15 MAC Pt. 21, R. 1.11.6
Radiation Survey Instruments
Cite as 15 Miss. Admin. Code Pt. 21, R. 1.11.6
Radiation Survey Instruments.
1.
Radiation survey instruments used to determine exposure rates pursuant to this
section shall be capable of measuring 1 microroentgen per hour through at
least 500 microroentgens per hour.
2.
Radiation survey instruments used to make surveys required by this section
shall be calibrated and operable.
3.
Each radiation survey instrument shall be calibrated:
a.
by person licensed by the Agency, another Agreement State, or the U.S.
Nuclear Regulatory Commission to perform such service;
b.
at energies and radiation levels appropriate for the licensee's use;
c.
at intervals not to exceed six months and after each instrument
servicing other than battery replacement; and
d.
to demonstrate an accuracy within plus or minus 20 percent of the true
radiation level on each scale.
4.
Records of these calibrations shall be maintained for 3 years after the calibration
date for inspection by the Agency.