NDAC 33.1-10-06-11
Computed tomography x-ray systems
Cite as N.D. Admin. Code § 33.1-10-06-11
1.
Definitions. In addition to the definitions provided in sections 33.1-10-01-04 and
33.1-10-06-02, the following definitions are applicable to this section:
a.
"Computed tomography dose index" means the integral from -7T to +7T of the dose
profile along a line perpendicular to the tomographic plane divided by the product of the
nominal tomographic section thickness and the number of tomograms produced in a
single scan, that is:
where:
z = Position along a line perpendicular to the tomographic plane.
D(z) = Dose at position z.
T = Nominal tomographic section thickness.
n = Number of tomograms produced in a single scan.
This definition assumes that the dose profile is centered around z = 0 and that, for a
multiple tomogram system, the scan increment between adjacent scans is nT.
b.
"Contrast scale" means the change in the linear attenuation coefficient per computed
tomography number relative to water, that is:
where:
μx = Linear attenuation coefficient of the material of interest.
μw = Linear attenuation coefficient of water.
(CTN)x = CTN of the material of interest.
(CTN)w = CTN of water.
c.
"CS" (see "contrast scale").
d.
"CT" means a radiologic imaging technique that produces images of "slices" through a
patient's body.
e.
"CT conditions of operation" means all selectable parameters governing the operation of
a CT x-ray system, including nominal tomographic section thickness, filtration, and the
technique factors as defined in section 33.1-10-06-02.
f.
"CTDI" (see "computed tomography dose index").
g.
"CT gantry" means the tube housing assemblies, beam-limiting devices, detectors, and
the supporting structures and frames which hold these components.
h.
"CTN" (see "CT number").
i.
"CT number" means the number used to represent the x-ray attenuation associated with
each elemental area of the CT image.
where:
k = A constant (The constant has a normal value of one thousand when the Houndsfield
scale of CTN is used.)
μx = Linear attenuation coefficient of the material of interest.
μw = Linear attenuation coefficient of water.
j.
"Dose profile" means the dose as a function of position along a line.
k.
"Elemental area" means the smallest area within a tomogram for which the x-ray
attenuation properties of a body are depicted. (see also "picture element").
l.
"Multiple tomogram system" means a computed tomography x-ray system which obtains
x-ray transmission data simultaneously during a single scan to produce more than one
tomogram.
m.
"Noise" means the standard deviation of the fluctuations in computed tomography
number expressed as a percentage of the attenuation coefficient of water. Its estimate
(Sn ) is calculated using the following expression:
where:
CS = Contrast scale.
μw = Linear attenuation coefficient of water.
s = Estimated standard deviation of the CTN of picture elements in a specified area of
the CT image.
n.
"Nominal tomographic section thickness" means the full width at half maximum of the
sensitivity profile taken at the center of the cross sectional volume over which x-ray
transmission data are collected.
o.
"Picture element" means an elemental area of a tomogram.
p.
"Reference plane" means a plane which is displaced from and parallel to the
tomographic plane.
q.
"Scan" means the complete process of collecting x-ray transmission data for the
production of a tomogram. Data can be collected simultaneously during a single scan for
the production of one or more tomograms.
r.
"Scan increment" means the amount of relative displacement of the patient with respect
to the CT x-ray system between successive scans measured along the direction of such
displacement.
s.
"Scan sequence" means a preselected set of two or more scans performed consecutively
under preselected CT conditions of operation.
t.
"Scan time" means the period of time between the beginning and end of x-ray
transmission data accumulation for a single scan.
u.
"Single tomogram system" means a CT x-ray system which obtains x-ray transmission
data during a scan to produce a single tomogram.
v.
"Tomographic plane" means that geometric plane which is identified as corresponding to
the output tomogram.
w.
"Tomographic section" means the volume of an object whose x-ray attenuation properties
are imaged in a tomogram.
2.
Requirements for equipment.
a.
Termination of exposure.
(1)
Means must be provided to terminate the x-ray exposure automatically by either
de-energizing the x-ray source or shuttering the x-ray beam in the event of
equipment failure affecting data collection. Such termination must occur within an
interval that limits the total scan time to no more than one hundred ten percent of its
preset value through the use of either a backup timer or devices which monitor
equipment function.
