R 333.5711
R 333.5711 Enclosures.
Cite as Mich. Admin. Code R 333.5711
Rule 711. (1) A fixed CT scanner enclosure shall be a permanent part of the building
or equipment. Portable shields shall not be used for permanent installations.
(2) The degree of protection required for a CT scanner enclosure shall be determined
by the workload, use and occupancy factors, and the tube potential, tube current,
mechanical movement, and distance. The design shall be subject to approval by the
department.
(3) Protective barriers shall be provided in the ceiling, floor, and walls of a fixed CT
scanner enclosure.
(4) The control panel for a fixed CT scanner shall be shielded by a protective barrier
which cannot be removed from a protective position between the operator and the
radiation source during machine operation.
(5) Movable barriers with electrical interlocks shall not be approved in place of
compliance with subrule (4) of this rule.
(6) The operator of a fixed CT scanner shall be able to see and communicate with the
patient from a shielded position at the control panel. When an observation window is
provided, it shall have a lead equivalence at least equal to that required of the control
barrier in which it is installed.
(7) Mobile or portable CT scanners used routinely in 1 location shall be considered a
fixed installation and shall meet the requirements of subrules (1) to (6) of this rule.