Regl. 6302, art. 405(b)(9)-20.56
Nickel
Length: 3,505 wordsOfficial source
Cite as Reglamento Núm. 6302, Art. 405(b)(9)-20.56
3.6
1.8
5.4
Phosphorus
18
9
27
Selenium
118 (0.5)
18 (0.3)
127 (0.8)
Silver
1.7
0.9
2.6
Thallium
19.6 (0.2)
14.8 (0.1)
114.4 (0.3)
Zinc
0.5
0.3
0.8
Mercury analysis only.
Detection limit when analyzed by GFAAS.
2 Detection limit when analyzed by CVAAS, estimated for Back-Half and Total Train. See Sections 2.2 and 5.4.3.
Note: Actual method In-stack detection limits may vary from these values, as described in Section 2.3.3.
2.3.2 To ensure optimum precision/resolution in the analyses, the target concentrations of metals in the analytical solutions
should be at least ten times their respective
analytical detection limits. Under certain
conditions, and with greater care in the analytical procedure. these concentrations can
be as low as approximately three times the
respective analytical detection limits without seriously impairing the precision of the
analyses. On at least one sample run in the
source test, and for each metal analyzed,
perform either repetitive analyses, Method
of Standard Additions, serial dilution, or matrix spike addition, etc., to document the
quality of the data.
2.3.3 Actual In-stack method detection
limits are based on actual source sampling
parameters and analytical results as described above. If required, the method instack detection limits can be improved over
those shown in Table 29-1 for a specific test
by either increasing the sampled stack gas
volume, reducing the total volume of the digested samples, improving the analytical detection limits, or any combination of the
three. For extremely low levels of Hg only.
the aliquot size selected for digestion and
analysis can be increased to as much as 10
ml. thus Improving the in-stack detection
limit by a factor of ten compared to a 1 ml
aliquot size.
2.3.3.1 A nominal one hour sampling run
will collect a stack gas sampling volume of
about 1.25 m3. If the sampling time is increased to four hours and 5 m 3 are collected.
the in-stack method detection limits would
be Improved by a factor of four compared to
the values shown in Table 29-1.
2.3.3.2 The In-stack detection limits assume that all of the sample is digested and
the final liquid volumes for analysis are the
normal values of 300 ml for Analytical Fraction 1. and 150 ml for Analytical Fraction 2A.
If the volume of Analytical Fraction 1 is reduced from 300 to 30 ml. the in-stack detection limits for that fraction of the sample
would be Improved by a factor of ten. If the
volume of Analytical Fraction 2A is reduced
from 150 to 25 ml. the in-stack detection Iimits for that fraction of the sample would be
improved by a factor of six. Matrix effect
checks are necessary on sample analyses and
typically are of much greater significance
for samples that have been concentrated to
Environmental Protection Agency, EPA
less than the normal original sample volume. Reduction of Analytical Fractions 1
and 2A to volumes of less than 30 and 25 ml.
respectively, could Interfere with the redissolving of the residue and could increase interference by other compounds to an intolerable level.
2.3.3.3 When both of the modifications described in Sections 2.3.3.1 and 2.3.3.2 are used
simultaneously on one sample. the resultant
Improvements are multiplicative. For example. an increase in stack gas volume by a factor of four and a reduction in the total liquid
sample digested volume of both Analytical
Fractions 1 and 2A by a factor of six would
result in an improvement by a factor of
twenty-four of the in-stack method detection
limit.
2.4 Precision. The precision (relative
standard deviation) for each metal detected
in a method development test performed at a
sewage sludge incinerator were found to be
as follows: Sb (12.7 percent). As (13.5 percent), Ba (20.6 percent), Cd (11.5 percent), Cr
(11.2 percent), Cu (11.5 percent), Pb (11.6 percent), P (14.6 percent), Se (15.3 percent). TI
(12.3 percent). and Zn (11.8 percent). The precision for Ni was 7.7 percent for another test
conducted at a source simulator. Be, Mn. and
Ag were not detected in the tests. However,
based on the analytical detection limits of
the ICAP for these metals, their precisions
could be similar to those for the other metals when detected at similar levels.
