R 333.5482 X-ray equipment.
Cite as Mich. Admin. Code R 333.5482
Rule 482. (1) Tube housing leakage from analytical x-ray sources shall not exceed 0.5
milliroentgen per hour at a 5 centimeter distance from the surface of the tube housing
with the beam ports blocked and the tube operating at its leakage technique factors. Also,
radiation originating from the high voltage power supplies shall not exceed this limit.
(2) For instruments in which the primary x-ray beam is completely enclosed, the
radiation shall be less than 2 milliroentgens per hour at a distance of 25 centimeters from
the cabinet surface.
(3) For enclosed equipment, interlocks shall be provided on all access panels which
terminate exposure and prevent operation while the panel is removed.
(4) Open beam analytical x-ray equipment shall meet all of the following:
(a) X-ray diffraction cameras shall have the appropriate ports arranged so that the
camera collimating system is in place before the x-ray tube can be energized or the
shutter can be opened.
(b) An adapter between the x-ray tube and the collimator of the diffractometer camera
or other accessory shall provide protection equivalent to that required by subrule (1) of
this rule.
(c) Safety interlocks shall not be used as routine cut-off switches during normal
operation. They shall be operated as safety devices only, and tested periodically. When
the interlock system terminates the x-ray beam, it shall be necessary to reset the "on"
switch at the control panel to resume operation.
(d) Tube head ports which are not in use shall be secured in a closed position and
interlocked to the x-ray generator or warning system.
(e) The shutter indicator shall be conspicuously displayed to disclose the "open" or
"closed" position of the shutter.
(f) The instrument shall display a conspicuous warning label such as "CAUTION
RADIATION
-
THIS
EQUIPMENT
PRODUCES
X-RADIATION
WHEN
ENERGIZED."
(g) A red warning light shall indicate "X-RAY ON" when the equipment is producing
x-rays. Other signal lights or alarms shall operate only to indicate a malfunction which
may produce a radiation, electrical, or other hazard.