R 336.2014
R 336.2014 Reference test method 5E.
Cite as Mich. Admin. Code R 336.2014
Rule 1014. Reference method 5E, determination of particulate matter emissions from
positive pressure fabric filters, reads as follows:
(a) The principle, applicability, and performance test criteria are as follows:
(i) Principle. Particulate matter is withdrawn isokinetically from the source and collected
on a glass fiber filter maintained at a temperature at or above the exhaust gas temperature
up to a nominal 248 ±25 degrees Fahrenheit. The particulate mass, which includes any
material that condenses at or above the filtration temperature, is determined gravimetrically
after the removal of uncombined water.
(ii) Applicability. This method is applicable for the determination of particulate
emissions from the stationary sources as identified in table 31 of R 336.1331. The method
is also applicable when specifically provided for in the department's rules, orders, a permit
to install, or a permit to operate.
(iii) Performance test criteria as follows:
(A) A performance test must meet the requirements under R 336.2003(2).
(B) For sources that are subject to an emission limitation calculated to 50% excess air,
the multipoint, integrated sampling procedure of R 336.2004(1)(c) must be used for gas
analysis. For all other sources that require a determination of the molecular weight of the
exhaust, an optional sampling procedure of R 336.2004(1)(c) may be used. Alternatives or
modifications to procedures are subject to the approval of the department.
(C) The minimum volume per sample must be 30 cubic feet actual gas. Minimum
sample time must be 60 minutes, which may be continuous or a combination of shorter
sampling periods for sources that operate in a cyclic manner. Smaller sampling times or
sample volumes, when necessitated by process variables or other factors, may be approved
by the department.
(D) For a source whose emission control device alters the moisture content of the
exhaust gas, a moisture determination must be performed in a location upstream from the
emission control device and in accordance with R 336.2004(1)(d) or an alternative method
approved by the department.
(b) The following provisions apply to apparatus:
(i) Sampling train. A schematic of the sampling train used in this method is shown in
figure 103 under R 336.2021. Construction details for many, but not all, of the train
components are given in APTD-0581, subdivision (g)(ii) of this rule. For changes from the
APTD-0581 document and for allowable modifications to figure 103 under
R 336.2021, consult with the department. The operating and maintenance procedures for
many, but not all, of the sampling train are described in APTD-0576, adopted by reference
in R 336.1902 and as referenced under subdivision (g)(iii) of this rule. Since correct usage
is important in obtaining valid results, all users shall read APTD-0576 and adopt the
applicable operating and maintenance procedures outlined in it, unless otherwise specified
in these rules. The sampling train consists of the following components:
(A) Probe nozzle. Stainless steel 316 or glass with sharp, tapered leading edge. The
angle of taper must be less than 30 degrees and the taper must be on the outside to preserve
a constant internal diameter. The probe nozzle must be of the button-hook design, unless
otherwise specified by the department. If made of stainless steel, the nozzle must be
constructed from seamless tubing. Other materials of construction may be used, subject to
the approval of the department. A range of nozzle sizes suitable for isokinetic sampling
must be available, for example, 0.32 to 1.27 centimeters, 1/8 to 1/2 inch, or larger if higher
volume sampling trains are used inside diameter nozzles in increments of 0.16 centimeter,
1/16 inch. Each nozzle must be calibrated according to the procedures outlined in
subdivision (e) of this rule.
(B) Probe liner. Borosilicate or quartz glass tubing with a heating system capable of
maintaining a gas temperature at the exit end during sampling of 120 ±14 degrees
Centigrade, 248 ±25 degrees Fahrenheit, another temperature as specified by the
department's rules, or a temperature approved by the department for a particular
application. The tester may opt to operate the equipment at a temperature lower than that
specified. Since the actual temperature at the outlet of the probe is not usually monitored
during sampling, probes constructed according to APTD-0581 that utilize the calibration
curves of APTD-0576, or calibrated according to the procedure outlined in APTD-0576,
are acceptable. Either borosilicate or quartz glass probe liners may be used for stack
temperatures up to about 480 degrees Centigrade, 900 degrees Fahrenheit. Quartz liners
must be used for temperatures between 480 and 900 degrees Centigrade, 900 and 1,650
degrees Fahrenheit. Both types of liners may be used at higher temperatures than specified
for short periods of time, subject to the approval of the department. The softening
temperature for borosilicate is 820 degrees Centigrade, 1,508 degrees Fahrenheit, and for
quartz it is 1,500 degrees Centigrade, 2,732 degrees Fahrenheit. When practical, every
effort must be made to use borosilicate or quartz glass probe liners. Alternatively, metal
liners, such as 316 stainless steel, Incoloy 825, or other corrosion-resistant materials made
of seamless tubing, may be used, subject to the approval of the department.
(C) Pitot tube. Type S, as described in method 2, or other device approved by the
department. The pitot tube must be attached to the probe, as shown in figure 103 under R
336.2021, to allow constant monitoring of the stack gas velocity. The impact, high
pressure, opening plane of the pitot tube must be even with or above the nozzle entry plane,
see method 2, figure 2-6 Velocity Traverse Data during sampling. The type S pitot tube
assembly must have a known coefficient, determined as outlined in method 2.
