R 333.5694
R 333.5694 Mammography equipment evaluations.
Cite as Mich. Admin. Code R 333.5694
Rule 694. (1) Additional evaluations of stereotactic breast biopsy units or image
processors shall be conducted when a new unit or processor is installed, a unit or
processor is disassembled and reassembled at the same or a new location, or major
components of a stereotactic breast biopsy unit or processor equipment are changed or
repaired. These evaluations shall be used to determine whether the new or changed
equipment meets the requirements of R 333.5676 to R 333.5678 and R 333.5687 to
R 333.5691, as applicable. Problems revealed by the evaluation shall be corrected before
the new or changed equipment is put into service for procedures or film processing.
(2) The equipment evaluations specified in subrule (1) of this rule shall be performed
by a medical physicist or by an individual under the direct supervision of a medical
physicist.