Rule 3701:1-58-70. Therapy-related computer systems for remote afterloader units, teletherapy units, and gamma stereotactic radiosurgery units
119.03 The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of: (A) The source-specific input parameters required by the dose calculation algorithm; (B) The accuracy of dose, dwell time, and treatment time calculations at representative points; (C) The accuracy of isodose plots and graphic displays; (D) The accuracy of the software used to determine sealed source positions from radiographic images; and (E) The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system. Last updated November 18, 2022 at 8:39 AM Supplemental Information Authorized By: 3748.02 , 3748.04 Amplifies: 3748.04 4/7/2027