(D) Differential pressure gauge. Incline manometer or equivalent devices, quantity 2,
as described in method 2. One manometer must be used for velocity head (p) readings, and
the other must be used for orifice differential pressure readings.
(E) Filter holders. Two separate filter holders in series or 1 filter holder with separate
filter supports and seals for 2 filters. One filter holder with 2 filters held in contact with
each other is not acceptable. Materials of construction may be stainless steel 316, glass,
Teflon, or other material approved by the department.
(F) Filter heating system. Any heating system capable of maintaining a temperature
around the filter holder during sampling of 120 ±14 degrees Centigrade, 248 ±25 degrees
Fahrenheit, another temperature as specified by the department's rules or a permit
condition, or a temperature approved by the department for a particular application.
Alternatively, the tester may opt to operate the equipment at a temperature lower than that
specified. A temperature gauge capable of measuring temperature to within 3 degrees
Centigrade, 5.4 degrees Fahrenheit, must be installed so that the temperature around the
filter holders can be regulated and monitored during sampling. Heating systems other than
the one shown in APTD-0581 may be used.
(G) Condenser. The following system must be used to determine the stack gas moisture
content: Three impingers connected in series with leak-free ground glass fittings or any
similar leak-free noncontaminating fittings. All impingers must be of the Greenburg-Smith
design and must be modified by replacing the tip with a 1.3 centimeters, 1/2 inch, inside
diameter glass tube extending to about 1.3 centimeters, 1/2 inch, from the bottom of the
flask. Modifications, such as using flexible connections between the impingers or using
materials other than glass, are allowed subject to the approval of the department. The first
impinger must contain a known quantity of water, as described in subdivision (d)(i)(C) of
this rule, the second must be empty, and the third must contain a known weight of silica
gel or equivalent desiccant. Alternatively, any system that cools the sample gas stream and
allows measurement of the water condensed and moisture leaving the condenser, each to
within 1 milliliter or 1 gram, may be used subject to the approval of the department. In any
case, the means for measuring the moisture leaving the condenser must be by passing the
sample gas stream through a tared silica gel, or equivalent desiccant, trap with exit gases
kept below 20 degrees Centigrade, 68 degrees Fahrenheit, and determining the weight gain.
If a determination of the particulate matter collected in the impingers is required by the
department's rules, a permit to install, or a permit to operate, the impinger system described
in this subdivision must be used, without modification. Contact the department as to the
sample recovery and analysis of the impinger contents.
(H) Metering system. Vacuum gauge, leak-free pump, thermometers capable of
measuring temperature to within 3 degrees Centigrade, 5.4 degrees Fahrenheit, dry-gas
meter capable of measuring volume to within 2%, and related equipment as shown in figure
103 under R 336.2021. Other metering systems capable of maintaining sampling rates
within 10% of isokinetic and capable of determining sample volumes to within 2% may be
used, subject to the approval of the department. When the metering system is used in
conjunction with a pitot tube, the system must enable checks of isokinetic rates. Sampling
trains utilizing metering systems designed for higher flow rates than those described in
APTD-0581 or APTD-0576, adopted by reference in R 336.1902, may be used if the
specifications of this rule are met.
(I) Barometer. Mercury, aneroid, or other barometer capable of measuring atmospheric
pressure to within 2.5 millimeters mercury, 0.1-inch mercury. In many cases, the
barometric reading may be obtained from a nearby national weather service station. When
obtained from this source, the station value, which is the absolute barometric pressure, must
be requested and an adjustment for elevation differences between the weather station and
sampling point must be applied at a rate of minus 2.5 millimeters mercury, 0.1-inch
mercury, per 30 meters, 100 foot, elevation increase or vice versa for elevation decrease.
(J) Gas density determination equipment. Temperature sensor and pressure gauge, as
described in method 2, and gas analyzer, if necessary, as described in method 3. The
temperature sensor must, preferably, be permanently attached to the pitot tube or sampling
probe in a fixed configuration so that the tip of the sensor extends beyond the leading edge
of the probe sheath and does not touch metal. Alternatively, the sensor may be attached
just before use in the field. Note, however, that if the temperature sensor is attached in the
field, the sensor must be placed in an interference-free arrangement with respect to the type
S pitot tube openings, see method 2, figure 2.6 Velocity Traverse Data. As a second
alternative, if a difference of not more than 1% in the average velocity measurement is to
be introduced, the temperature gauge need not be attached to the probe or pitot tube. This
alternative is subject to the approval of the department.
(ii) Sample recovery. The following items are needed:
(A) Probe-liner and probe-nozzle brushes. Nylon bristle brushes with stainless steel wire
handles. The probe brush must have extensions, at least as long as the probe, made of
stainless steel, nylon, Teflon, or similarly inert material. The brushes must be properly
sized and shaped to brush out the probe liner and nozzle.
(B) Wash bottles - 2. Glass wash bottles are recommended. Polyethylene wash bottles
may be used at the option of the tester. It is recommended that acetone not be stored in
polyethylene bottles for longer than a month.
(C) Glass sample storage containers. Chemically resistant, borosilicate glass bottles, for
acetone washes, 500 milliliters or 1000 milliliters. Screw cap liners must either be rubber-
backed Teflon or must be constructed so as to be leak-free and resistant to chemical attack
by acetone. Narrow-mouth glass bottles have been found to be less prone to leakage.
