180 NAC 7-061

180 NAC 7-061. THERAPY-RELATED COMPUTER SYSTEMS

Last amended: 2023Year: 2026Length: 87 wordsOfficial source

Cite as Neb. Admin. Code tit. 180, ch. 7, § 061

061. THERAPY-RELATED COMPUTER SYSTEMS. The licensee must perform acceptance testing on the treatment planning system of therapy-related computer systems according to published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of: (A) The source-specific input parameters required by the dose calculation algorithm;(B) The accuracy of dose, dwell time, and treatment time calculations at representative points;(C) The accuracy of isodose plots and graphic displays; and(D) The accuracy of the software used to determine sealed source positions from radiographic images.