Regl. 6302, art. 405(b)(9)-4
Enmienda las Reglas 102 y 405 del Reglamento para el Control de la Contaminación Atmosférica
Length: 1,567 wordsOfficial source
Cite as Reglamento Núm. 6302, Art. 405(b)(9)-4
1.2 Principle. The opacity of particulate matter in
stack emissions is continuously monitored by a measurement system based upon the principle of transmissometry.
Light having specific spectral characteristics is projected
from a lamp through the effluent in the stack or duct, and
the intensity of the projected light is measured by a sensor. The projected light is attenuated because of absorption and scattered by the particulate matter in the effluent;
the percentage of visible light attenuated is defined as the
opacity of the emission. Transparent stack emissions that
do not attenuate light will have a transmittance of 100
percent or an opacity of zero percent. Opaque stack emissions that attenuate all of the visible light will have a
transmittance of zero percent or an opacity of 100 percent.
This specification establishes specific design criteria for
the transmissometer system. Any opacity CEMS that is
expected to meet this specification is first checked to verify that the design specifications are met. Then, the opacity CEMS is calibrated, installed, and operated for a specified length of time. During this specified time period, the
system is evaluated to determine conformance with the
established performance specifications.
2.1 Continuous Emission Monitoring System. The
total equipment required for the determination of opacity.
The system consists of the following major subsystems:
2.1.1 Sample Interface. That portion of CEMS that
protects the analyzer from the effects of the stack effluent
and aids in keeping the optical surfaces clean.
2.1.2 Analyzer. That portion of the CEMS that senses
the pollutant and generates an output that is a function of
the opacity.
2.1.3 Data Recorder. That portion of the CEMS that
provides a permanent record of the analyzer output in
terms of opacity. The data recorder may include automatic data-reduction capabilities.
2.2 Transmissometer. That portion of the CEMS that
includes the sample interface and the analyzer.
2.3 Transmittance. The fraction of incident light that
is transmitted through an optical medium.
2.4 Opacity. The fraction of incident light that is attenuated by an optical medium. Opacity (Op) and transmittance (Tr) are related by: Op=1 Tr.
25 Optical Density. A logarithmic measure of the
amount of incident light attenuated. Optical density (D) is
related to the transmittance and opacity as follows:
D=10g10 Tr=logio (1-Op).
2.6 Peak Spectral Response. The wavelength of maximum sensitivity of the transmissometer.
2.7 Mean Spectral Response. The wavelength that is
the arithmetic mean value of the wavelength distribution
for the effective spectral response curve of the transmissometer.
2.8 Angle of View. The angle that contains all of the
radiation detected by the photodetector assembly of the
analyzer at a level greater than 25 percent of the peak
detector response.
2.9 Angle of Projection. The angle that contains all of
the radiation projected from the lamp assembly of the analyzer at a level of greater than 2.5 percent of the peak
illuminance.
2.10 Span Value. The opacity value at which the
CEMS is set to produce the maximum data display output
as specified in the applicable subpart.
2.11 Upscale Calibration Value. The opacity value at
which a calibration check of the CEMS is performed by
simulating an upscale opacity condition as viewed by the
receiver.
2. Definitions
Pt. 60, App. B, Spec. 1
2.12 Calibration Error. The difference between the
opacity values indicated by the CEMS and the known values of a series of calibration attenuators (filters or
screens).
2.13 Zero Drift. The difference in the CEMS output
readings from the zero calibration value after a stated period of normal continuous operation during which no umscheduled maintenance, repair, or adjustment took place.
A calibration value of 10 percent opacity or less may be
used in place of the zero calibration value.
2.14 Calibration Drift. The difference in the CEMS
output readings from the upscale calibration value after a
stated period of normal continuous operation during
which no unscheduled maintenance, repair, or adjustment
took place.
2.15 Response Time. The amount of time it takes the
CEMS to display on the data recorder 95 percent of a
step change in opacity.
2.16 Conditioning Period. A period of time (168
hours minimum) during which the CEMS is operated
without any unscheduled maintenance, repair, or adjustment prior to initiation of the operational test period.
2.17 Operational Test Period. A period of time (168
hours) during which the CEMS is expected to operate
within the established performance specifications without
any unscheduled maintenance, repair, or adjustment.
