OAC 310:667-23-2
Radiological, computerized tomography, magnetic resonance
Cite as Okla. Admin. Code § 310:667-23-2
imaging services
(a)
Radiology.
The
hospital
shall
maintain
or
have
available
radiological services according to needs of the hospital.
(1)
Hazards for patients and personnel.
(A)
The radiological department shall be free of health and
safety hazards for patients and personnel.
(B)
Proper safety precautions shall be maintained against fire
and explosion hazards, electrical hazards, and radiation hazards.
(C)
Inspection of x-ray equipment shall be made once every two
(2) years by a certified health physicist or members of the
Diagnostic x-ray section of the Department, and hazards so
identified shall be promptly corrected.
(D)
The hospital shall identify those employees who are subject
to significant occupational exposure to radiation while performing
their job duties. All such workers shall be checked periodically
for amounts of radiation exposure by the use of exposure meters or
badge tests.
(E)
With fluoroscopes, attention shall be paid to modern safety
design and good operating procedures; records shall be maintained
for the output of all fluoroscopes.
(F)
Regulations based upon medical staff recommendations shall be
established as to the administration of the application and
removal of radium element, its disintegration products, and other
radioactive isotopes.
(G)
If mammography is performed at the facility, the facility
shall have a current certificate from the Food and Drug
Administration as required by the Mammography Quality Standards
Act.
(2)
Personnel.
(A)
Personnel adequate to supervise and conduct the services
shall
be
provided,
and
the
interpretation
of
radiological
examinations shall be made by physicians or licensed independent
practitioners competent in the field according to individually
granted clinical privileges.
(B)
The hospital shall have a qualified radiologist, either full-
time, part-time or on a consulting basis, both to supervise the
department and to interpret films and studies that require
specialized knowledge for accurate reading.
(C)
If the activities of the radiology department extend to
radiotherapy, the physician in charge shall be appropriately
qualified.
(D)
The amount of qualified radiologist and technologist time
shall be sufficient to meet the hospital's requirements. A
technologist shall be on duty or on call at all times.
(E)
The use of all x-ray apparatus shall be limited to personnel
OAC 310:667
OKLAHOMA STATE DEPARTMENT OF HEALTH
54
September 13, 2019
designated as qualified by the radiologist or by an appropriately
constituted committee of the medical staff. The same limitation
shall apply to personnel applying and removing radium element, its
disintegration products, and radioactive isotopes. Radiology
technologists
shall
not
independently
perform
fluoroscopic
procedures. Fluoroscopic procedures may be performed by radiology
technologists only upon the written authorization of a qualified
radiologist, and in the presence of a physician or licensed
independent practitioner or by real time visualization through
electronic means.
(3)
Signed reports.
(A)
Signed reports shall be filed with the patient's medical
record and exact duplicates of the signed reports shall be kept in
the department.
(B)
Requests by the attending physician or licensed independent
practitioner for x-ray examination shall contain a concise
statement of reason for the examination.
(C)
Reports of interpretations shall be written or dictated and
signed by the radiologist, physician, or licensed independent
practitioner making the interpretation.
(D)
X-ray reports shall be preserved or microfilmed in accordance
with the statute of limitations and OAC 310:667-19.
(E)
X-rays and other images shall be maintained at least five (5)
years. These images may be maintained in a digital or electronic
format as long as a duplicate can be reproduced.
(b)
Ultrasound imaging.
(1)
Ultrasound imaging shall be performed only upon order of a
physician or licensed independent practitioner.
(2)
Ultrasound imaging shall be performed by a physician or licensed
independent practitioner or by a technologist that has specific
training in ultrasound imaging and designated as qualified by the
radiologist.
(3)
Reports of findings of ultrasound imaging shall be included in
the patient's medical record.
(c)
Computerized tomography and magnetic resonance imaging.
(1)
Computerized tomography and magnetic resonance imaging may be
provided.
(2)
If used by the facility, all computerized tomography (CT) and
magnetic resonance imaging (MRI) examinations shall be authorized by
a written and signed order from a physician or licensed independent
practitioner.
(3)
CT and MRI examinations shall be performed under the direction
of and interpreted by a qualified radiologist who is a member of the
hospital active or consulting medical staff.
(4)
CT and MRI examinations shall be performed by a radiology
technologist with documented CT or MRI training and experience and
designated as qualified by the radiologist.
(5)
A qualified physician or licensed independent practitioner shall
be available during the administration of intravenous contrast media.
(6)
Oxygen and emergency medical supplies shall be maintained within
the CT/MRI suite or be readily available. If the CT/MRI suite is a
mobile unit, the mobile unit shall contain oxygen and emergency
medical supplies.
OAC 310:667
OKLAHOMA STATE DEPARTMENT OF HEALTH
55
September 13, 2019
(7)
CT/MRI films and reports shall be maintained at the hospital in
the same manner as x-ray films and reports.
(8)
If the CT/MRI is a mobile unit, written infection control policy
and procedures and safety plans shall be maintained as part of the
overall hospital plans.