2.5 Interferences. Iron (Fe) can be a spectral interference during the analysis of As,
Cr, and Cd by ICAP. Aluminum (Al) can be a
spectral interference during the analysis of
As and Pb by ICAP. Generally, these interferences can be reduced by diluting the analytical sample. but such dilution raises the
in-stack detection limits. Background and
overlap corrections may be used to adjust for
spectral interferences. Refer to Method 6010
in EPA Publication SW-846 Third Edition
(November 1986) including updates I, II, IIA
and IIB, as incorporated by reference in
$60.17(i) the other analytical methods used
for details on potential interferences to this
method. For all GFAAS analyses, use matrix
modifiers to limit Interferences, and matrix
match all standards.
3. Apparatus
3.1 Sampling. A schematic of the sampling train is shown In Figure 29-1. It has
general similarities to the Method 5 train.
ER25AP96.000
3.1.1 Probe Nozzle (Probe Tip) and
Borosilicate or Quartz Glass Probe Liner.
Same as Method 5. Sections 2.1.1 and 2.1.2.
except that glass nozzles are required unless
alternate tips are constructed of materials
that are free from contamination and will
not interfere with the sample. If a probe tip
other than glass is used. no correction to the
Environmental Protection Agency, EPA
sample test results to compensate for the
nozzle's effect on the sample is allowed.
Probe fittings of plastic such as Teflon, polypropylene, etc. are recommended instead of
metal fittings to prevent contamination. If
one chooses to do so, a single glass piece consisting of a combined probe tip and probe
liner may be used.
3.1.2 Pitot Tube and Differential Pressure
Gauge. Same as Method 2, Sections 2.1 and
2.2, respectively.
3.1.3 Filter Holder. Glass, same as Method
5. Section 2.1.5, except use a Teflon filter
support or other non-metallic, non-contaminating support in place of the glass frit.
3.1.4 Filter Heating System. Same as
Method 5, Section 2.1.6.
3.1.5 Condenser. Use the following system
for condensing and collecting gaseous metals
and determining the moisture content of the
stack gas. The condensing system shall consist of four to seven impingers connected in
series with leak-free ground glass fittings or
other leak-free, non-contaminating fittings.
Use the first impinger as a moisture trap.
The second impinger (which is the first
HNO₃/H₂O₂ impinger) shall be identical to the
first impinger in Method 5. The third Impinger (which is the second HNO₃/H₂ O₂ Impinger) shall be a Greenburg Smith impinger
with the standard tip as described for the
second impinger in Method 5, Section 2.1.7.
The fourth (empty) impinger and the fifth
and sixth (both acidified KMnO4 ) impingers
are the same as the first impinger in Method
5. Place a thermometer capable of measuring
to within 1 °C (2 °F) at the outlet of the last
impinger. If no Hg analysis is planned, then
the fourth, fifth, and sixth impingers are not
used.
3.1.6 Metering System, Barometer, and
Gas Density Determination Equipment.
Same as Method 5, Sections 2.1.8 through
2.1.10, respectively.
3.1.7 Teflon Tape. For capping openings
and sealing connections. If necessary, on the
sampling train.
3.2. Sample Recovery. Same as Method 5,
Sections 2.2.1 through 2.2.8 (Probe-Liner and
Probe-Nozzle Brushes or Swabs. Wash Bottles, Sample Storage Containers, Petri
Dishes, Glass Graduated Cylinder, Plastic
Storage Containers, Funnel and Rubber Policeman. and Glass Funnel), respectively,
with the following exceptions and additions:
3.2.1 Non-metallic Probe-Liner and Probe-
Nozzle Brushes or Swabs. Use non-metallic
probe-liner and probe-nozzle brushes or
swabs for quantitative recovery of materials
collected in the front-half of the sampling
train.