Alternatively, polyethylene bottles may be used.
(D) Filter containers. Glass, polyethylene, or aluminum tube containers, unless
otherwise specified by the department.
(E) Graduated cylinder or balance. To measure condensed water to within 1 milliliter
or 1 gram. Graduated cylinders must have subdivisions of not more than 2 milliliters. Most
laboratory balances are capable of weighing to the nearest 0.5 gram or less. Any of these
balances are suitable for use here and in paragraph (iii)(D) of this subdivision.
(F) Plastic storage containers. Airtight containers to store silica gel.
(G) Funnel and rubber policeman, to aid in the transfer of silica gel to container, but not
necessary if silica gel is weighed in the field.
(H) Funnel. Glass or polyethylene, to aid in sample recovery.
(iii) Analysis. The following equipment is needed for analysis:
(A) Glass weighing dishes.
(B) Desiccator.
(C) Analytical balance. To measure to within 0.1 milligrams.
(D) Balance. To measure to within 0.5 milligrams.
(E) Beakers. 250 milliliters.
(F) Hygrometer, to measure the relative humidity of the laboratory environment.
(G) Temperature gauge, to measure the temperature of the laboratory environment.
(c) The following provisions apply to reagents:
(i) Sampling. The reagents used in sampling are as follows:
(A) Filters. Two outstack filters may be any combination of alundum ceramic thimble
filters, type RA-98 or glass fiber filters, type A without organic binder. The size of such
filters must allow proper sampling rates to maintain isokinetics using the nozzle sizes
specified in subdivision (b)(i)(A) of this rule. Alternatively, other types of filters may be
used, subject to the approval of the department.
(B) Silica gel. Indicating type, 6 to 16 mesh. If previously used, dry at 175 degrees
Centigrade, 350 degrees Fahrenheit, for 2 hours. New silica gel may be used as received.
Alternatively, other types of desiccants, the equivalent or better of silica gel, may be used,
subject to the approval of the department.
(C) Water. When analysis of the material caught in the impingers is required, distilled
water must be used. Run blanks before field use to eliminate a high blank on test samples.
(D) Crushed ice.
(E) Stopcock grease. Acetone-insoluble, heat-stable silicone grease. This is not
necessary if screw-on connectors with Teflon sleeves, or equivalent, are used.
Alternatively, other types of stopcock grease may be used, subject to the approval of the
department.
(ii) Sample recovery. Washing solvent. Either acetone or distilled water may be used for
sample recovery. If acetone is used for washing solvent, then reagent grade, less than
0.001% residue, in glass bottles is required. Acetone from metal containers generally has
a high residue blank and must not be used. Suppliers sometimes transfer acetone to glass
bottles from metal containers. Thus, acetone blanks must be run before field use, and only
acetone with low blank values, less than 0.001%, must be used. In no case must a blank
value of more than 0.001% of the weight of acetone used be subtracted from the sample
weight. If distilled water is used for washing solvent, use distilled water with less than
0.001% residue. Run blanks before field use to eliminate a high blank on test samples.
(iii) Analysis. Two reagents are required for the analysis:
(A) Solvent. Same as paragraph (ii) of this subdivision for quantitative transfer.
(B) Desiccant. Anhydrous calcium sulfate, indicating type. Alternatively, other types of
desiccants may be used, subject to the approval of the department.
(d) The following provisions apply to procedures:
(i) Determination of single measurement sites. The measurement site for a positive
pressure fabric filter with an exhaust stack meeting method 1 criteria must be in accordance
with method 1. The measurement site for positive pressure fabric filters with short stacks
or physical configuration not amenable to the requirements of method 1 must be
determined from the following alternatives, or as approved by the department:
(A) Short stacks not meeting method 1 criteria: Short stacks may be extended in
accordance with the procedures set forth in method 1 or by the use of flow straightening
vanes. The flow straightening vanes must be of the egg crate design, see figure 109 under
R 336.2021. The measurement site, when using straightening vanes, must be at a distance
not less than 2 times the average equivalent diameter of the vane opening and not less than
half of the overall stack diameter upstream of the stack outlet.
(B) Roof monitor or monovent exhaust outlets: For positive pressure fabric filters
equipped with peaked roof monitors, ridge vents, or other types of monovents, use a
measurement site at the base of the monovent. Examples of the locations are shown in
figure 108 under R 336.2021. The measurement site must be upstream of any exhaust point.
(C) Measurement site in fabric filter compartment housing. Sample immediately
downstream of the filter bags directly as shown in the examples in figure 108 under R
336.2021. Depending on the housing design, use sampling ports in the housing walls or
locate the sampling equipment within the compartment housing.
(ii) Determination of number and location of traverse points. The number and location
of traverse points for single exhaust stacks on positive pressure fabric filters meeting
method 1 criteria must be in accordance with method 1. The number of traverse points for
other single measurement sites not meeting method 1 criteria must not be less than 24. For
example, a rectangular measurement site, such as a monovent, would require the use of a
balanced 5-by-5 traverse point matrix. All traverse points must be sampled for each test
run.
(iii) Multiple measurement sites. Sampling from 2 or more stacks or measurement sites
may be combined for a test run, if all of the following requirements are met:
(A) All measurement sites up to 12 must be sampled. For more than 12 measurement
sites, conduct sampling on not less than 12 sites or 50% of the sites, whichever is greater.
