R 333.5692 Use of quality assurance test results.

Last amended: 2016Year: 2026Length: 132 wordsOfficial source

Cite as Mich. Admin. Code R 333.5692

Rule 692. (1) After completion of tests specified in R 333.5687 to R 333.5691, the facility shall compare the test results to the corresponding specified action limits or the limits established by the facility to verify the image quality of mobile units following a move. (2) If the test results fall outside of the action limits, the source of the problem shall be identified and corrective actions shall be taken within the following time frames: (a) Before any further examinations are performed or any films are processed using a component of the mammography system that failed any of the tests described in R 333.5687(a),(b),(d),(f)(i), (f)(iii), (f)(iv); R 333.5688(g) and (h); or R 333.5691. (b) Within 30 days of the test date for all other tests described in R 333.5687 to R 333.5691.