180 NAC 7-104. RECORDS OF CALIBRATION MEASUREMENTS OF BRACHYTHERAPY SOURCES. A licensee must maintain a record of the calibrations of brachytherapy sources required by 180 NAC 7-060 for three years after the last use of the source

Last amended: 2023Year: 2026Length: 84 wordsOfficial source

Cite as Neb. Admin. Code tit. 180, ch. 7, § 104

104. RECORDS OF CALIBRATION MEASUREMENTS OF BRACHYTHERAPY SOURCES. A licensee must maintain a record of the calibrations of brachytherapy sources required by 180 NAC 7-060 for three years after the last use of the source. The record must include: (A) The date of the calibration;(B) The manufacturer's name, model number, and serial number for the source and the instruments used to calibrate the source;(C) The source output or activity;(D) The source positioning accuracy within the applicators; and(E) The signature of the authorized medical physicist.