R 333.5696 Calibration of air kerma measuring instruments.

Last amended: 2016Year: 2026Length: 71 wordsOfficial source

Cite as Mich. Admin. Code R 333.5696

Rule 696. Instruments used by a medical physicist in his or her annual survey to measure the air kerma or air kerma rate from a stereotactic breast biopsy unit shall be calibrated once every 2 years and each time the instrument is repaired. The instrument calibration shall be traceable to a national standard and calibrated with an accuracy of plus or minus 6% (95% confidence level) in the mammography energy range.