3.2.2 Sample Storage Containers. Use
glass bottles (see the Precaution: in Section
4.3.2 of this Method) with Teflon-lined caps
that are non-reactive to the oxidizing solutions, with capacities of 1000- and 500-ml, for
storage of acidified KMnO4- containing samples and blanks. Glass or polyethylene bottles may be used for other sample types.
3.2.3 Graduated Cylinder. Glass or equivalent.
3.2.4 Funnel. Glass or equivalent.
3.2.5 Labels. For identifying samples.
3.2.6 Polypropylene Tweezers and/or Plastic Gloves. For recovery of the filter from
the sampling train filter holder.
3.3 Sample Preparation and Analysis.
3.3.1 Volumetric Flasks, 100-ml, 250-ml,
and 100-ml. For preparation of standards and
sample dilutions.
3.3.2 Graduated Cylinders. For preparation of reagents.
3.3.3 Parr Bombs or Microwave Pressure
Relief Vessels with Capping Station (CEM
Corporation model or equivalent). For sample digestion.
3.3.4 Beakers and Watch Glasses. 250-ml
beakers, with watch glass covers, for sample
digestion.
3.3.5 Ring Stands and Clamps. For securing equipment such as filtration apparatus.
3.3.6 Filter Funnels. For holding filter
paper.
3.3.7 Disposable Pasteur Pipets and Bulbs.
3.3.8 Volumetric Pipets.
3.3.9 Analytical Balance. Accurate to
within .01 mg.
3.3.10 Microwave or Conventional Oven.
For heating samples at fixed power levels or
temperatures, respectively.
3.3.11 Hot Plates.
3.3.12 Atomic Absorption Spectrometer
(AAS). Equipped with a background corrector.
3.3.12.1 Graphite Furnace Attachment.
With Sb. As, Cd. Co. Pb, Se, and TI hollow
cathode lamps (HCLs) or electrodeless discharge lamps (EDLs). Same as Methods 7041
(Sb), 7060 (As), 7131 (Cd). 7201 (Co), 7421 (Pb),
7740 (Se), and 7841 (TI) in EPA publication
SW-846 Third Edition (November 1886) including updates I, II, IIA and IIB, as incorporated by reference in $60.17(i).
3.3.12.2 Cold Vapor Mercury Attachment.
With a mercury HCL or EDL, an air recirculation pump, a quartz cell, an aerator apparatus, and a heat lamp or desiccator tube.
The heat lamp shall be capable of raising the
temperature at the quartz cell by 10 °C above
ambient. so that no condensation forms on
the wall of the quartz cell. Same as Method
6020 in EPA publication SW-846 Third Edition (November 1986) including updates I. II,
IIA and IIB, as incorporated by reference in
$60.17(i). See NOTE No. 2: Section 5.4.3 for
other acceptable approaches for analysis of
Hg in which analytical detection limits of
0.002 ng/ml were obtained.
3.3.13 Inductively Coupled Argon Plasma
Spectrometer. With either a direct or sequential reader and an alumina torch. Same
as EPA Method 6010 in EPA publication SW-
846 Third Edition (November 1986) including
Pt. 60, App. A, Meth. 29
updates I, II. IIA and IIB, as Incorporated by
reference in $60.17(i).
3.3.14 Inductively Coupled Plasma-Mass
Spectrometer. Same as EPA Method 6020 in
EPA publication SW-846 Third Edition (November 1986) including updates I. II, IIA and
ПВ, as incorporated by reference in $60.17(i).
d. Reagents
4.1 Unless otherwise indicated, it is intended that all reagents conform to the specifications established by the Committee on
Analytical Reagents of the American Chemical Society, where such specifications are
available. Otherwise. use the best available
grade.
4.2 Sampling Reagents.
4.2.1 Sample Filters. Without organic
binders. The filters shall contain less than
1.3 µg/In.² of each of the metals to be measured. Analytical results provided by filter
manufacturers stating metals content of the
filters are acceptable. However, If no such results are available, analyze filter blanks for
each target metal prior to emission testing.
