31 MAC Pt. 506, R. 8.1
Certified Breath Alcohol Standards
Cite as 31 Miss. Admin. Code Pt. 506, R. 8.1
Certified Breath Alcohol Standards
Breath alcohol reference standard solutions are purchased in the concentrations utilized in the
calibration of the Intoxilyzer 8000. These solutions are certified by the manufacturer to be
consistent of each concentration throughout the lot. The Mississippi Crime Laboratory analyzes
random samples of each lot number of breath alcohol reference standard solutions to ascertain
the concentration value for use in the calibration of the Intoxilyzer 8000. The analysis of the
breath alcohol reference standard solutions is performed using gas chromatography. The result of
analysis is reported on the Implied Consent Certificate of Analysis (Form IP-61) and is on file
with the Implied Consent Section of the Mississippi Crime Laboratory or designee.