NDAC 33.1-10-15-06
Therapeutic radiation machines of less than five hundred kilovolts
Cite as N.D. Admin. Code ยง 33.1-10-15-06
1.
Leakage radiation. When the x-ray tube is operated at its maximum rated tube current for the
maximum kilovolts, the leakage air kerma rate shall not exceed the value specified at the
distance specified for that classification of therapeutic radiation machine:
a.
Five to fifty kilovolts systems. The leakage air kerma rate measured at any position
five centimeters from the tube housing assembly shall not exceed one milligray (100
mrad) in any one hour.
b.
Greater than fifty and less than five hundred kilovolts systems. The leakage air kerma
rate measured at a distance of one meter from the target in any direction shall not
exceed one centigray (1 rad) in any one hour. This air kerma rate measurement may be
averaged over areas no larger than one hundred square centimeters. In addition, the air
kerma rate at a distance of five centimeters from the surface of the tube housing
assembly shall not exceed thirty centagray (30 rad) per hour.
c.
For each therapeutic radiation machine, the registrant shall determine, or obtain from the
manufacturer, the leakage radiation existing at the positions specified in subdivisions a
and b for the specified operating conditions. Records on leakage radiation
measurements shall be maintained at the installation for inspection by the department.
2.
Permanent beam-limiting devices. Permanent diaphragms or cones used for limiting the useful
beam shall provide at least the same degree of attenuation as required for the tube housing
assembly.
3.
Adjustable or removable beam-limiting devices.
a.
All adjustable or removable beam-limiting devices, diaphragms, cones, or blocks shall
not transmit more than five percent of the useful beam for the most penetrating beam
used; and
b.
When adjustable beam-limiting devices are used, the position and shape of the radiation
field shall be indicated by a light beam.
4.
Filter system. The filter system shall be so designed that:
a.
Filters cannot be accidentally displaced at any possible tube orientation;
b.
An interlock system prevents irradiation if the proper filter is not in place;
c.
The air kerma rate escaping from the filter slot shall not exceed one centigray (1 rad) per
hour at one meter under any operating conditions; and
d.
Each filter shall be marked as to its material of construction and its thickness. For wedge
filters, the nominal wedge angle shall appear on the wedge or wedge tray, if permanently
mounted to the tray.
5.
Tube immobilization.
a.
The x-ray tube shall be so mounted that it cannot accidentally turn or slide with respect to
the housing aperture; and
b.
The tube housing assembly shall be capable of being immobilized for stationary portal
treatments.
6.
Source marking. The tube housing assembly shall be so marked that it is possible to
determine the location of the source to within five millimeters, and such marking shall be
readily accessible for use during calibration procedures.
7.
Beam block. Contact therapy tube housing assemblies shall have a removable shield of
material, equivalent in attenuation to five-tenths millimeters of lead at one hundred kilovolts,
which can be positioned over the entire useful beam exit port during periods when the beam is
not in use.
8.
Timer. A suitable irradiation control device shall be provided to terminate the irradiation after a
preset time interval.
a.
A timer with a display shall be provided at the treatment control panel. The timer shall
have a preset time selector and an elapsed time or time remaining indicator;
b.
The timer shall be a cumulative timer that activates with an indication of "BEAM-ON" and
retains its reading after irradiation is interrupted or terminated. After irradiation is
terminated and before irradiation can be reinitiated, it shall be necessary to reset the
elapsed time indicator;
c.
The timer shall terminate irradiation when a preselected time has elapsed, if any
dose-monitoring system present has not previously terminated irradiation;
d.
The timer shall permit accurate presetting and determination of exposure times as short
as one second;
e.
The timer shall not permit an exposure if set at zero;
f.
The timer shall not activate until the shutter is opened when irradiation is controlled by a
shutter mechanism unless calibration includes a timer error correction to compensate for
mechanical lag; and
g.
Timer shall be accurate to within one percent of the selected value or one second,
whichever is greater.
9.
Control panel functions. The control panel, in addition to the displays required by other
provisions in this section, shall have:
a.
An indication of whether electrical power is available at the control panel and if activation
of the x-ray tube is possible;
b.
An indication of whether x-rays are being produced;
c.
A means for indicating x-ray tube potential and current;
d.
The means for terminating an exposure at any time;
e.
A locking device which will prevent unauthorized use of the therapeutic radiation
machine; and
f.