Quartz fiber filters meeting these requirements are recommended. However, if glass
fiber filters become available which meet
these requirements, they may be used. Filter
efficiencies and unreactiveness to sulfur dioxide (SO₂) or sulfur trioxide (SO 3) shall be
as described in Section 3.1.1 of Method 5.
4.2.2 Water. To conform to ASTM Specification D1193-77, Type II (incorporated by
reference-See $60.17). If necessary, analyze
the water for all target metals prior to field
use. All target metals should be less than 1
ng/ml.
4.2.3 Nitric Acid (HNO₂). Concentrated.
Baker Instra-analyzed or equivalent.
4.2.4 Hydrochloric Acid (HCL). Concentrated. Baker Instra-analyzed or equivalent.
4.2.5 Hydrogen Peroxide (H₂O₂), 30 Percent
(V/V).
4.2.6 Potassium Permanganate (KMnO4 ).
4.2.7 Sulfuric Acid (H2SO4). Concentrated.
4.2.8 Silica Gel and Crushed Ice. Same as
Method 5, Sections 3.1.2 and 3.1.4. respectively.
4.3 Pretest Preparation of Sampling Reagents.
4.3.1 Absorbing Solution, 5 Percent HNO3/10 Percent H₂O₂. Add carefully
with stirring 50 ml of concentrated HNO3 to
a 1000-ml volumeric flask containing approximately 500 ml of water, and then add
carefully with stirring 333 ml of 30 percent
H2O2. Dilute to volume with water. Mix well.
This reagent shall contain less than 2 ng/ml
of each target metal.
4.3.2 Acidic KMnO4 Absorbing Solution, 4
Percent KMnO₄ (W/V). 10 Percent H2SO4 (V/
V). Prepare fresh daily. Mix carefully. with
stirring, 100 ml of concentrated H2SO4 into
approximately 800 ml of water. and add
water with stirring to make a volume of 1
liter: this solution is 10 percent H2SO4 (V/V).
Dissolve, with stirring, 40 g of KMnO4 into 10
percent H₂ SO4 (V/V) and add 10 percent H2SO4
(V/V) with stirring to make a volume of 1
liter. Prepare and store in glass bottles to
prevent degradation. This reagent shall contain less than 2 ng/ml of Hg.
Precaution: To prevent autocatalytic decomposition of the permanganate solution, filter
the solution through Whatman 541 filter
paper. Also. due to the potential reaction of
the potassium permanganate with the acid,
there could be pressure buildup in the solution storage bottle. Therefore these bottles
shall not be fully filled and shall be vented
to relieve excess pressure and prevent explosion potentials. Venting is required, but not
in a manner that will allow contamination of
the solution. A No. 70-72 hole drilled in the
container cap and Teflon liner has been used.
4.3.3 HNO3, 0.1 N. Add with stirring 6.3 ml
of concentrated HNO3 (70 percent) to a flask
containing approximately 900 ml of water.
Dilute to 1000 ml with water. Mix well. This
reagent shall contain less than 2 ng/ml of
each target metal.
4.3.4 HCI, 8 N. Carefully add with stirring
690 ml of concentrated HCI to a flask containing 250 ml of water. Dilute to 1000 ml
with water. Mix well. This reagent shall contain less than 2 ng/ml of Hg.
4.4 Glassware Cleaning Reagents.
4.4.1 HNO3. Concentrated. Fisher ACS
grade or equivalent.
4.4.2 Water. To conform to ASTM Specification D1193-77, Type II (incorporated by
reference-See $60.17).
4.4.3 HNO3, 10 Percent (V/V). Add with
stirring 500 ml of concentrated HNO3 to a
flask containing approximately 4000 ml of
water. Dilute to 5000 ml with water. Mix
well. This reagent shall contain less than 2
ng/ml of each target metal.
4.5 Sample Digestion and Analysis Reagents.
The metals standards, except Hg, may also
be made from solid chemicals as described in
Citation 3 of the Bibliography. Refer to Citations 1. 2. or 5 of the Bibliography for additional information on Hg standards. The 1000
µg/ml Hg stock solution standard may be
made according to Section 6.2.5 of Method
101A.
