Ohio Admin. Code 3701:1-58-96

Rule 3701:1-58-96. Records of periodic spot-checks for remote afterloader units

Last amended: 2005Year: 2026Length: 168 wordsOfficial source
119.03 (A) A licensee shall retain a record of each spot-check for remote afterloader units required by rule 3701:1-58-65 of the Administrative Code for three years. (B) The record must include, as applicable: (1) The date of the spot-check; (2) The manufacturer's name, model number, and serial number for the remote afterloader unit and source; (3) An assessment of timer accuracy; (4) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit's computer; and (5) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check. (C) A licensee shall retain a copy of the procedures required by paragraph (B) of rule 3701:1-58-65 of the Administrative Code until the licensee no longer possesses the remote afterloader unit. Last updated November 18, 2022 at 8:42 AM Supplemental Information Authorized By: 3748.02 , 3748.04 Amplifies: 3748.04 4/7/2027