R 333.5452
R 333.5452 Radiation surveys.
Cite as Mich. Admin. Code R 333.5452
Rule 452. (1) A registrant shall maintain at each accelerator installation or class D
operation appropriate calibrated and operable portable radiation monitoring instruments
to make physical radiation surveys as required by this part and R 333.5051 to
R 333.5089.
(2) These instruments shall be capable by design, calibration, and operation of
measuring the intensity of the various types and energies of radiation produced by the
accelerator. These instruments shall be tested for proper operation at the beginning of
each day they are to be used and calibrated at least quarterly.
(3) During repair or calibration of a radiation monitoring instrument, a spare
calibrated and operable instrument shall be provided or accelerator operations which
require the instrument shall be terminated until required instrumentation is available.
(4) A radiation protection survey shall be performed and documented pursuant to
R 333.5063 when changes have been made in shielding, operation, equipment, or
occupancy of adjacent areas.
(5) Radiation levels in all accessible high radiation areas shall be continuously
monitored except in installations designed for human exposure. The monitoring devices
shall be independent and capable of providing a remote and local readout with visual or
audible alarms, or both, at the control panel and at the monitoring stations.
(6) All area monitors shall be calibrated at established periodic intervals approved by
the department.
(7) All area surveys shall be made as specified in the written procedures established
by a health physics consultant or the radiation protection supervisor of the accelerator
facility and approved by the department.
(8) Records of all radiation protection surveys, calibration results, instrumentation
tests, and smear results shall be kept current and on file at each accelerator facility.