19 CSR 25-30.051
Breath Analyzer Calibration and Accuracy Verification Standards
PURPOSE: This rule defines the standard simulator solutions or
compressed ethanol-gas mixtures to be used in verifying and calibrating breath analyzers, as well as the annual checks required
on simulators used in conjunction with the standard simulator
solution.
(1) Standards used for the purpose of verifying and calibrating
breath analyzers shall consist of standard simulator solutions
or compressed ethanol-gas standard mixtures.
(2) Standard simulator solutions, used to verify and calibrate
evidential breath analyzers, shall be solutions from approved
suppliers. The standard simulator solutions used shall have a
vapor concentration within five percent (5%) of the following
values:
(A) 0.10%;
(B) 0.08%; or
(C) 0.04%.
(3) Approved suppliers of standard simulator solutions are—
(A) Alcohol Countermeasure Systems, Inc.;
(B) Guth Laboratories, Inc.;
(C) RepCo Marketing, Inc.; and
(D) Draeger Safety, Inc.
(4) Any breath alcohol simulator used in the verification or
calibration of evidential breath analyzers with the standard
simulator solutions referred to in sections (2) and (3) of this
rule shall be certified against a National Institute of Standards
and Technology (NIST) traceable reference thermometer or
AND SENIOR SERVICES
thermocouple between January 1, 2013, and December 31,
2013, and annually thereafter. Proof that the simulator is in
certification shall be shown by entry of the simulator serial
number and expiration date of the certification period on the
maintenance report required under 19 CSR 25-30.031, or by
certification report as issued by the department.
(5) Compressed ethanol-gas standard mixtures used to verify
and calibrate evidential breath analyzers shall be mixtures provided from approved suppliers. The compressed ethanol-gas
mixtures used shall have a concentration within five percent
(5%) of the following values:
(A) 0.10%;
(B) 0.08%; or
(C) 0.04%.
(6) Approved suppliers of standard compressed ethanol-gas
mixtures are—
(A) Intoximeters, Inc.;
(B) CMI, Inc.;
(C) Draeger Safety Diagnostic, Inc.;
(D) ILMO Products Company, Inc.; and
(E) Guth Laboratories, Inc.
(7) Maintenance reports completed prior to the effective date of
this rule shall be considered valid under this rule if the maintenance report was completed in compliance with the rules in
effect at the time the maintenance was conducted.
AUTHORITY: sections 192.006, 577.020, and 577.037, RSMo 2016.*
Emergency rule filed Aug. 22, 1997, effective Sept. 1, 1997, expired
Feb. 27, 1998. Original rule filed Aug. 25, 1997, effective Feb. 28,
1998. Emergency rescission and emergency rule filed April 17,
1998, effective May 4, 1998, expired Oct. 30, 1998. Rescinded and
readopted: Filed May 1, 1998, effective Oct. 30, 1998. Amended:
Filed Jan. 15, 2004, effective July 30, 2004. Amended: Filed May
31, 2012, effective Dec. 30, 2012. Amended: Filed Sept. 4, 2013, effective Feb. 28, 2014. Amended: Filed Oct. 1, 2019, effective March
30, 2020.
*Original authority: 192.006, RSMo 1993, amended 1995; 577.020, RSMo 1977,
amended 1982, 1983, 1996, 1998, 2001, 2006, 2014; and 577.037, RSMo 1982, amended
1983, 1988, 1993, 1996, 2001, 2014, 2016.