2.18 Path Length. The depth of effluent in the light
beam between the receiver and the transmitter of a singlepass transmissometer. or the depth of effluent between the
transceiver and reflector of a double-pass transmissometer. Two path lengths are referenced by this specification as follows:
2.18.1 Monitor Path Length. The path length (depth
of effluent) at the installed location of the CEMS.
2.18.2 Emission Outlet Path Length. The path length
(depth of effluent) at the location where emissions are released to the atmosphere. For noncircular outlets,
D₆=(2LW)+(L+W), where L is the length of the outlet
and W is the width of the outlet. Note that this definition
does not apply to pressure baghouse outlets with multiple
stacks, side discharge vents, ridge roof monitors, etc.
3. Apparatus
3.1 Opacity Continuous Emission Monitoring System.
Any opacity CEMS that is expected to meet the design
and performance specifications in Section 5 and a suitable
data recorder, such as an analog strip chart recorder or
other suitable device (e.g., digital computer) with an input
signal range compatible with the analyzer output.
3.2 Calibration Attenuators. Minimum of three. These
attenuators must be optical filters or screens with neutral
spectral characteristics selected and calibrated according
to the procedures in Sections 7.1.2 and 7.1.3, and of sufficient size to attenuate the entire light beam received by
the detector of the transmissometer.
3.3 Upscale Calibration Value Attenuator. An optical
filter with neutral spectral characteristics, a screen, or
other device that produces an opacity value (corrected for
path length. if necessary) that is greater than or equal to
the applicable opacity standard but less than or equal to
one-half the applicable instrument span value.
3.4 Calibration Spectrophotometer. A laboratory
spectrophotometer meeting the following minimum design
specifications:
Parameter
Specification
Wavelength range
400-700 nm.
Detector angle of view
<10°.
Accuracy
<0.5 percent transmittance,
NBS traceable calibration.
4. Installation Specifications
Install the CEMS at a location where the opacity measurements are representative of the total emissions from
the affected facility. These requirements can be met as
follows:
4.1 Measurement Location. Select a measurement 10-
cation that is (a) downstream from all particulate control
equipment, (b) where condensed water vapor is not
present, (c) free of interference from ambient light (applicable only if transmissometer is responsive to ambient
light). and (d) accessible in order to permit routine maintenance. Accessibility is an important criterion because
easy access for lens cleaning, alignment checks, calibration checks, and blower maintenance will help assure
quality data.
4.2 Measurement Path. The primary concern in locating a transmissometer is determining a location of wellmixed stack gas. Two factors contribute to complete mixing of emission gases: turbulence and sufficient mixing
time. The criteria listed below define conditions under
which well-mixed emissions can be expected.
Select a measurement path that passes through a centroidal area equal to 25 percent of the cross section. Additional requirements or modifications must be met for certain locations as follows:
4.2.1 If the location is in a straight vertical section of
stack or duct and is less than 4 equivalent diameters
downstream from a bend, use a path that is in the plane
defined by the upstream bend (see Figure I-1).
4.2.2 If the location is in a straight vertical section of
stack or duct and is less than 4 equivalent diameters upstream from a bend, use a path that is in the plane defined by the bend (see Figure 1-2).
Pt. 60, App. B, Spec. 1
Insertillus.0035
Pt. 60, App. B, Spec. 1
Insertillus.0038
4.2.3 If the location is in a straight vertical section of
stack or duct and is less than 4 diameters downstream and
is also less than 1 diameter upstream from a bend, use
a path in the plane defined by the upstream bend (see
Figure 1-3).
4.2.4 If the location is in a horizontal section of duct
and is at least 4 diameters downstream from a vertical
bend, use a path in the horizontal plane that is between
one-third and one-half the distance up the vertical axis
from the bottom of the duct (see Figure 1-4).
4.2.5 If the location is in a horizontal section of duct
and is less than 4 diameters downstream from a vertical
bend, use a path in the horizontal plane that is between
one-half and two-thirds the distance up the vertical axis
from the bottom of the duct for upward flow in the vertical section, and is between one-third and one-half the distance up the vertical axis from the bottom of the duct for
downward flow (Figure I-5).
4.3 Alternative Locations and Measurement Paths.
Other locations and measurement paths may be selected
by demonstrating to the Administrator that the average
opacity measured at the alternative location or path is
equivalent to the opacity as measured at a location meeting the criteria of Sections 4.1 and 4.2. The opacity at the
alternative location is considered equivalent if the average
value measured at the alternative location is within the
range defined by the average measured opacity ±10 percent at the location meeting the installation criteria in