19 CSR 30-30.070
Physical Standards for Abortion Facilities
PURPOSE: Section 197.225, RSMo authorizes the Department of
Health and Senior Services to establish physical standards for
abortion facilities in order to provide acceptable care in a safe
environment. Abortion facilities are defined in section 197.200(1),
RSMo and are subject to licensure under section 197.205, RSMo.
PUBLISHER’S NOTE: The publication of the full text of the material
that the adopting agency has incorporated by reference in this
rule would be unduly cumbersome or expensive. Therefore, the
full text of that material will be made available to any interested
person at both the Office of the Secretary of State and the office of
the adopting agency, pursuant to section 536.031.4, RSMo. Such
material will be provided at the cost established by state law.
(1) This regulation does not apply to abortion facilities that
do not perform surgical abortions or surgical intervention for
abortion complications.
(2) Requests for deviations from requirements on physical
facilities shall be in writing to the Department of Health and
Senior Services. Approvals for deviations shall be in writing
and both requests and approvals shall be made a part of the
permanent Department of Health and Senior Services records
for the abortion facility.
(3) Any abortion facility constructed or renovated after October
25, 1987 shall have plans prepared by an architect or engineer
registered in Missouri. These plans shall be submitted to the
department for review and approval prior to construction. New
abortion facilities shall have the following:
(A) At least two (2) remote exits shall be provided for each
floor directly to the outside or through an enclosed stairway or
passageway to the outside;
(B) Corridors serving patients shall be at least six feet (6')
wide;
(C) All doors through which patients pass shall be at least
forty-four inches (44") wide and of solid-core construction;
(D) One- (1-) story buildings shall be at least of Type II
(111) protected noncombustible construction as described in
Standard on Types of Building Construction 1979 published by
the National Fire Protection Association;
(E) Multistory buildings shall be constructed of at least Type
II (222) fire-resistive construction as described in Standard on
Types of Building Construction published by the NFPA, or shall
be protected throughout by an approved automatic sprinkler
system;
(F) Multistory buildings shall have at least one (1) elevator.
The elevator cab shall be at least five feet by seven feet (5' × 7')
clear inside. The car door shall have a clear opening of not less
than forty-four inches (44");
(G) Trickle-charge battery pack units shall be located to
provide emergency lighting in the procedure room, recovery
room, exit corridors, and exit stairs to grade;
(H) A manual fire alarm break station shall be located near
each exit and connected to a local audible alarm which can be
heard throughout the facility;
(I) At least two (2) ABC-type fire extinguishers are to be
located in the facility, one (1) in the clinical area;
(J) Illuminated exit signs shall be located above each exit
and illuminated directional exit signs shall be located where
needed to direct patients and personnel to exits in event of an
emergency;
(K) Ceiling, wall, and floor finishes in the clinical area
including the procedure rooms, recovery room, personnel
change rooms, central sterile and supply, janitor’s closet, and
laboratory shall be smooth and easily cleanable;
(L) Scrub-up facilities shall be knee- or foot-operated and
provided at the rate of one (1) per procedure room. Scrub-up
facilities shall be located outside but immediately available to
the procedure room;
(M) Procedure rooms shall have the following:
1. A minimum length and width of twelve feet (12');
2. A minimum ceiling height of nine feet (9');
3. A door with a minimum width of forty-four inches (44");
and
4. There shall be no windows in the room except there may
be a fixed-view window in the wall between the procedure
room and the adjacent corridor;
(N) The recovery room shall be separated from the procedure
room and be of sufficient size to accommodate at least four (4)
recovery beds or recliners for each procedure room. There shall
be three feet (3') of clear space on both sides and at the foot of
each recovery bed or recliner;
(O) The procedure room and recovery room shall be provided
with a minimum of six (6) air changes per hour. Air supplied to
all areas shall be filtered through a filter with at least a twentyfive percent (25%) efficiency rating;
(P) Personnel change rooms shall be provided for each sex
and located convenient to the procedure room. Each change
room shall be equipped with a toilet and lavatory;
(Q) The laboratory shall be equipped with a counter, sink,
and refrigerator;
(R) The procedure room shall be equipped with a ceilingmounted surgical light, operating table or a conventional
gynecological examining table with accessories, closed
cabinets for equipment, and sufficient tables to hold an
emergency tray and other necessary equipment;
(S) There shall be one (1) electrical outlet in the procedure
room for the emergency light and at least one (1) duplex outlet
on each wall;
(T) There shall be one (1) electrical outlet in the recovery
room for the emergency light and at least one (1) duplex outlet
for each two (2) recovery beds or recliners;
(U) Piped-in or portable oxygen and suction equipment shall
be located in the recovery room;
(V) The sterilizing room shall be equipped with a steam
sterilizer, counter and sink, and storage space for clean
supplies. Air pressure in this room shall be positive in relation
to adjacent areas;
(W) The soiled/decontamination room shall be equipped
with a counter and sink. This room shall be equipped with a
constant running exhaust;
(X) A patient toilet with lavatory shall be located convenient
to the recovery room. This room shall be equipped with a
constant running exhaust;
(Y) At least two (2) patient change rooms with secure storage
for personal effects shall be provided; and
(Z) Office space, waiting room, record storage space, and
counseling rooms shall be provided. Counseling rooms shall
be separate and not smaller than ten feet by ten feet (10' × 10').