(2)
A visible signal must indicate when the x-ray exposure has been terminated through
the means required by paragraph 1.
(3)
The operator must be able to terminate the x-ray exposure at any time during a
scan, or series of scans under computed tomography x-ray system control, of
greater than one-half second duration.
b.
Tomographic plane indication and alignment.
(1)
For any single tomogram system, means must be provided to permit visual
determination of the tomographic plane or a reference plane offset from the
tomographic plane.
(2)
For any multiple tomogram system, means shall be provided to permit visual
determination of the location of a reference plane. This reference plane can be
offset from the location of the tomographic planes.
(3)
If a device using a light source is used to satisfy paragraph 1 or 2, the light source
must provide illumination levels sufficient to permit visual determination of the
location of the tomographic plane or reference plane under ambient light conditions
of up to five hundred lux.
c.
Beam-on and shutter status indicators and control switches.
(1)
The computed tomography x-ray control and gantry must provide visual indication
whenever x-rays are produced and, if applicable, whether the shutter is open or
closed.
(2)
Each emergency button or switch must be clearly labeled as to its function.
d.
Indication of computed tomography conditions of operation. The computed tomography
x-ray system must be designed such that the computed tomography conditions of
operation to be used during a scan or a scan sequence must be indicated prior to the
initiation of a scan or a scan sequence. On equipment having all or some of these
conditions of operation at fixed values, this requirement may be met by permanent
markings. Indication of computed tomography conditions of operation must be visible
from any position from which scan initiation is possible.
e.
Extraneous radiation. When data are not being collected for image production, the
radiation adjacent to the tube port may not exceed that permitted by subsection 3 of
section 33.1-10-06-04.
f.
Maximum surface computed tomography dose index identification. The angular position
where the maximum surface computed tomography dose index occurs must be identified
to allow for reproducible positioning of a computed tomography dosimetry phantom.
g.
Additional requirements applicable to computed tomography x-ray systems containing a
gantry manufactured after September 3, 1985.
(1)
The total error in the indicated location of the tomographic plane or reference plane
may not exceed five millimeters.
(2)
If the x-ray production period is less than one-half second, the indication of x-ray
production must be actuated for at least one-half second. Indicators at or near the
gantry must be discernible from any point external to the patient opening where
insertion of any part of the human body into the primary beam is possible.
(3)
The deviation of indicated scan increment versus actual increment may not exceed
plus or minus one millimeter with any mass from zero to one hundred kilograms
resting on the support device. The patient support device must be incremented from
a typical starting position to the maximum incremented distance or thirty
centimeters, whichever is less, and then returned to the starting position.
Measurement of actual versus indicated scan increment may be taken anywhere
along this travel.
(4)
Premature termination of the x-ray exposure by the operator shall necessitate
resetting of the computed tomography conditions of operation prior to the initiation
of another scan.
h.
Facility design requirements.
(1)
All CT capable systems shall be required to have the x-ray control permanently
mounted in a protected area during the entire exposure (see appendix B).
(2)
Aural communication. Provision must be made for two-way aural communication
between the patient and the operator at the control panel.
i.
Surveys, calibrations, spot checks, and operating procedures.
(1)
Surveys.
(a)
All computed tomography x-ray systems installed after March 1, 1992, and
those systems not previously surveyed shall have a survey made by, or under
the direction of, a qualified expert. In addition, such surveys must be done after
any change in the facility or equipment which might cause a significant
increase in radiation hazard.
(b)
The registrant shall obtain a written report of the survey from the qualified
expert, and a copy of the report must be made available to the department
upon request.
(2)
Radiation calibrations.
(a)
The calibration of the radiation output of the computed tomography x-ray
system must be performed by, or under the direction of, a qualified expert who
is physically present at the facility during such calibration.
(b)
The calibration of a computed tomography x-ray system must be performed at
intervals specified by a qualified expert and after any change or replacement of
components which, in the opinion of the qualified expert, could cause a change
in the radiation output.
(c)
The calibration of the radiation output of a computed tomography x-ray system
must be performed with a calibrated dosimetry system. The calibration of such
system must be traceable to a national standard. The dosimetry system must
have been calibrated within the preceding two years.