The measurement sites sampled must be evenly, or nearly evenly, distributed among the
available sites, if not all of the sites are to be sampled.
(B) The same number of measurement sites must be sampled for each test run.
(C) The minimum number of traverse points per test run is 24. An exception to the 24-
point minimum would be a test combining the sampling from 2 stacks meeting method 1
criteria for acceptable stack length, and method 1 specifies fewer than 12 points per site.
(D) As long as the 24 traverse points per test run criterion is met, the number of traverse
points per measurement site may be reduced to 8. Alternatively, conduct a test run for each
measurement site individually using the criteria in this paragraph and paragraph (ii) of this
subdivision for the number of traverse points. Each test must count toward the total of 3
required for a performance test. If more than 3 measurement sites are sampled, the number
of traverse points per measurement site may be reduced to 8 if not less than 72 traverse
points are sampled for all 3 tests.
(iv) Sampling. The complexity of this method is such that, in order to obtain reliable
results, testers must be trained and experienced with the test procedures. Sampling must
comply with the following provisions:
(A) Pretest preparation. All the components must be maintained and calibrated
according to the applicable procedures described in APTD-0576, adopted by reference in
R 336.1902, unless otherwise specified in this rule. Weigh several 200 to 300 gram portions
of silica gel in airtight containers to the nearest 0.5 gram. Record the total weight of the
silica gel plus container on each container. As an alternative, the silica gel need not be
preweighed, but may be weighed directly in its impinger or sampling holder just before
train assembly. Check filters visually against light for irregularities, flaws, pinhole leaks,
or cracks. Label filters of the proper size on the back side using numbering machine ink.
As an alternative, label the shipping containers, subdivision (b)(ii)(D) of this rule, and keep
the filters in these containers at all times, except during sampling and weighing. Dry the
filters in an oven at 105 degrees Centigrade, 220 degrees Fahrenheit, for a minimum of 2
hours, cool for at least 1 hour in a desiccator containing anhydrous calciumsulfate, and
individually weigh and record each weight to the nearest 0.1 milligram. During the
weighing, the filters must not be exposed to the laboratory atmosphere for a period of more
than 2 minutes and a relative humidity above 50%. Procedures, other than those specified,
that account for relative humidity effects may be used, subject to the approval of the
department.
(B) Preliminary determinations. Select the sampling site and the minimum number of
sampling points according to method 1 or as specified by the department. Determine the
stack pressure, temperature, and the range of velocity heads using method 2. It is
recommended that a leak check of the pitot lines, see method 2, be performed. Determine
the moisture content using approximation method 4, or its alternatives, for the purpose of
making isokinetic sampling rate settings. Determine the stack gas dry molecular weight, as
described in method 2. If integrated method 3 sampling is used for molecular weight
determination, the integrated bag sample must be taken simultaneously with, and for the
same total length of time as, the particulate sample run. Select a nozzle size based on the
range of velocity heads so that it is not necessary to change the nozzle size in order to
maintain isokinetic sampling rates. During the run, do not change the nozzle size. Ensure
that the proper differential pressure gauge is chosen for the range of velocity heads
encountered, see method 2. Select a suitable probe liner and probe length so that all traverse
points can be sampled. For large stacks, consider sampling from opposite sides of the stack
to reduce the length of probes. Select a total sampling time greater than or equal to the
minimum total sampling time specified in the test procedures for the specific industry so
that the sampling time per point is not less than 5 minutes, unless approved by the
department, or some greater time interval as specified by the department, and so that the
sample volume taken, corrected to standard conditions, exceeds the required minimum total
gas sample volume. The latter is based on an approximate average sampling rate. It is
recommended that the number of minutes sampled at each point be an integer or an integer
plus 1/2 minute to avoid timekeeping errors. In some circumstances, such as in batch
cycles, it may be necessary to sample for shorter times at the traverse points and to obtain
smaller gas sample volumes. In these cases, the department's approval must first be
obtained.
(C) Preparation of collection train. During preparation and assembly of the sampling
train, keep all openings where contamination can occur covered until just before assembly
or until sampling is about to begin. Place 100 milliliters of water in the first impinger, leave
the second impinger empty, and transfer approximately 200 to 300 grams of preweighed
silica gel from its container to the third impinger. More silica gel may be used, but care
should be taken to ensure that it is not entrained and carried out from the impinger during
sampling. Place the container in a clean place for later use in the sample recovery.
Alternatively, the weight of the silica gel plus impinger may be determined to the nearest
0.5 gram and recorded. Using tweezers or clean disposable surgical gloves, place a labeled,
identified, and weighed filter in the filter holder. Be sure that the filter is properly centered
and the gasket properly placed so as to prevent the sample gas stream from circumventing
the filter. Check the filter for tears after assembly is completed. When glass liners are used,
install the selected nozzle using a Viton A O-ring when stack temperatures are less than
260 degrees Centigrade, 500 degrees Fahrenheit, and a heat-resistant fiberglass, graphite,
or other material string gasket when temperatures are higher. See APTD-0576, adopted by
reference in R 336.1902, for details. Other connecting systems using either 310 stainless
steel or Teflon ferrules may be used. When metal liners are used, install the nozzle in the
same manner as for glass liners or by a leak-free direct mechanical connection. Mark the
probe with heat-resistant tape or by some other method to denote the proper distance into
the stack or duct for each sampling point. Set up the train as in figure 103 under R 336.2021.
