R 333.5690

R 333.5690 Dosimetry.

Last amended: 2016Year: 2026Length: 44 wordsOfficial source

Cite as Mich. Admin. Code R 333.5690

Rule 690. The average glandular dose delivered during a single exposure of a department-approved phantom simulating a standard breast shall not exceed 3.0 milligrays (300 millirads) per exposure. The dose shall be determined with technique factors and conditions used clinically for a standard breast.
R 333.5690: R 333.5690 Dosimetry. | Justis AI