(d)
Computed tomography dosimetry phantoms must be used in determining the
radiation output of a computed tomography x-ray system. Such phantoms must
meet the following specifications and conditions of use:
[1]
Computed tomography dosimetry phantoms must be right circular
cylinders of polymethyl methacrylate of density one point nineteen plus or
minus point zero one grams per cubic centimeter. The phantoms must be
at least fourteen centimeters in length and must have diameters of
thirty-two centimeters for testing computed tomography x-ray systems
designed to image any section of the body and sixteen centimeters for
systems designed to image the head or for whole-body scanners
operated in the head scanning mode.
[2]
Computed tomography dosimetry phantoms must provide means for the
placement of a dosimeter along the axis of rotation and along a line
parallel to the axis of rotation one centimeter from the outer surface and
within the phantom. Means for the placement of dosimeters or alignment
devices at other locations may be provided.
[3]
Any effects on the doses measured due to the removal of phantom
material to accommodate dosimeters must be accounted for through
appropriate corrections to the reported data or included in the statement
of maximum deviation for the values obtained using the phantom.
[4]
All dose measurements must be performed with the computed
tomography dosimetry phantom placed on the patient couch or support
device without additional attenuation materials present.
(e)
The calibration shall be required for each type of head, body, or whole-body
scan performed at the facility.
(f)
Calibration must meet the following requirements:
[1]
The dose profile along the center axis of the computed tomography
dosimetry phantom for the minimum, maximum, and midrange values of
the nominal tomographic section thickness used by the registrant shall be
measurable. Where less than three nominal tomographic thicknesses can
be selected, the dose profile determination must be performed for each
available nominal tomographic section thickness.
[2]
The computed tomography dose index (For the purpose of determining
the computed tomography dose index, the manufacturer's statement as to
the nominal tomographic section thickness for that particular system may
be utilized.) along the two axes specified in item 2 of subparagraph d
must be measured. The computed tomography dosimetry phantom must
be oriented so that the measurement point one centimeter from the outer
surface and within the phantom is in the same angular position within the
gantry as the point of maximum surface computed tomography dose
index identified. The computed tomography conditions of operation must
correspond to typical values used by the registrant.
[3]
The spot checks specified in paragraph 3 of subdivision i must be made.
(g)
Calibration procedures must be in writing. Records of calibrations performed
must be maintained for inspection by the department.
(3)
Spot checks.
(a)
The spot check procedures must be in writing and must have been developed
by a qualified expert.
(b)
The spot check procedures must incorporate the use of a computed
tomography dosimetry phantom which has a capability of providing an
indication of contrast scale, noise, nominal tomographic section thickness, the
resolution capability of the system for low and high contrast objects, and
measuring the mean computed tomography number for water or other
reference material.
(c)
All spot checks must be included in the calibration required by paragraph 2 and
at time intervals and under system conditions specified by a qualified expert.
(d)
Spot checks must include acquisition of images obtained with the computed
tomography dosimetry phantoms using the same processing mode and
computed tomography conditions of operation as are used to perform
calibrations required by paragraph 2 of subdivision i. The images must be
retained, until a new calibration is performed, in two forms as follows:
[1]
Photographic copies of the images obtained from the image display
device; and
[2]
Images stored in digital form on a storage medium compatible with the
computed tomography x-ray system.
(e)
Written records of the spot checks performed shall be maintained for
inspection by the department.
(4)
Operating procedures.
(a)
The computed tomography x-ray system may not be operated except by an
individual who has been specifically trained in its operation.
(b)
Information must be available at the control panel regarding the operation and
calibration of the system. Such information must include the following:
[1]
Dates of the latest calibration and spot checks and the location within the
facility where the results of those tests may be obtained;
[2]
Instructions on the use of the computed tomography dosimetry phantoms,
including a schedule of spot checks appropriate for the system, allowable
variations for the indicated parameters, and the results of at least the
most recent spot checks conducted on the system;
[3]
The distance in millimeters between the tomographic plane and the
reference plane if a reference plane is utilized; and
[4]
A current technique chart available at the control panel which specifies for
each routine examination the computed tomography conditions of
operation and the number of scans per examination.
(c)
If the calibration or spot check of the computed tomography x-ray system
identifies that a system operating parameter has exceeded a tolerance
established by the qualified expert, use of the computed tomography x-ray
system on patients must be limited to those uses permitted by established
written instructions of the qualified expert.