15 MAC Pt. 16, R. 1.43.4
Treatment Or Disposal Of Infectious Medical Waste Shall Be by One Of the
Cite as 15 Miss. Admin. Code Pt. 16, R. 1.43.4
Treatment Or Disposal Of Infectious Medical Waste Shall Be by One Of the
Following Methods:
1. By incineration in an approved incinerator which provides combustion of the
waste to carbonized or mineralized ash.
2. By sterilization by heating in a steam sterilizer, so as to render the waste non-
infectious. Infectious medical waste so rendered non-infectious shall be
disposable as medical waste. Operating procedures for steam sterilizers shall
include, but not be limited to the following:
a. Adoption of standard written operating procedures or each steam sterilizer
including time, temperature, pressure, type of waste, type of container(s),
closure on container(s), pattern of loading, water content, and maximum load
quantity;
b. Check or recording and/or indicating thermometers during each complete cycle
to ensure the attainment of a temperature of 121 C (250 degrees F) for one-half
hour or longer, depending on quantity and density of the load, in order to
achieve sterilization of the entire load. Thermometers shall be checked for
calibration at least annually;
c. Use of heat sensitive tape or other device for each container that is processed to
indicate the attainment of adequate sterilization conditions;
d. Use of the biological indicator Bacillus stearothermophilus placed at the center
of a load processed under standard operating conditions at least monthly to
confirm the attainment of adequate sterilization conditions;
e. Maintenance of records of procedures specified in (a), (b), (c) and (d) above for
period of not less than a year;
f. By discharge to the approved sewerage system if the waste is liquid or semi-
liquid, except as prohibited by the Mississippi State Department of Health.
3. Recognizable human anatomical remains shall be disposed of by incineration or
internment, unless burial at an approved landfill is specifically authorized by the
Mississippi State Department of Health.
4. Chemical sterilization shall use only those chemical sterilants recognized by the
U. S. Environmental Protection Agency, Office of Pesticides and Toxic
Substances.
5. Ethylene oxide, glutaraldehyde, and hydrogen peroxide are examples of sterilants
that, used in accordance with manufacturer recommendation, will render
infectious waste non-infectious. Testing with Bacillus Subtilis Spores or other
equivalent organisms shall be conducted quarterly to ensure the sterilization
effectiveness of gas or steam treatment.