If necessary, use a very light coat of silicone grease on all ground glass joints. Grease only
the outer portion, see APTD-0576, to avoid the possibility of contamination by the silicone
grease. Place crushed ice around the impingers.
(D) Leak check procedures:
(I) Pretest leak check. A pretest leak check is strongly recommended, but not required,
to prevent invalid sampling and wasted time. If the tester opts to conduct the pretest leak
check, the following procedure must be used: After the sampling train has been assembled,
turn it on and set the filter and probe heating systems at the desired operating temperatures.
Allow time for the temperatures to stabilize. If a Viton A O-ring or other leak-free
connection is used in assembling the probe nozzle to the probe liner, leak check the train
at the sampling site by plugging the nozzle and pulling a 380 millimeter mercury, 15 inch
mercury, vacuum. A lower vacuum may be used if it is not exceeded during the test. If a
heat-resistant fiberglass, graphite, or other material string is used, do not connect the probe
to the train during the leak check. Instead, leak check the train by first plugging the inlet to
the filter holder, cyclone, if applicable, and pulling a 380 millimeter mercury, 15 inch
mercury, vacuum. A lower vacuum may be used if it is not exceeded during the test. Then
connect the probe to the train and leak check at about a 25 millimeter mercury, 1 inch
mercury, vacuum. Alternatively, the probe may be leak checked with the rest of the
sampling train, in 1 step, at a 380 millimeter mercury, 15 inch mercury, vacuum. Leakage
rates in excess of 4% of the average sampling rate or 0.00057 cubic meters per minute,
0.02 cubic feet per minute,, whichever is less, are unacceptable. The following leak check
instructions for the sampling train described in APTD-0576 and APTD-058 may be helpful.
Start the pump with the bypass valve fully open and the coarse adjust valve completely
closed. Partially open the coarse adjust valve and slowly close the bypass valve until the
desired vacuum is reached. Do not reverse the direction of the bypass valve, as this will
cause water to back up into the filter holder. If the desired vacuum is exceeded, either leak
check at this higher vacuum or end the leak check and start over. When the leak check is
completed, first slowly remove the plug from the inlet to the probe, filter holder, or cyclone,
if applicable, and immediately turn off the vacuum pump. This prevents the water in the
impingers from being forced backward into the filter holder and prevents silica gel from
being entrained backward into the third impinger.
(II) Leak checks during sample run. If, during the sampling run, a component, such as
a filter assembly or impinger, change becomes necessary, a leak check must be conducted
immediately before the change is made. The leak check must be done according to the
procedure outlined in paragraph (iv)(D)(I) of this subdivision, except that it must be done
at a vacuum equal to or greater than the maximum value recorded up to that point in the
test. If the leakage rate is found to be not more than 0.00057 cubic meters per minute, 0.02
cubic feet per minute, or 4% of the average sampling rate, whichever is less, the results are
acceptable and no correction need be applied to the total volume of dry gas metered. If,
however, a higher leakage rate is obtained, the tester shall either record the leakage rate
and plan to correct the sample volume, as shown in subdivision (f)(iii) of R 336.2011, or
shall void the sampling run. Immediately after component changes, leak checks are
optional. If the leak checks are done, the procedure outlined in paragraph (iv)(D)(I) of this
subdivision must be used.
(III) Post-test leak check. A leak check is mandatory at the conclusion of each sampling
run. The leak check must be done in accordance with the procedures outlined in paragraph
(iv)(D)(I) of this subdivision, except that it must be conducted at a vacuum equal to or
greater than the maximum value reached during the sampling run. If the leakage rate is
found to be not more than 0.00057 cubic meters per minute, 0.02 cubic feet per minute, or
4% of the average sampling rate, whichever is less, the results are acceptable and no
correction need be applied to the total volume of dry gas metered. If, however, a higher
leakage rate is obtained, the tester shall either record the leakage rate and correct the sample
volume, as shown in subdivision (f)(iii) of R 336.2011, or shall void the sampling run.
(E) Particulate train operation. During the sampling run, maintain an isokinetic
sampling rate that is within 10% of true isokinetic, unless otherwise specified by the
department. For each run, record the data required on a data sheet such as the one shown
in figure 104 under R 336.2021. Be sure to record the initial dry-gas meter reading. Record
the dry-gas meter readings at the beginning and end of each sampling time increment, when
changes in flow rates are made, before and after each leak check, and when sampling is
halted. Take other readings required by figure 104 under R 336.2021 at least once at each
sample point during each time increment, and take additional readings when significant
changes, 20% variation in velocity head readings, necessitate additional adjustments in
flow rate. Level and zero the manometer. Because the manometer level and zero may drift
due to vibrations and temperature changes, make periodic checks during the traverse. Clean
the portholes before the test run to minimize the chance of sampling deposited material. To
begin sampling, remove the nozzle cap and verify that the pitot tube and probe are properly
positioned. Position the nozzle at the first traverse point with the tip pointing directly into
the gas stream. Immediately start the pump and adjust the flow to isokinetic conditions.
