R 333.5505 Equipment.
Cite as Mich. Admin. Code R 333.5505
Rule 505. (1) During all phases of operation of an electron microscope at the
maximum rated continuous tube current for the maximum rated peak tube potential the
radiation exposure rate as measured in air at a distance of 5 centimeters from an
accessible point on the external surface of the microscope shall not exceed 0.5
milliroentgen per hour.
(2) Interlocks shall be provided on all potential radiation hazard access panels which
terminate exposure and prevent operation while the panel is removed.
(3) The instrument shall display a conspicuous warning label such as “CAUTION
RADIATION
-
THIS
EQUIPMENT
PRODUCES
X-RADIATION
WHEN
ENERGIZED.”
(4) Electron microscopes that are not capable of exceeding an operating potential of
50 kilovolts are exempt from the requirements of R 333.5505 to R 333.5508.