(4) Any abortion facility in operation at the time these rules are
adopted shall comply with the following:
(A) Smoke detectors shall be located in all rooms and in
corridors at thirty-feet (30') intervals unless the building is
rated Type II (222) fire-resistive or if it is a one- (1-) story building
rated Type II (111) protected-noncombustible as described in
Standard on Types of Building Construction 1979 published by
the NFPA. If the building is multistoried and rated combustible,
it shall be protected throughout by an approved automatic
sprinkler system;
(B) There shall be a system of corridors, passageways, and
elevators adequate in size and arrangement to allow a patient
on a stretcher to be moved from any point in the abortion
facility to a street-level exit;
(C) Space shall be provided for waiting, registration,
counseling, medical evaluation, examination, and referral.
This space shall be equipped with suitable furnishings and
accommodations;
(D) Dressing rooms shall be provided for the privacy, physical
comfort, and convenience of patients and personnel;
(E) At least one (1) procedure room shall be adequately
equipped, supplied, and staffed to safely perform abortions.
The procedure room shall be equipped with an operating
table or a conventional gynecologic examining table with
accessories, a closed cabinet for equipment, and tables to
hold an emergency tray and other necessary equipment. The
procedure room shall be well-lighted and maintained at a
comfortable temperature;
(F) Personnel change rooms and scrub-up facilities shall be
located convenient to the procedure room;
(G) A utility room with facilities for steam sterilization
and space for storage of clean and sterilized supplies shall
be provided. There shall be sufficient surgical instruments
sterilized and available for each patient who presents herself
for an abortion. The room shall be arranged to prevent cross
traffic of clean and dirty material;
(H) The recovery room shall be separate from the procedure
room and be of sufficient size to accommodate at least four
(4) recovery beds or recliners for each procedure room. The
recovery room shall be well-lighted and maintained at a
comfortable temperature. Recovery beds or recliners shall be
spaced to permit easy staff access to each patient;
(I) Piped-in or portable oxygen and suction equipment shall
be located in the recovery room;
(J) Trickle charge battery pack units shall be located to
SENIOR SERVICES
provide emergency lighting in the procedure room, recovery
room, exit corridors, and exit stairs to grade;
(K) A manual fire alarm break station shall be located near
each exit and connected to a local audible alarm which can be
heard throughout the facility;
(L) At least two (2) ABC-type fire extinguishers shall be
located in the facility, one (1) in the clinical area;
(M) Illuminated exit signs shall be located above each exit
door and illuminated directional exit signs shall be located
where needed to direct patients and personnel to exits in event
of an emergency;
(N) Wall and floor finishes in the procedure room, recovery
room, and the sterilization area shall be smooth and easily
cleanable;
(O) The laboratory shall be equipped with a counter, sink,
and refrigerator; and
(P) At least two (2) remote exits shall be provided for each
floor. Each exit shall discharge directly to the outside or
through an enclosed stairway or passageway to the outside.
AUTHORITY: section 197.225, RSMo Supp. 2017.* Original rule filed
July 15, 1987, effective Oct. 25, 1987. Amended: Filed Oct. 24, 2017,
effective April 30, 2018.
*Original authority: 197.225, RSMo 1975, amended 1986, 2017.