Nomographs that aid in the rapid adjustment of the isokinetic sampling rate without
excessive computations are available. These nomographs are designed for use when the
type S pitot tube coefficient is 0.85 ±0.02 and the stack gas equivalent density, dry
molecular weight, is equal to 29 ±4. APTD-0576, adopted by reference in R 336.1902,
details the procedure for using the nomographs. If Cp and Md are outside the above stated
ranges, do not use the nomographs unless appropriate steps, see subdivision (g)(iv) of this
rule, are taken to compensate for the deviations. When the stack is under significant
negative pressure, height of impinger stem, take care to pull low flow when inserting the
probe into the stack to prevent water from backing into the sample tubing and to avoid
pulsation through the filter and possible loss of materials. When the probe is in position,
block off the openings around the probe and porthole to prevent unrepresentative dilution
of the gas stream. Traverse the stack cross section, as required by method 1 or as specified
by the department, being careful not to bump the probe nozzle into the stack walls when
sampling near the walls or when removing or inserting the probe through the portholes.
This minimizes the chance of extracting deposited material. During the test run, add more
ice and, if necessary, salt to maintain a temperature of less than 20 degrees Centigrade, 68
degrees Fahrenheit, at the condenser/silica gel outlet. Also, periodically check the level and
zero of the manometer. If the pressure drop across the filter becomes too high and makes
isokinetic sampling difficult to maintain, the filter may be replaced in the midst of a sample
run. It is recommended that another complete filter assembly be used rather than attempting
to change the filter itself. Before a new filter assembly is installed, conduct a leak check,
see paragraph (iv)(D)(II) of this subdivision. The total particulate weight must include the
summation of all filter assembly catches. A single train must be used for the entire sample
run, except in cases where simultaneous sampling is required in 2 or more separate ducts,
at 2 or more different locations within the same duct, or where equipment failure
necessitates a change of trains. In all other situations, the use of 2 or more trains must be
subject to the approval of the department. Note that when 2 or more trains are used, separate
analyses of the front-half catches from the individual trains may be combined, as may the
impinger catches, and 1 analysis of the front-half catch and 1 analysis of impinger catch
may be performed. Consult with the department for details concerning the calculation of
results when 2 or more trains are used. At the end of the sample run, turn off the coarse
adjust valve, remove the probe and nozzle from the stack, turn off the pump, record the
final dry-gas meter reading, and conduct a post-test leak check, as outlined in paragraph
(iv)(D)(III) of this subdivision. Also, leak-check the pitot lines as described in method 2
The lines must pass this leak check to validate the velocity head data.
(F) Calculation of percent isokinetic. Calculate percent isokinetic, see subdivision (f) of
this rule, to determine whether the run was valid or whether another test run should be
made. If there was difficulty in maintaining isokinetic rates due to source conditions,
consult with the department for possible variance on the isokinetic rates.
(v) Sample recovery. Proper cleanup procedure begins as soon as the probe is removed
from the stack at the end of the sampling period. Allow the probe to cool. When the probe
can be safely handled, wipe off all external particulate matter near the tip of the probe
nozzle and place a cap over it to prevent losing or gaining particulate matter. Do not cap
off the probe tip tightly while the sampling train is cooling down as this creates a vacuum
in the filter holder and draws water from the impingers into the filter holder. Before moving
the sample train to the cleanup site, remove the probe from the sample train, wipe off the
silicone grease, and cap the open outlet of the probe. Be careful not to lose any condensate
that might be present. Wipe off the silicone grease from the filter inlet where the probe was
fastened and cap it. Remove the umbilical cord from the last impinger and cap the impinger.
If a flexible line is used between the first impinger or condenser and the filter holder,
disconnect the line at the filter holder and let any condensed water or liquid drain into the
impingers or condenser. After wiping off the silicone grease, cap off the filter holder outlet
and impinger inlet. Ground-glass stoppers, plastic caps, or serum caps may be used to close
these openings. Transfer the probe and filter-impinger assembly to the cleanup area. This
area must be clean and protected from the wind so that the chances of contaminating or
losing the sample are minimized. Save a portion of the solvent used for cleanup as a blank.
