Regl. 6302, art. 405(b)(9)-3.4
Using this calibration spectrophotometer,
Length: 841 wordsOfficial source
Cite as Reglamento Núm. 6302, Art. 405(b)(9)-3.4
calibrate the required filters or screens. Make measurements at wavelength intervals of 20 nm or less. As an alternative procedure, use the calibration spectrophotometer
to measure the C.I.E. Daylight c luminous transmittance of
the attenuators. Check the attenuators several times, at
different locations on the attermator.
The attenuator manufacturer must specify the period of
time over which the attenuator values can be considered
stable, as well as any special handling and storing procedures required to enhance attenuator stability. To assure
stability. recheck attenuator values at intervals less than or
equal to the period stability guaranteed by the manufacturer. Recheck at least every 3 months. If desired, perform the stability checks with an instrument (secondary)
other than the calibration spectrophotometer. This secondary instrument must be a high-quality laboratory transmissometer or spectrophotometer, and the same instrument
must always be used for the stability checks. If a secondary instrument is to be used for stability checks, the value
of the calibrated attenuator must be measured on this secondary instrument immediately following initial calibration. If over a period of time an attenuator value changes
by more than ±2 percent opacity, recalibrate the attenuator on the calibration spectrophotometer or replace it
with a new attenuator.
If this procedure is conducted by the filter or screen
manufacturer or by an independent laboratory, obtain a
statement certifying the values and certifying that the
specified procedure, or equivalent, is used.
7.1.4 Calibration Error Test. Insert the calibration attenuators (low. mid, and high range) in the transmissometer path at or as near the midpoint of the path as feasible. Place the attenuator in the measurement path at a
point where the effluent will be measured; i.e., do not
place the calibration attenuator in the instrument housing.
If the instrument manufacturer recommends a procedure
wherein the attenuators are placed in the instrument housing. the manufacturer must provide data showing this alternative procedure is acceptable. While inserting the attenuator, assure that the entire beam received by the detector will pass through the attenuator and that the attemuator is inserted in a manner which minimizes interference
from reflected light. Make a total of five nonconsecutive
readings for each filter. Record the monitoring system
output readings in percent opacity (see example Figure 1-.
6). Then, if the path length is not adjusted by the measurement system. subtract the actual calibration attenuator
value from the value indicated by the measurement system recorder for each of the 15 readings obtained. If the
path length is adjusted by the measurement system, subtract the "path adjusted" calibration attenuator values
from the values indicated by the measurement system recorder (the "path adjusted" calibration attenuator values
are calculated using Equation 1-6 or 1-7). Calculate the
arithmetic mean difference, standard deviation, and confidence coefficient of the five tests at each attenuator
value using Equations 1-2, I-3, and 1-4 (Sections 8.1-
8.3). Calculate the sum of the absolute value of the mean
difference and the absolute value of the confidence coefficient for each of the three test attenuators: report these
three values as the calibration error.
Insert illus. 0053
Pt. 60, App. B, Spec. 1
7.1.5 System Response Test. Insert the high-range
calibration attenuator in the transmissometer path five
times, and record the time required for the system to respond to 95 percent of final zero and high-range filter
values (see example Figure 1-7). Then calculate the mean
time of the 10 upscale and downscale tests and report this
value as the system response time.
Insertillus.0055
7.2 Preliminary Field Adjustments. Install the CEMS
on the affected facility according to the manufacturer's
written instructions and the specifications in Section 4,
and perform the following preliminary adjustments:
7.2.1 Optical and Zero Alignment. When the facility
is not in operation, optically align the light beam of the
transmissometer upon the optical surface located across
the duct or stack (i.e., the retroreflector or photodetector,
as applicable) in accordance with the manufacturer's instructions; verify the alignment with the optical alignment
sight. Under clear stack conditions, verify the zero alignment (performed in Section 7.1.1) by assuring that the
monitoring system response for the simulated zero check
coincides with the actual zero measured by the transmissometer across the clear stack. Adjust the zero alignment,
if necessary. Then, after the affected facility has been
started up and the effluent stream reaches normal operating temperature. recheck the optical alignment. If the optical alignment has shifted, realign the optics. Note: Careful
consideration should be given to whether a "clear stack"
condition exists. It is suggested that the stack be monitored and the data output (instantaneous real-time basis)
be examined to determine whether fluctuations from zero
opacity are occurring before a clear stack condition is assumed to exist.
7.2.2 Optical and Zero Alignment (Alternative Procedure). The procedure given in 7.2.1 is the preferred procedure and should be used whenever possible; however, if
the facility is operating and a zero stack condition cannot
practicably be obtained, use the zero alignment obtained
during the preliminary adjustments (Section 7.1.1) before
installing the transmissometer on the stack. After completing all the preliminary adjustments and tests required in