A positive display of specific filters in the beam.
10.
Multiple tubes. When a control panel may energize more than one x-ray tube:
a.
It shall be possible to activate only one x-ray tube at any time;
b.
There shall be an indication at the control panel identifying which x-ray tube is activated;
and
c.
There shall be an indication at the tube housing assembly when that tube is energized.
11.
Target-to-skin distance (TSD). There shall be a means of determining the central axis
target-to-skin distance to within one centimeter and of reproducing this measurement to within
two millimeters thereafter.
12.
Shutters. Unless it is possible to bring the x-ray output to the prescribed exposure parameters
within five seconds after the x-ray "ON" switch is energized, the beam shall be attenuated by a
shutter having a lead equivalency not less than that of the tube housing assembly. In addition,
after the unit is at operating parameters, the shutter shall be controlled by the operator from
the control panel. An indication of shutter position shall appear at the control panel.
13.
Low filtration x-ray tubes. Each therapeutic radiation machine equipped with a beryllium or
other low-filtration window shall be clearly labeled as such upon the tube housing assembly
and shall be provided with a permanent warning device on the control panel that is activated
when no additional filtration is present, to indicate that the dose rate is very high.
14.
Facility design requirements for therapeutic radiation machines capable of operating in the
range fifty kilovolts to five hundred kilovolts. In addition to shielding adequate to meet
requirements of section 33.1-10-15-09, the treatment room shall meet the following design
requirements:
a.
Aural communication. Provision shall be made for continuous two-way aural
communication between the patient and the operator at the control panel.
b.
Viewing systems. Provision shall be made to permit continuous observation of the patient
during irradiation and the viewing system shall be so located that the operator can
observe the patient from the control panel. The therapeutic radiation machine shall not
be used for patient irradiation unless at least one viewing system is operational.
15.
Additional requirements. Treatment rooms that contain a therapeutic radiation machine
capable of operating above one hundred fifty kilovolts shall meet the following additional
requirements:
a.
All protective barriers shall be fixed except for entrance doors or beam interceptors;
b.
The control panel shall be located outside the treatment room or in a totally enclosed
booth, which has a ceiling, inside the room;
c.
Interlocks shall be provided such that all entrance doors, including doors to any interior
booths, shall be closed before treatment can be initiated or continued. If the radiation
beam is interrupted by any door opening, it shall not be possible to restore the machine
to operation without closing the door and reinitiating irradiation by manual action at the
control panel; and
d.
When any door referred to in subdivision c is opened while the x-ray tube is activated,
the air kerma rate at a distance of one meter from the source shall be reduced to less
than one milligray (100 mrad) per hour.
16.
Full calibration measurements.
a.
Full calibration of a therapeutic radiation machine subject to this section shall be
performed by, or under the direct supervision of, a radiation therapy physicist:
(1)
Before the first medical use following installation or reinstallation of the therapeutic
radiation machine;
(2)
At intervals not exceeding one year; and
(3)
Before medical use under the following conditions:
(a)
Whenever quality assurance check measurements indicate that the radiation
output differs by more than five percent from the value obtained at the last full
calibration and the difference cannot be reconciled; and
(b)
Following any component replacement, major repair, or modification of
components that could significantly affect the characteristics of the radiation
beam.
(4)
Notwithstanding the requirements of paragraph 3:
(a)
Full calibration of therapeutic radiation machines with multi-energy capabilities
is required only for those modes or energies, or both, that are not within their
acceptable range; and
(b)
If the repair, replacement, or modification does not affect all energies, full
calibration shall be performed on the affected energy that is in most frequent
clinical use at the facility. The remaining energies may be validated with quality
assurance check procedures against the criteria in subparagraph a of
paragraph 3.
b.
To satisfy the requirement of subdivision a, full calibration shall include all measurements
recommended for annual calibration by NCRP report 69, "dosimetry of x-ray and gamma
ray beams for radiation therapy in the energy range ten keV to fifty MeV" (1981).
c.
The registrant shall maintain a record of each calibration for the duration of the
registration. The record shall include the date of the calibration; the manufacturer's name,
model number, and serial number for both the therapeutic radiation machine and the
x-ray tube; the model numbers and serial numbers of the instruments used to calibrate
the therapeutic radiation machine; and the signature of the radiation therapy physicist
responsible for performing the calibration.