4.5.1 HCL. Concentrated.
4.5.2 Hydrofluoric Acid (HF), Concentrated.
4.5.3 HNO3. Concentrated. Baker Instraanalyzed or equivalent.
4.5.4 HNO3. 50 Percent (V/V). Add with
stirring 125 ml of concentrated HNO3 to 100
ml of water. Dilute to 250 ml with water. Mix
well. This reagent shall contain less than 2
ng/ml of each target metal.
4.5.5 HNO3, 5 Percent (V/V). Add with stirring 50 ml of concentrated HNO3 to 800 ml of
water. Dilute to 1000 ml with water. Mix
Environmental Protection Agency, EPA
well. This reagent shall contain less than 2
ng/ml of each target metal.
4.5.6 Water. To conform to ASTM Specification D1193-77, Type II (incorporated by
reference-See $60.17).
4.5.7 Hydroxylamine Hydrochloride and
Sodium Chloride Solution. See Citation 2 of
the Bibliography for preparation.
4.5.8 Stannous Chloride. See Citation 2 of
the Bibliography for preparation.
4.5.9 KMnO4, 5 Percent (W/V). See Citation
2 of the Bibliography for preparation.
4.5.10 H2SO4. Concentrated.
4.5.11 Potassium Persulfate, 5 Percent (W/
V). See Citation 2 of the Bibliography for
preparation.
4.5.12 Nickel Nitrate, NI (NO₃ )₂ 6H₂O.
4.5.13 Lanthanum Oxide. Laz O3.
4.5.14 Hg Standard (AAS Grade), 1000 µg/
ml.
4.5.15 Pb Standard (AAS Grade), 1000 µg/
ml.
4.5.16 As Standard (AAS Grade), 1000 µg/
ml.
4.5.17 Cd Standard (AAS Grade), 1000 µg/
ml.
4.5.18 Cr Standard (AAS Grade), 1000 µg/
ml.
4.5.19 Sb Standard (AAS Grade), 1000 µg/
ml.
4.5.20 Ba Standard (AAS Grade), 1000 µg/
ml.
4.5.21 Be Standard (AAS Grade), 1000 µg/
ml.
4.5.22 Co Standard (AAS Grade), 1000 µg/
ml.
ml.
4.5.28 Ag Standard (AAS Grade), 1000 µg/
ml.
4.5.29 TI Standard (AAS Grade). 1000 µg/
ml.
4.5.30 Zn Standard (AAS Grade). 1000 µg/
ml.
4.5.31 AI Standard (AAS Grade), 1000 µg/
ml.
4.5.32 Fe Standard (AAS Grade), 1000 µg/
ml.
4.5.23 Cu Standard (AAS Grade). 1000 µg/
ml.
4.5.24 Mn Standard (AAS Grade), 1000 µg/
ml.
4.5.25 Ni Standard (AAS Grade), 1000 µg/
4.5.26 P Standard (AAS Grade), 1000 µg/ml.
4.5.27 Se Standard (AAS Grade). 1000 µg/
ml.
4.5.33 Hg Standards and Quality Control
Samples. Prepare fresh weekly a 10 µg/ml intermediate Hg standard by adding 5 ml of
1000 µg/ml Hg stock solution prepared according to Method 101A to a 500-ml volumetric flask; dilute with stirring to 500 ml
by first carefully adding 20 ml of 15 percent
HNO3 and then adding water to the 500-ml
volume. Mix well. Prepare a 200 ng/ml working Hg standard solution fresh daily: add 5
ml of the 10 µg/ml Intermediate standard to
a 250-ml volumetric flask, and dilute to 250
ml with 5 ml of 4 percent KMnO4. 5 ml of 15
percent HNO3. and then water. Mix well. Use
at least five separate aliquots of the working
Hg standard solution and a blank to prepare
the standard curve. These aliquots and blank
shall contain 0.0, 1.0, 2.0, 3.0, 4.0. and 5.0 ml
of the working standard solution containing
0. 200, 400, 600, 800, and 1000 ng Hg, respectively. Prepare quality control samples by
making a separate 10 µg/ml standard and diluting until in the calibration range.
