31 MAC Pt. 506, R. 10.1

Intoxilyzer Calibration Check

Year: 2026Length: 150 wordsOfficial source

Cite as 31 Miss. Admin. Code Pt. 506, R. 10.1

Intoxilyzer Calibration Check Miss. Code Ann. ยง 63-11-19 requires that calibration checks be performed at least once a quarter on breath alcohol instruments. The calibration checks of Intoxilyzer 8000 instruments are performed when an instrument is installed or removed from a location, when a dry gas ethanol standard is changed and with each breath test. The quarterly calibration checks are performed remotely by the Implied Consent Section of the Mississippi Crime Laboratory or designee. These are recorded on Form IP-16E. Certifiers of calibration perform calibration checks on instruments: a) when they are installed, b) removed from a location if possible, and C) when a dry gas ethanol standard is changed. These calibration checks are recorded on Form IP-03E. The calibration checks performed with each breath test are automatically performed by the instrument during the breath test sequence. These calibration check results are recorded on each breath test on Form IP-01E.