Mich. Admin. Code R 333.5325
Rule 325. (1) All x-ray tube housings in fixed radiographic installations shall be of
the diagnostic type.
(2) The aluminum equivalent of the total filtration in the useful beam shall not be less
than the values shown in table 325-1.
TABLE 325-1
Operating Potential
Minimum Total Filter
(Inherent plus added)
Below 50 kVp
0.5 mm aluminum
50-70 kVp
1.5 mm aluminum
Above 70 kVp
2.5 mm aluminum
(3) If the filter in the machine is not accessible for examination and the total filtration
is not known, subrule (2) of this rule may be assumed to have been met if the half-value
layer is not less than any of the following:
(a) 0.6 mm aluminum at 49 kVp.
(b) 1.6 mm aluminum at 70 kVp.
(c) 2.6 mm aluminum at 90 kVp.
(4) Under conditions of subrule (3) of this rule for tube potentials above 90 kVp,
subrule (2) of this rule may be assumed to have been met if the half-value layer is not less
than that specified in table 325-2.
(5) The half-value layer of the useful beam for a given x-ray tube potential shall not
be less than the values shown in table 325-2.
TABLE 325-2
Design Operating Range
(kVp)
Measured
Potential
(kVp)
Half-value
Layer
(mm aluminum)
Below 50
30
0.3
40
0.4
49
0.5
50 to 70
50
1.2
60
1.3
70
1.5
Above 70
80
2.3
90
2.5
100
2.7
110
3.0
120
3.2
130
3.5
140
3.8
150
4.1
(6) To determine the half-value layer at an x-ray tube potential which is not listed in
table 325-2, linear interpolation or extrapolation may be made. Positive means shall be
provided to ensure that at least the minimum filtration needed to achieve these beam
quality requirements is in the useful beam during each exposure.
(7) Machines equipped with beryllium window x-ray tubes with removable filters
shall contain keyed filter interlock switches in the tube housing and suitable indication on
the control panel of the added filter in the useful beam. The total filtration permanently
in the useful beam shall not be less than 0.5 millimeter aluminum equivalent and shall be
clearly indicated on the tube housing.
(8) Beryllium window x-ray tubes shall not be used routinely for general purpose
diagnostic examinations. Such a tube may comprise an x-ray subsystem if needed for
special soft tissue technique in accord with subrule (7) of this rule.
(9) Beam-limiting devices, such as diaphragms, cones, or adjustable collimators,
capable of restricting the useful beam to the area radiographically recorded shall be
provided to define the beam and shall provide the same degree of attenuation as that
required of the tube housing.
(10) Beam-limiting devices shall be calibrated in terms of the size of the projected
useful beam at specified source-image receptor distances (SID). This calibration shall be
clearly and permanently recorded on the beam-limiting device. Calibration of adjustable
beam-limiting devices shall permit reproducible settings.
(11) X-ray systems designed for only 1 image receptor size at a fixed SID shall be
provided with means to limit the field at the plane of the image receptor to dimensions no
greater than those of the image receptor, and to align the center of the x-ray field with the
center of the image receptor to within 2% of the SID.
(12) General purpose radiographic x-ray systems shall be equipped with adjustable
beam-limiting devices containing light localizers that define the entire field.
(13) The size of the x-ray beam projected by fixed aperture beam-limiting devices,
except those used for stereoradiography, shall not exceed the dimensions of the image
receptor by more than 2% of the SID when the axis of the x-ray beam is perpendicular to
the plane of the image receptor.
(14) The calibrated field size indicator on adjustable beam-limiting devices shall be
accurate to within 2% of the SID. The light field shall be aligned with the x-ray field
with the same degree of accuracy. The field size projected by automatic adjustable
beam-limiting devices shall provide the same precision.
(15) For radiographic procedures resulting in multiple views on a single image
receptor, the beam-limiting device shall limit the x-ray field size to the recorded
radiographic image size within 2% of the SID. Covering a portion of the image receptor
with radio-opaque material is not a substitute for proper x-ray field limitation.
(16) Radiographic x-ray machines used for purposes other than mammography or
extremity radiography only shall be capable of operation at not less than an average
current of 100 milliamperes (mA) during all radiographic techniques used. A machine
not capable of sustained operation at not less than an average of 100 mA for the duration
of a given technique shall not be used for that technique.
(17) A device shall be provided which terminates the exposure at a preset time
interval or exposure limit. The operator shall be able to terminate the exposure at any
time by discontinuing pressure upon the exposure switch except that during serial
radiography means may be provided to permit completion of a single exposure in
progress.
(18) The exposure switch, except for those used in conjunction with spot film devices
in fluoroscopy, shall be securely fixed so that the operator is required to be behind a fixed
shield which intercepts the useful beam and any radiation which has been scattered only
once.
(19) When 2 or more x-ray tube heads are operated from a single exposure switch,
there shall be indication at the control panel showing which tube is connected and ready
to be energized and means to prevent energizing more than 1 tube head simultaneously.
Machines designed for simultaneous multiple tube operation shall have positive means
for selecting single tube or multiple tube operation.
(20) The control panel shall provide positive visual identification of the production of
x-rays when the x-ray tube is energized. A milliammeter may comply with this subrule.
(21) A signal audible to the operator shall indicate that the exposure has ended.
(22) The technique factors to be used during an exposure shall be indicated before the
exposure begins. When automatic exposure controls are used, only those technique
factors which are set before the exposure shall be indicated. On equipment having fixed
technique factors, this requirement may be met by permanent markings. Indication of
technique factors shall be visible from the operator's position.