NDAC 33.1-10-15-08
Calibration of survey instruments
Cite as N.D. Admin. Code ยง 33.1-10-15-08
1.
The registrant shall ensure that the survey instruments used to show compliance with this
chapter have been calibrated before first use, at intervals not to exceed twelve months, and
following repair;
2.
To satisfy the requirements of subsection 1, the registrant shall:
a.
Calibrate all required scale readings up to ten millisieverts (1,000 mrem) per hour with an
appropriate radiation source that is traceable to the national institute of standards and
technology (NIST); and
b.
Calibrate at least two points on each scale to be calibrated. These points should be at
approximately one-third and two-thirds of full scale;
3.
To satisfy the requirements of subsection 2, the registrant shall:
a.
Consider a point as calibrated if the indicated dose rate differs from the calculated dose
rate by not more than ten percent; and
b.
Consider a point as calibrated if the indicated dose rate differs from the calculated dose
rate by not more than twenty percent if a correction factor or graph is conspicuously
attached to the instrument;
4.
The registrant shall retain a record of each calibration required in subsection 1 for three years.
The record shall include:
a.
A description of the calibration procedure; and
b.
A description of the source used and the certified dose rates from the source, and the
rates indicated by the instrument being calibrated, the correction factors deduced from
the calibration data, the signature of the individual who performed the calibration, and the
date of calibration; and
5.
The registrant may obtain the services of individuals registered by the department, or licensed
by the United States nuclear regulatory commission, an agreement state, or a licensing state
to perform calibrations of survey instruments. Records of calibrations that contain information
required by subsection 4 shall be maintained by the registrant.