Regl. 5293, art. 110-16-660
X-RAY EQUIPMENT
Cite as Reglamento Núm. 5293, Art. 110-16-660
70-675
conductor, or shall be permitted as a separate conductor and shall be permitted to be in contact with the cable.
(d) Cable Covering. Each cable shall be provided with an outer covering, either overall or on each of any subassemblies of grouped conductors.
Tape shall be permitted in place of a covering. Where not installed in metal
raceway, the covering shall be flame-retardant or the cable or each cable
subassembly shall be covered with a closely wound fireproof tape.
650-6. Installation of Conductors. Cables shall be securely fastened in
place and shall be permitted to be attached directly to the organ structure
without insulating supports. Cables shall not be placed in contact with
other conductors.
650-7. Overcurrent Protection. Circuits shall be so arranged that all conductors shall be protected from overcurrent by an overcurrent device rated
at not more than 6 amperes.
Exception: The main supply conductors and the common-return conductors.
ARTICLE 660 - X-RAY EQUIPMENT
A. General
660-1. Scope. This article covers all X-ray equipment operating at any
frequency or voltage for industrial or other nonmedical or nondental use.
(FPN): See Article 517, Part E, for X-ray installations in health care facilities.
I
Nothing in this article shall be construed as specifying safeguards against
the useful beam or stray X-ray radiation.
(FPN No. 1): Radiation safety and performance requirements of several classes
of X-ray equipment are regulated under Public Law 90-602 and are enforced by the
Department of Health and Human Services.
(FPN No. 2): In addition, information on radiation protection by the National
Council on Radiation Protection and Measurements is published as Reports of the
National Council on Radiation Protection and Measurement. These reports are
obtainable from NCRP Publications, 7910 Woodmont Ave., Suite 1016, Bethesda,
MD 20814.
660-2. Definitions.
Long-Time Rating: A rating based on an operating interval of 5 minutes
or longer.
Mobile: X-ray equipment mounted on a permanent base with wheels
and/or casters for moving while completely assembled.
Momentary Rating: A rating based on an operating interval that does
not exceed 5 seconds.
Portable: X-ray equipment designed to be hand-carried.
Transportable: X-ray equipment to be installed in a vehicle or that may
be readily disassembled for transport in a vehicle.
70-676
NATIONAL ELECTRICAL CODE
660-3. Hazardous (Classified) Locations. Unless approved for the location, X-ray and related equipment shall not be installed or operated in hazardous (classified) locations.
(FPN): See Article 517, Part D.
660-4. Connection to Supply Circuit.
(a) Fixed and Stationary Equipment. Fixed and stationary X-ray equipment shall be connected to the power supply by means of a wiring method
meeting the general requirements of this Code.
Exception: Equipment properly supplied by a branch circuit rated at not
over 30 amperes shall be permitted to be supplied through a suitable
attachment plug cap and hard-service cable or cord.
(b) Portable, Mobile, and Transportable Equipment. Individual branch
circuits shall not be required for portable, mobile, and transportable X-ray
equipment requiring a capacity of not over 60 amperes. Portable and
mobile types of X-ray equipment of any capacity shall be supplied through
a suitable hard-service cable or cord. Transportable X-ray equipment of
any capacity shall be permitted to be connected to its power supply by
suitable connections and hard-service cable or cord.
(c) Over 600 Volts, Nominal. Circuits and equipment operated at more
than 600 volts, nominal, shall comply with Article 710.
660-5. Disconnecting Means. A disconnecting means of adequate capacity for at least 50 percent of the input required for the momentary rating or
100 percent of the input required for the long-time rating of the X-ray
equipment, whichever is greater, shall be provided in the supply circuit.
The disconnecting means shall be operable from a location readily accessible from the X-ray control. For equipment connected to a 120-volt, nominal, branch circuit of 30 amperes or less, a grounding-type attachment plug
cap and receptacle of proper rating shall be permitted to serve as a disconnecting means.
660-6. Rating of Supply Conductors and Overcurrent Protection.
(a) Branch-Circuit Conductors. The ampacity of supply branch-circuit
conductors and the overcurrent protective devices shall not be less than 50
percent of the momentary rating or 100 percent of the long-time rating,
whichever is the greater.
(b) Feeder Conductors. The rated ampacity of conductors and overcurrent devices of a feeder for two or more branch circuits supplying X-ray
units shall not be less than 100 percent of the momentary demand rating [as
determined by (a)] of the two largest X-ray apparatus plus 20 percent of the
momentary ratings of other X-ray apparatus.
(FPN): The minimum conductor size for branch and feeder circuits is also governed by voltage regulation requirements. For a specific installation, the manufacturer usually specifies minimum distribution transformer and conductor sizes, rating
of disconnect means, and overcurrent protection.
ARTICLE 660 - X-RAY EQUIPMENT
70-677
660-7. Wiring Terminals. X-ray equipment shall be provided with suitable
wiring terminals or leads for the connection of power-supply conductors of
the size required by the rating of the branch circuit for the equipment.
Exception: Where provided with a permanently attached cord or a
cord set.
660-8. Number of Conductors in Raceway. The number of control circuit
conductors installed in a raceway shall be determined in accordance with