4.5.34 ICAP Standards and Quality Control Samples. Calibration standards for ICAP
analysis can be combined into four different
mixed standard solutions as follows:
MIXED STANDARD SOLUTIONS FOR ICAP
ANALYSIS
Solution
Elements
I
As, Be, Cd, Mn. Pb, Se, Zn.
u
Ba, Co, Cu, Fe.
m
AI, Cr, NI.
IV
Ag, P, Sb, TL
Prepare these standards by combining and
diluting the appropriate volumes of the 1000
µg/ml solutions with 5 percent HNO3. A minimum of one standard and a blank can be
used to form each calibration curve. However, prepare a separate quality control sample spiked with known amounts of the target
metals in quantities In the mid-range of the
calibration curve. Suggested standard levels
are 25 µg/ml for Al. Cr and Pb. 15 µg/ml for
Fe, and 10 µg/ml for the remaining elements.
Prepare any standards containing less than 1
µg/ml of metal on a daily basis. Standards
containing greater than 1 µg/ml of metal
should be stable for a minimum of 1 to 2
weeks. For ICP-MS. follow Method 6020 in
EPA Publication SW-846 Third Edition (November 1986) including updates I, II, IIA and
IIB, as incorporated by reference in $60.17(i).
4.5.35 GFAAS Standards. Sb, As, Cd, Co,
Pb, Se, and TI. Prepare a 10 µg/ml standard
by adding 1 ml of 1000 µg/ml standard to a
100-ml volumetric flask. Dilute with stirring
to 100 ml with 10 percent HNO3. For GFAAS,
matrix match the standards. Prepare a 100
ng/ml standard by adding 1 ml of the 10 µg/ml
standard to a 100-ml volumetric flask, and
dilute to 100 ml with the appropriate matrix
solution. Prepare other standards by diluting
the 100 ng/ml standards. Use at least five
standards to make up the standard curve.
Suggested levels are 0. 10. 50. 75. and 100 ng/
ml. Prepare quality control samples by making a separate 10 µg/ml standard and diluting
until it is in the range of the samples. Prepare any standards containing less than 1 µg/
ml of metal on a daily basis. Standards containing greater than 1 µg/ml of metal should
be stable for a minimum of I to 2 weeks.
4.5.36 Matrix Modifiers.
4.5.36.1 Nickel Nitrate, 1 Percent (V/V).
Dissolve 4.956 g of Ni or
Pt. 60, App. A, Meth. 29
other nickel compound suitable for preparation of this matrix modifier in approximately 50 ml of water in a 100-ml volumetric
flask. Dilute to 100 ml with water.
4.5.36.2 Nickel Nitrate, 0.1 Percent (V/V).
Dilute 10 ml of 1 percent nickel nitrate solution to 100 ml with water. Inject an equal
amount of sample and this modifier into the
graphite furnace during GFAAS analysis for
As.
4.5.36.3 Lanthanum. Carefully dissolve
0.5864 g of La2 O3 in 10 ml of concentrated
HNO3, and dilute the solution by adding It
with stirring to approximately 50 ml of
water. Dilute to 100 ml with water, and mix
well. Inject an equal amount of sample and
this modifier into the graphite furnace during GFAAS analysis for Pb.
4.5.37 Whatman 40 and 541 Filter Papers
(or equivalent). For filtration of digested
samples.
5. Procedure
5.1 Sampling. The complexity of this
method is such that, to obtain reliable results, both testers and analysts must be
trained and experienced with the test procedures, including source sampling: reagent
preparation and handling; sample handling:
safety equipment and procedures: analytical
calculations; reporting: and the specific procedural descriptions throughout this method.
5.1.1 Pretest Preparation. Follow the
same general procedure given in Method 5,