R 333.5717
R 333.5717 Quality control program.
Cite as Mich. Admin. Code R 333.5717
Rule 717. (1) A CT facility shall establish and implement a quality control program
under the supervision of the medical physicist. The documented program shall include
evaluation of all of the following:
(a) Image quality.
(b) Patient radiation dose.
(c) Personnel radiation protection.
(d) Compliance with the provisions of this part.
(e) Ongoing quality control.
(2) Evaluations and tests shall be performed following written procedures and
methods. Corrective action shall be taken and documented according to instructions
provided by the medical physicist if the results of an evaluation or test fall outside the
control limits.
(3) The medical physicist shall determine the frequency of each test and who may
perform the test. An on-site CT radiologic technologist shall be identified to be
responsible for the ongoing quality control testing. The tests shall be performed by this
technologist or by other personnel qualified by training and experience following written
procedures and methods under subrule (2) of this rule.