20 CSR 2267-5.030
Cleaning and Sterilization
PURPOSE: This rule outlines the proper cleaning and sterilization
of equipment used by a tattooist, body piercer or brander.
(1) Cleaning.
(A) Reusable equipment used in a tattooing, body piercing
and branding procedure shall be cleaned immediately following each use to remove blood and/or body tissue residue prior
to sterilization.
(B) Reusable bars, tubes, branding irons, other branding
instruments and body piercing equipment shall be placed in a
covered container until they are sterilized.
(C) All containers holding contaminated tubes, branding
irons, other branding equipment, reusable body piercing
equipment, and container lids shall be cleaned and disinfected
with an approved disinfectant as defined in 20 CSR 2267-1.010.
(D) Gloved personnel shall clean bars, tubes, branding irons,
other branding equipment, and reusable body piercing equipment prior to sterilization as follows:
1. Manually preclean the items with care taken to ensure
removal of residue; thoroughly rinse the items with warm
water and then drain the water; clean the items by soaking
them in a protein-dissolving deter-gent-enzyme cleaner used
according to the manufacturer’s instructions; and clean the
items further in an ultrasonic cleaning device that operates at
forty to sixty (40–60) hertz and is used according to the manufacturer’s instructions; and
2. Rinsing and drying the items.
(E) Prior to autoclaving, all tubes shall be packaged either
individually or in quantities appropriate for individual procedures. Packages shall be identifiable and dated.
(2) Sterilization.
(A) Equipment requiring sterilization shall be pressure-sterilized at the establishment in an autoclave and in accordance
with manufacturer’s instructions. Practitioners shall have procedures in place to ensure autoclaves have been properly
disinfected and spore-tested as required in subsection (2)(C) of
this rule.
(B) Each batch of sterilized equipment shall be monitored for
sterilization by use of heat-sensitive indicators capable of indicating approximate time and temperature achieved.
(C) Autoclaves shall be spore-tested at least weekly. Spore kill
test effectiveness shall be conducted by an independent laboratory. If a positive spore test is received, the practitioner shall
immediately cease using the autoclave device and notify the
office within forty-eight (48) hours.
(D) Sterilized equipment shall be wrapped and stored in a
manner that ensures it will remain sterile until used.
(E) Each tattoo, body piercing and branding establishment
shall maintain sterilization records including spore tests for at
least two (2) years from the date of the last entry, which shall
include the following information:
1. Date of sterilization;
2. Name of person operating the equipment; and
3. Result of heat-sensitive indicator.
(F) Sterilized equipment shall be resterilized if the package is
opened, damaged or becomes wet.
(G) All methods of sterilization other than steam autoclaving
are prohibited.
AUTHORITY: section 324.522, RSMo Supp. 2007.* This rule originally filed as 4 CSR 267-5.030. Original rule filed Aug. 15, 2002,
effective Feb. 28, 2003. Moved to 20 CSR 2267-5.030, effective Aug.
28, 2006. Amended: Filed June 16, 2008, effective Dec. 30, 2008.
*Original authority: 324.522, RSMo 1998, amended 1999, 2001, 2007.