Take 200 milliliters of this solvent directly from the wash bottle being used and place it in
a glass sample container labeled "solvent blank." Inspect the train before and during
disassembly and note any abnormal conditions. Treat the samples as follows: Container
numbers 1, 1A. Carefully remove the filters from the filter holders and place each filter in
its identified container. Use a pair of tweezers or clean disposable surgical gloves, or both,
to handle the filters. Carefully transfer to the container any particulate matter or filter fibers,
or both, that adhere to the filter holder gasket by using a dry nylon bristle brush or sharp-
edged blade, or both. Seal the container. Container number 2. Taking care to see that dust
on the outside of the probe or other exterior surfaces does not get into the sample, the tester
shall quantitatively recover from particulate matter or condensate from the nozzle, probe
fitting, probe liner, and from both filter holders by washing these components with solvent
and placing the wash in a glass container. Perform the solvent rinses as follows: Carefully
remove the probe nozzle and clean the inside surface by rinsing with solvent from a wash
bottle and brushing with a nylon bristle brush. Brush until the solvent rinse shows no visible
particles and then make a final rinse of the inside surface with solvent. Brush and rinse the
inside parts of the Swagelok fitting with solvent in a similar way until no visible particles
remain. Rinse the probe liner with solvent by tilting and rotating the probe while squirting
solvent into its upper end so that all inside surfaces are wetted with acetone. Let the solvent
drain from the lower end into the sample container. A glass or polyethylene funnel may be
used to aid in transferring liquid washes to the container. Follow the solvent rinse with a
probe brush. Hold the probe in an inclined position and squirt solvent into the upper end as
the probe brush is being pushed with a twisting action through the probe. Hold a sample
container underneath the lower end of the probe and catch any solvent and particulate
matter that is brushed from the probe. Run the brush through the probe 3 or more times
until no visible particulate matter is carried out with the solvent or until none remains in
the probe liner on visual inspection. With stainless steel or other metal probes, run the brush
through, in the manner set forth in this paragraph, not less than 6 times, since metal probes
have small crevices in which particulate matter can be entrapped. Rinse the brush with
solvent and quantitatively collect these washings in the sample container. After the
brushing, make a final solvent rinse of the probe as described above. It is recommended
that 2 people be used to clean the probe to minimize sample losses. Between sampling runs,
keep brushes clean and protected from contamination. After ensuring that all joints have
been wiped clean of silicone grease, clean the inside of both filter holders by rubbing the
surfaces with a nylon bristle brush and rinsing with solvent. Rinse each surface 3 times, or
more if needed, to remove visible particulate. Make a final rinse of the brush and filter
holder. After all solvent washings and particulate matter have been collected in the sample
container, tighten the lid on the sample container so that solvent will not leak out when it
is shipped to the laboratory. Mark the height of the fluid level to determine whether or not
leakage occurred during transport. Label the container to clearly identify its contents.
Container number 3. Note the color of the indicating silica gel to determine if it has been
completely spent and make a notation of its condition. Transfer the silica gel from the third
impinger to its original container and seal. A funnel may make it easier to pour the silica
gel without spilling it. A rubber policeman may be used as an aid in removing the silica gel
from the impinger. It is not necessary to remove the small amount of dust particles that
adhere to the impinger wall and are difficult to remove. Since the gain in weight is to be
used for moisture calculations, do not use any water or other liquids to transfer the silica
gel. If a balance is available in the field, follow the procedure for container number 3 in
paragraph (vi) of this subdivision. Impinger water. Treat the impingers as follows: Make a
notation of any color or film in the liquid catch. Measure the liquid that is in the first 2
impingers to within ±1 milliliter by using a graduated cylinder or by weighing it to within
±1.0 gram by using a balance if none is available. Record the volume or weight of liquid
present. This information is required to calculate the moisture content of the effluent gas.
Discard the liquid after measuring and recording the volume or weight, unless analysis of
the impinger catch is required, see subdivision (b)(i)(G) of this rule. If a different type of
condenser is used, measure the amount of moisture condensed either volumetrically or
gravimetrically. When possible, containers must be shipped in a manner that keeps them
upright at all times.
(vi) Analysis. Record the data required on a sheet such as the one shown in figure 106
under R 336.2021. Handle each sample container as follows: Container numbers 1, 1A.
Analyze and report each filter separately. Transfer the filter and any loose particulate from
the sample container to a tared-glass weighing dish. Dry the filter in an oven at 105 degrees
Centigrade, 220 degrees Fahrenheit, for a minimum of 2 hours, cool for at least 1 hour in
a desiccator containing anhydrous calcium sulfate, and weigh and record its weight to the
nearest 0.1 milligram. During the weighing, the filter must not be exposed to the laboratory
atmosphere for a period of more than 2 minutes or a relative humidity above 50%.
Procedures, other than those specified, that account for relative humidity effects may be
used, subject to the approval of the department. The method used for the drying and
weighing of filters must be consistent before and after the test. Container number 2. Note
the level of liquid in the container and confirm on the analysis sheet whether or not leakage
occurred during transport. If a noticeable amount of leakage has occurred, either void the
sample or use methods, subject to the approval of the department, to correct the final
results. Measure the liquid in this container either volumetrically to ±1 milliliters or
gravimetrically to ±1.0 grams. Transfer the contents to a tared 250 milliliter beaker and
evaporate to dryness either at ambient temperature and pressure for acetone or at 95 degrees
Centigrade, 203 degrees Fahrenheit, in an oven for distilled water. Then subject the sample
to 250 degrees Centigrade, 482 degrees Fahrenheit, in an oven for 2 to 3 hours. Desiccate
for 24 hours and weigh to a constant weight. Report the results to the nearest 0.1 milligram.
Container number 3. Weigh the spent silica gel, or silica gel plus impinger, to the nearest
0.5 gram using a balance. This step may be conducted in the field. "Solvent blank"
container. Measure solvent in this container either volumetrically or gravimetrically.
Transfer the contents to a tared 250 milliliter beaker and evaporate to dryness either at
ambient temperature and pressure for acetone or at 95 degrees Centigrade, 203 degrees
Fahrenheit, in an oven for distilled water. Then subject the sample to 250 degrees
Centigrade, 482 degrees Fahrenheit, in an oven for 2 to 3 hours. Desiccate for 24 hours and
weigh to a constant weight. Report the results to the nearest 0.1 milligram. If acetone is
used, the contents of container number 2, as well as the acetone blank container, may be
evaporated at temperatures higher than ambient. If evaporation is done at an elevated
temperature, the temperature must be closely supervised, and the contents of the beaker
must be swirled occasionally to maintain an even temperature. Use extreme care, as acetone
is highly flammable and has a low flash point.
