NDAC 33.1-10-15-04
General technical requirements
Cite as N.D. Admin. Code ยง 33.1-10-15-04
1.
Protection surveys.
a.
The registrant shall ensure that radiation protection surveys of all new facilities, and
existing facilities not previously surveyed, are performed with an operable radiation
measurement survey instrument calibrated in accordance with section 33.1-10-15-08.
The radiation protection survey shall be performed by, or under the direction of, a
radiation therapy physicist or a qualified expert and shall verify that, with the therapeutic
radiation machine in a "BEAM-ON" condition, with the largest clinically available
treatment field and with a scattering phantom in the useful beam of radiation:
(1)
Radiation levels in restricted areas are not likely to cause personnel exposures in
excess of the limits specified in chapter 33.1-10-04.2-01 [10 CFR 20.1201]; and
(2)
Radiation levels in unrestricted areas do not exceed the limits specified in chapter
33.1-10-04.2-01 [10 CFR 20.1301].
b.
In addition to the requirements of subdivision a, a radiation protection survey shall also
be performed prior to any subsequent medical use and:
(1)
After making any change in the treatment room shielding;
(2)
After making any change in the location of the therapeutic radiation machine within
the treatment room;
(3)
After relocating the therapeutic radiation machine; or
(4)
Before using the therapeutic radiation machine in a manner that could result in
increased radiation levels in areas outside the external beam radiation therapy
treatment room.
c.
The survey record shall indicate all instances where the facility, in the opinion of the
radiation therapy physicist or a qualified expert, is in violation of applicable regulations.
The survey record shall also include the date of the measurements, the reason the
survey is required, the manufacturer's name, model number and serial number of the
therapeutic radiation machine, the instruments used to measure radiation levels, a plan
of the areas surrounding the treatment room that were surveyed, the measured dose rate
at several points in each area expressed in microsieverts or millirems per hour, the
calculated maximum level of radiation over a period of one week for each restricted and
unrestricted area, and the signature of the individual responsible for conducting the
survey;
d.
If the results of the surveys required by subdivision a or b indicate any radiation levels in
excess of the respective limit specified in subdivision a, the registrant shall lock the
control in the "OFF" position and not use the unit:
(1)
Except as may be necessary to repair, replace, or test the therapeutic radiation
machine, the therapeutic radiation machine shielding, or the treatment room
shielding; or
(2)
Until the registrant has received a specific exemption from the department.
2.
Modification of radiation therapy unit or room before beginning a treatment program. If the
survey required by subsection 1 indicates that an individual in an unrestricted area may be
exposed to levels of radiation greater than those permitted by chapter 33.1-10-04.2-01
[10 CFR 20.1301], before beginning the treatment program the registrant shall:
a.
Either equip the unit with beam direction interlocks or add additional radiation shielding to
ensure compliance with chapter 33.1-10-04.2-01 [10 CFR 20.1301];
b.
Perform the survey required by subsection 1 again; and
c.
Include in the report required by subsection 4 the results of the initial survey, a
description of the modification made to comply with subdivision a, and the results of the
second survey; or
d.
Request and receive a registration amendment under chapter 33.1-10-04.2-01 [10 CFR
20.1301] that authorizes radiation levels in unrestricted areas greater than those
permitted by chapter 33.1-10-04.2-01 [10 CFR 20.1301(a) and 10 CFR 20.1301(b)].
3.
Dosimetry equipment.
a.
The registrant shall have a calibrated dosimetry system available for use. The system
shall have been calibrated by the national institute for standards and technology (NIST)
or by an American association of physicists in medicine (AAPM) accredited dosimetry
calibration laboratory (ADCL). The calibration shall have been performed within the
previous twenty-four months and after any servicing that may have affected system
calibration. An independent survey shall be conducted by a qualified expert or radiation
therapy physicist other than the person performing the original survey prior to the
equipment being used except as described in subdivision d of subsection 1.
(1)
For beams with energies greater than one million volts (1 Mv) or one million electron
volts (1 MeV), the dosimetry system shall have been calibrated for cobalt-60; or
(2)
For beams with energies equal to or less than one million volts (1 Mv) or one million
electron volts (1 MeV), the dosimetry system shall have been calibrated at an
energy appropriate for the radiation being measured;
b.
The registrant shall have available for use a dosimetry system for quality assurance
check measurements. To meet this requirement, the system may be compared with a
system that has been calibrated in accordance with subdivision a. This comparison shall
have been performed within the previous twelve months and after each servicing that
may have affected system calibration. The quality assurance check system may be the
same system used to meet the requirement in subdivision a; and
c.
The registrant shall maintain a record of each dosimetry system calibration,
intercomparison, and comparison for the duration of the registration. For each calibration,
intercomparison, or comparison, the record shall include the date; the model numbers
and serial numbers of the instruments that were calibrated, intercompared, or compared
as required by subdivisions a and b; the correction factors that were determined; the
names of the individuals who performed the calibration, intercomparison, or comparison;
and evidence that the intercomparison was performed by, or under the direct supervision
and in the physical presence of, a radiation therapy physicist.
4.
Reports of external beam radiation therapy surveys and measurements. The registrant for any
therapeutic radiation machine subject to section 33.1-10-15-06 or 33.1-10-15-07 shall furnish
a copy of the records required in subsections 1 and 2 to the department within thirty days
following completion of the action that initiated the record requirement.