R 333.5365
R 333.5365 Enclosures.
Cite as Mich. Admin. Code R 333.5365
Rule 365. (1) The degree of protection required for an enclosure shall be determined
by the workload, use and occupancy factors, and the tube potential, tube current,
mechanical movement, and distance, and shall be subject to design approval by the
department.
(2) In a radiographic room, wall and floor areas exposed to the useful beam plus an
area extending at least 30 centimeters (1 foot) beyond shall be provided with a primary
protective barrier where necessary as determined by workload, use and occupancy
factors, and distance.
(3) Secondary protective barriers shall be provided in the radiographic room ceiling
and in those walls not requiring primary barriers. Common building materials often
fulfill this requirement.
(4) A fixed barrier of 1.6 millimeters (1/16 inch) lead equivalence, such as a shielded
wall partition or immobilized portable shield, is recommended for operator protection.
When this protection is provided, the operator shall be able to see and communicate with
the patient from a shielded position.