(e) Calibration. Maintain a laboratory log of all calibrations. The following provisions
apply to calibrations:
(i) Probe nozzle. A probe nozzle must be calibrated before its initial use in the field.
Using a micrometer, measure the inside diameter of the nozzle to the nearest 0.025
millimeter, 0.001 inch. Make 3 separate measurements using different diameters each time
and obtain the average of the measurements. The difference between the high and low
numbers must not exceed 0.1 millimeter, 0.004 inch. When nozzles become nicked, dented,
or corroded, the nozzles must be reshaped, sharpened, and recalibrated before use. Each
nozzle must be permanently and uniquely identified.
(ii) Pitot tube. The type S pitot tube assembly must be calibrated according to the
procedure outlined in method 2.
(iii) Metering system. Before its initial use in the field, the metering system must be
calibrated according to the procedure outlined in APTD-0576, adopted by reference in R
336.1902. Instead of physically adjusting the dry-gas meter dial readings to correspond to
the wet-test meter readings, calibration factors may be used to mathematically correct the
gas meter dial readings to the proper values. Before calibrating the metering system, it is
suggested that a leak check be conducted. For metering systems having diaphragm or rotary
pumps, the normal leak check procedure will not detect leakages within the pump. For
these cases, the following leak check procedure is suggested: Make a 10-minute calibration
run at 0.00057 cubic meters per minute, 0.02 cubic feet per minute, at the end of the run,
take the difference of the measured wet-test meter and dry-gas meter volumes, and divide
the difference by 10 to get the leak rate. The leak rate must not exceed 0.00057 cubic meters
per minute, 0.02 cubic feet per minute. After each field use, the calibration of the metering
system must be checked by performing 3 calibration runs at a single, intermediate orifice
setting, based on the previous field test, with the vacuum set at the maximum value reached
during the test series. To adjust the vacuum, insert a valve between the wet-test meter and
the inlet of the metering system. Calculate the average value of the calibration factor. If the
calibration has changed by more than 5%, recalibrate the meter over the full range of orifice
settings, as outlined in APTD-0576. Alternatively, a spirometer may be substituted for a
wet-test meter in the above mentioned calibration procedures. Alternative procedures, such
as using the orifice meter coefficients, may be used, subject to the approval of the
department. If the dry-gas meter coefficient values obtained before and after a test series
differ by more than 5%, the test series must be performed using whichever meter
coefficient value, before or after, gives the lower value of total sample volume.
(iv) Probe heater calibration. The probe heating system must be calibrated before its
initial use in the field according to the procedures outlined in APTD-0576, adopted by
reference in R 336.1902. Probes constructed according to APTD-0581 need not be
calibrated if the calibration curves in APTD-0576 are used.
(v) Temperature gauges. Use the procedure in method 2 to calibrate instack temperature
gauges. Dial thermometers, such as those used for the dry-gas meter and condenser outlet,
must be calibrated against mercury-in-glass thermometers or other thermometers that are
calibrated using a National Institute of Standards and Technology calibrated reference
thermometer.
(vi) Leak check of metering system shown in figure 103 under R 336.2021. That portion
of the sampling train from the pump to the orifice meter must be leak checked before initial
use and after each shipment. Leakage after the pump will result in less volume being
recorded than is actually sampled. The following procedure is suggested, also see figure
107 under R 336.2021: Close the main valve on the meter box. Insert a 1-hole rubber
stopper with rubber tubing attached into the orifice exhaust pipe. Disconnect and vent the
low side of the orifice manometer. Close off the low side orifice tap. Pressurize the system
to 13 to 18 centimeters, 5 to 7 inches, water column by blowing into the rubber tubing.
Pinch off the tubing and observe the manometer for 1 minute. A loss of pressure on the
manometer indicates a leak in the meter box. Leaks, if present, must be corrected.
(vii) Barometer. Calibrate against a mercury barometer.
(f) Calculations. When carrying out calculations, retain at least 1 extra decimal figure
beyond that of the acquired data. Round off figures after the final calculation. Other forms
of the equations may be used if the other forms of the equations give equivalent results. All
of the provisions under R 336.2011(f) apply to calculations for this rule.
(g) Bibliography:
(i) Federal Register, Volume 42, No. 160, Part 60, Chapter 1, Title 40, Appendix A
Method 5, August 18, 1977.
(ii) Martin, Robert M. Construction Details of Isokinetic Source Sampling Equipment.
Environmental Protection Agency. Research Triangle Park, N.C.APTD-0581. April, 1971.
(iii) Rom, Jerome J. Maintenance, Calibration, and Operation of Isokinetic Source
Sampling Equipment. Environmental Protection Agency.Research Triangle Park, N.C.
APTD-0576. March, 1972.
(iv) Shigehara, R. T. "Adjustments in the EPA Nomograph for Different Pitot Tube
Coefficients and Dry Molecular Weights." Stack Sampling News 2:4-11.October, 1974.
(v) Guidelines for Source Testing of Particulate. Michigan Department of Natural
Resources, Air Quality Division. June 1, 1977.