15 MAC Pt. 21, R. 1.8.5
Operating Requirements
Cite as 15 Miss. Admin. Code Pt. 21, R. 1.8.5
Operating Requirements.
1. Procedures. Normal operating procedures shall be written and available to all
analytical x-ray equipment workers. No individual shall be permitted to
operate analytical x-ray equipment in any manner other than that specified in the
procedures unless such individual has obtained written approval of the radiation
safety officer.
2. Bypassing. No individual shall bypass a safety device or interlock unless such
individual has obtained the approval of the radiation safety officer. Such approval
shall be for a specified period of time. When a safety device or interlock
has been bypassed, a readily discernible sign bearing the words "SAFETY
DEVICE NOT WORKING", or words having a similar intent, shall be placed
on the radiation source housing.
3. Repair or Modification of X-Ray Tube Systems. Except as specified in 1.8.5(2),
no operation involving removal of covers, shielding materials or tube housings
or modifications to shutters, collimators, or beam stops shall be performed
without ascertaining that the tube is off and will remain off until safe conditions
have been restored. The main switch, rather than interlocks, shall be used for
routine shutdown in preparation for repairs.
4. Radioactive Source Replacement, Testing, or Repair. Radioactive source
housings shall be opened for source replacement, leak testing, or other maintenance
or repair procedures only by individuals authorized to specifically conduct such
procedures under a license issued by the U.S. Nuclear Regulatory Commission
(NRC), an Agreement State, or a Licensing State.