R 333.5687
R 333.5687 Radiologic technologist quality control tests.
Cite as Mich. Admin. Code R 333.5687
Rule 687. A stereotactic breast biopsy facility shall have a radiologic technologist
perform the following quality control tests at the intervals specified in this rule:
(a) A localization accuracy test shall be performed daily before the equipment is used
on patients. Each of the indicated needle tip coordinates shall be within 1 millimeter of
the actual preset needle tip location.
(b) A phantom image evaluation shall be performed at least weekly. The phantom
image shall achieve at least the minimum score established in R 333.5689.
(c) A hard copy output quality test shall be performed at least monthly, if hard copies
are produced from digital data.
(d) A compression test shall be performed at least semiannually. The maximum
compression force for the power drive mode shall be between 25 pounds and 45 pounds.
(e) A repeat analysis shall be performed at least semiannually. If the overall repeat or
reject rate exceeds 20% based on an image volume of not less than 150 patients, the
reason for the change shall be determined. A repeat analysis shall be assessed
semiannually even if fewer than 150 patients are examined during that period.
(f) If stereotactic breast biopsy is performed using a screen-film system, the
following tests shall be required:
(i) A processor quality control test shall be performed at least daily. Film processors
used to develop stereotactic breast biopsy films shall be adjusted and maintained to meet
the technical development specifications for the mammography film in use. A processor
performance test shall be performed at the beginning of each operational day before
processing any clinical images. The test shall use the mammography film used clinically
at the facility and shall include an assessment of base plus fog density, mid-density, and
density difference as follows:
(A) The base plus fog density shall be within 0.03 of the established operating level.
(B) The mid-density shall be within plus or minus 0.15 of the established operating
level.
(C) The density difference shall be within plus or minus 0.15 of the established
operating level.
(ii) An analysis of fixer retention in film assessed at least quarterly. The residual
fixer shall be not more than 5 micrograms per square centimeter.
(iii) A screen-film contact test shall be performed at least semiannually. Testing for
screen-film contact shall be conducted using 40 mesh copper screen. All cassettes used
in the facility for stereotactic breast biopsy shall be tested.
(iv) A test of darkroom fog shall be performed at least semiannually. The optical
density attributable to darkroom fog shall not exceed 0.05 when a mammography film of
the type used in the facility, which has a mid-density of not less than 1.2 optical density,
is exposed to typical darkroom conditions for 2 minutes while the film is placed on the
counter top emulsion side up.