R 333.5610
R 333.5610 Supplemental machine information; effect of failure to submit information.
Cite as Mich. Admin. Code R 333.5610
Rule 610. (1) Upon notice from the department that an application for mammography
authorization is complete and complies with these rules and at the specific request of the
department, the applicant shall, within 45 days of the department's request, provide all of
the following information for each radiation machine for which mammography
authorization is being sought:
(a) Confirmation that a department-approved mammography phantom is on-site when
mammography is performed and is used in the facility's ongoing quality control program.
(b) Processor or laser film printer quality control data and corrective actions, if any,
taken as a result of that data for a 30-day period beginning after the date the application
was sent to the department.
(c) An x-ray image of a department-approved mammography phantom which is taken
during the 30-day period for which processor quality control data is required under
subdivision (b) of this subrule. The phantom image shall be taken using routine machine
settings being used by the facility for that mammography machine for a 4.2-centimeter
compressed breast of average density. The phantom image shall be accompanied by
documentation of the date that the image was taken and the machine settings that were
used.
(d) Determinations of the half-value layer, radiation exposure at skin entrance, and
mean glandular dose. These determinations shall be made with the use of a department-
approved dosimetry device exposed on the phantom during the same exposure that is
used to produce an x-ray image to be submitted under subdivision (c) of this subrule, or
that are made by other methods as specified or approved by the department.
(e) A set of clinical images produced on or after the date that the application was sent
to the department. Mammography images shall be without pathology for each of 2
representative patients, 1 with dense breasts and 1 with fatty breasts. Stereotactic breast
biopsy images shall be from 1 calcification biopsy case that demonstrates accurate needle
location and includes the case’s corresponding mammograms. The submitted images
shall meet all of the following:
(i) The cases are examples of the facility’s best work.
(ii) The images are from actual patients.
(iii) Both screen-film and digital images are labeled with the identification
information required in R 333.5657 for mammography images or R 333.5683 for
stereotactic breast biopsy images.
(iv) The lead interpreting physician reviews and approves the clinical images.
(f) A copy of the medical physicist’s most recent equipment survey report.
(2) The department may waive the requirements of subrule (1) of this rule if the
mammography machine is accredited, or is in the process of becoming accredited, by the
American college of radiology. To have the requirements of subrule (1) of this rule
waived, an applicant shall provide, to the department, within 45 days of the department's
request, copies of the applicant's current accreditation application, current accreditation-
related correspondence to and from the American college of radiology, or current
accreditation certificate that is issued by the American college of radiology.
(3) Failure of an applicant to submit the information required by the provisions of
either subrule (1) or (2) of this rule within 45 days of the department's request may be
considered a basis for withdrawal or denial of the mammography authorization, unless
the time limit is extended by the department for cause.