180 NAC 7-029. POSSESSION, USE, AND CALIBRATION OF INSTRUMENTS USED TO MEASURE THE ACTIVITY OF UNSEALED RADIOACTIVE MATERIAL
Cite as Neb. Admin. Code tit. 180, ch. 7, § 029
029. POSSESSION, USE, AND CALIBRATION OF INSTRUMENTS USED TO MEASURE THE ACTIVITY OF UNSEALED RADIOACTIVE MATERIAL. To perform direct measurements according to 180 NAC 7-031, a licensee must:
(A) Possess and use instrumentation to measure the activity of unsealed radioactive material before administration to each patient or human research subject;(B) Calibrate the instrumentation according to nationally recognized standards or the manufacturer’s instructions. The calibrations must, at minimum and as appropriate to demonstrate proper operation of the instrument, test for:
(i) Constancy;(ii) Linearity;(iii) Accuracy; and(iv) Geometry dependence; and
(C) Retain a record of each instrument test required by 180 NAC 7-029 according to 180 NAC 7-091.