17.
Periodic quality assurance checks.
a.
Periodic quality assurance checks shall be performed on therapeutic radiation machines
subject to this section, which are capable of operation at greater than or equal to fifty
kilovolts.
b.
To satisfy the requirement of subdivision a, quality assurance checks shall meet the
following requirements:
(1)
The registrant shall perform quality assurance checks in accordance with written
procedures established by the radiation therapy physicist; and
(2)
The quality assurance check procedures shall specify the frequency at which tests
or measurements are to be performed. The quality assurance check procedures
shall specify that the quality assurance check shall be performed during the
calibration specified in subdivision a of subsection 16. The acceptable tolerance for
each parameter measured in the quality assurance check, when compared to the
value for that parameter determined in the calibration specified in subdivision a of
subsection 16, shall be stated.
c.
The cause for a parameter exceeding a tolerance set by the radiation therapy physicist
shall be investigated and corrected before the system is used for patient irradiation.
d.
Whenever a quality assurance check indicates a significant change in the operating
characteristics of a system, as specified in the radiation therapy physicist's quality
assurance check procedures, the system shall be recalibrated as required in
subdivision a of subsection 16.
e.
The registrant shall use the dosimetry system described in subdivision b of subsection 3
of section 33.1-10-15-04 to make the quality assurance check required in subdivision b.
f.
The registrant shall have the radiation therapy physicist review and sign the results of
each radiation output quality assurance check within one month of the date that the
check was performed.
g.
The registrant shall ensure that safety quality assurance checks of therapeutic radiation
machines subject to section 33.1-10-15-06 are performed at intervals not to exceed one
month.
h.
Notwithstanding the requirements of subdivisions f and g, the registrant shall ensure that
no therapeutic radiation machine is used to administer radiation to humans unless the
quality assurance checks required by subdivisions f and g have been performed within
the thirty-day period immediately prior to said administration.
i.
To satisfy the requirement of subdivision g, safety quality assurance checks shall ensure
proper operation of:
(1)
Electrical interlocks at each external beam radiation therapy room entrance;
(2)
The "BEAM-ON" and termination switches;
(3)
Beam condition indicator lights on the access doors, control console, and in the
radiation therapy room;
(4)
Viewing systems; and
(5)
If applicable, electrically operated treatment room doors from inside and outside the
treatment room.
j.
The registrant shall maintain a record of each quality assurance check required by
subdivision g for three years. The record shall include the date of the quality assurance
check; the manufacturer's name, model number, and serial number of the therapeutic
radiation machine; the manufacturer's name, model number, and serial number for the
instruments used to measure the radiation output of the therapeutic radiation machine;
and the signature of the individual who performed the periodic quality assurance check.
18.
Operating procedures.
a.
The therapeutic radiation machine shall not be used for irradiation of patients unless the
requirements of subsection 16 of section 33.1-10-15-06 and subsection 17 of this section
have been met;
b.
Therapeutic radiation machines shall not be left unattended unless secured pursuant to
subdivision e of subsection 9;
c.
When a patient must be held in position for radiation therapy, mechanical supporting or
restraining devices shall be used;
d.
The tube housing assembly shall not be held by an individual during operation unless the
assembly is designed to require such holding and the peak tube potential of the system
does not exceed fifty kilovolts. In such cases, the holder shall wear protective gloves and
apron of not less than five-tenths millimeters lead equivalency at one hundred kilovolts;
e.
A copy of the current operating and emergency procedures shall be maintained at the
therapeutic radiation machine control console; and
f.
No individual other than the patient shall be in the treatment room during exposures from
therapeutic radiation machines operating above one hundred fifty kilovolts. At energies
less than or equal to one hundred fifty kilovolts, any individual, other than the patient, in
the treatment room shall be protected by a barrier sufficient to meet the requirements of
chapter 33.1-10-04.2-01 [10 CFR 20.1201].
19.
Possession of survey instruments. Each facility location authorized to use a therapeutic
radiation machine in accordance with this section shall possess appropriately calibrated
portable monitoring equipment. As a minimum, such equipment shall include a portable
radiation measurement survey instrument capable of measuring dose rates over the range ten
microsievert (1 mrem) per hour to ten millisievert (1,000 mrem) per hour. The survey
instruments shall be operable and calibrated in accordance with section 33.1-10-15-08.