Rule 409. (1) An operator shall properly utilize the beam-limiting devices provided to
restrict the useful beam to the smallest area consistent with clinical requirements.
Particular care shall be taken to align accurately the x-ray beam with the patient and film.
(2) The operator shall ensure the presence of adequate filtration before a radiographic
procedure pursuant to R 333.5405(2).
(3) When a patient or film is held in position for radiography, mechanical supporting
or restraining devices shall be available and shall be used unless contraindicated. Proper
use of these devices shall permit the operator to stand behind the primary control shield
during most radiographic procedures.
(4) If the patient or film is held by 1 or more individuals, each individual shall wear
protective gloves and body aprons of 0.5 millimeter minimum lead equivalence. Each
individual shall be so positioned that no part of his or her body is struck by the useful
beam and that his or her body is as far as possible from the edge of the useful beam.
(5) Only individuals whose presence is necessary shall be permitted in the
radiographic room during an exposure. These individuals shall be protected as specified
in subrule (4) of this rule unless protected by an approved primary barrier.
(6) If an auxiliary foot switch is provided as specified in R 333.5405(12), it shall be
used only by a licensed veterinarian and only at times when sufficient personnel are not
available to permit use of the primary exposure switch specified in R 333.5405(11).
(7) To protect the feet of the veterinarian or his or her assistant from the primary
beam while restraining patients, the underside of the radiographic table shall be protected
by at least 1.6 millimeter (1/16 inch) lead or equivalent protection approved by the
department.
(8) Personnel monitoring shall be performed in controlled areas for each individual
occupationally exposed to ionizing radiation from diagnostic x-ray equipment.
Individual monitoring devices, such as film badge dosimeters or thermoluminescent
dosimeters, shall be permanently assigned to each occupationally exposed individual.
This monitoring shall be continuous during employment as a radiation worker.
(9) Personnel exposure records shall be kept on permanent available file at the facility
where the exposure occurs.
(10) Monitoring devices used to estimate whole body exposure shall normally be
worn on the chest or abdomen. Monitoring of all other body parts shall comply with
R 333.5065.
(11) Monitoring devices worn to estimate personnel occupational exposure shall not
be worn by the individual when he or she is exposed as a patient for a medical or dental
reason.
(12) Medical x-ray screen type films and intensifying screens shall be employed to
reduce patient exposure except in cases where a noticeable decrease in image definition
resulting from increased sensitivity may reduce the clinical value of the examination.
(13) Film processing materials and techniques shall be those recommended by the x-
ray film and processing materials manufacturers unless otherwise tested to ensure
maximum information content of the developed film. Sight developing is not permitted
except under extreme emergency conditions. Correct temperature control and
development time are necessary to minimize radiation dose to the patient.
(14) A radiographic x-ray system shall not be left unattended without locking the
apparatus, room, or building in some manner which prevents use of the apparatus by
unauthorized persons.