R 333.5384 X-ray equipment.
Cite as Mich. Admin. Code R 333.5384
Rule 384. (1) X-ray equipment in panoramic installations shall meet the requirements
of R 333.5373 excluding subrules (7) to (11).
(2) For purposes of this rule, "image receptor" means that portion of the x-ray film or
digital receptor instantaneously exposed by the x-ray beam subtended by a beam-limiting
diaphragm immediately adjacent to the front of the radiographic film or digital receptor,
if the panoramic technique requires this diaphragm.
(3) The x-ray tube housing shall be provided with a beam-limiting diaphragm which
shall limit the field at the plane of the image receptor to dimensions not exceeding the
dimensions of the image receptor and shall align the center of the x-ray field with the
center of the image receptor to within 2% of the SID.
(4) Mechanical support of the tube head and image receptor shall maintain beam
alignment without drift or vibration of sufficient magnitude to cause the need for
manually restraining the tube or retaking the x-ray.
(5) A device shall be provided which terminates the exposure at a preset time interval
or exposure limit. The operator shall be able to terminate the exposure at any time by
discontinuing pressure upon the exposure switch.