R 333.5333
R 333.5333 Conditions of operation.
Cite as Mich. Admin. Code R 333.5333
Rule 333. (1) An operator shall properly utilize the beam-limiting devices provided to
restrict the useful beam to the smallest area consistent with clinical requirements.
Particular care shall be taken to align accurately the x-ray beam with the patient and film.
(2) The operator shall ensure the presence of adequate filtration before a radiographic
procedure.
(3) Staff personnel routinely working with or around radiation sources shall not be
required by the registrant to hold film or restrain patients during radiography. If such
procedure is permitted, personnel exposure shall not exceed the limits in R 333.5057 to
R 333.5059 or the procedure shall be prohibited.
(4) When a patient must be held in position for radiography, mechanical supporting or
restraining devices shall be available and shall be used unless contraindicated. If the
patient must be held by an individual, this individual shall wear protective gloves and a
protective apron of 0.5 millimeter minimum lead equivalence and shall be positioned so
that no part of his or her body can be struck by the useful beam and that his or her body is
as far as possible from the edge of the useful beam.
(5) Only individuals whose presence is necessary shall be permitted in the
radiographic room during an exposure. An individual, except the patient, shall be
protected by 0.5 millimeter minimum lead equivalent aprons unless protected by an
approved primary barrier.
(6) Personnel monitoring shall be performed in controlled areas for each individual
occupationally exposed to ionizing radiation from diagnostic x-ray equipment.
Individual monitoring devices such as film badge dosimeters or thermoluminescent
dosimeters shall be permanently assigned to each occupationally exposed individual.
This monitoring shall be continuous during employment as a radiation worker.
(7) Personnel exposure records shall be kept on permanent available file at the facility
where the exposure occurs.
(8) Monitoring devices used to estimate whole body exposure shall normally be worn
on the chest or abdomen. Monitoring of all other body parts shall comply with
R 333.5065.
(9) Monitoring devices worn to estimate personnel occupational exposure shall not be
worn by the individual when he or she is exposed as a patient for a medical or dental
reason.
(10) The gonads of children and individuals who have not passed the reproductive
age shall be protected from the useful beam either by the use of shielding (0.5 mm lead
equivalent), collimation, or special gonad shields when this does not interfere with the
conditions or objectives of the examination.
(11) Intensifying screens shall be employed to reduce patient exposure except in cases
where a noticeable decrease in image definition may reduce the clinical value of the
examination. Film and screen speed combinations shall be carefully selected to produce
the necessary clinical information with the least exposure to the patient consistent with
current clinical judgment.
(12) Film processing materials and techniques shall be those recommended by the x-
ray film and processing materials manufacturers unless otherwise tested to ensure
maximum information content of the developed film. Sight developing is not permitted
except under extreme emergency conditions. Correct temperature control and
development time shall be used.
(13) A radiographic x-ray system shall not be left unattended without locking the
apparatus, room, or building in some manner which prevents use of the apparatus by
unauthorized persons.