Regl. 6302, art. 405(b)(9)-4.2
Prepare standards to span the 0-
Length: 12,005 wordsOfficial source
Cite as Reglamento Núm. 6302, Art. 405(b)(9)-4.2
to 400- or 400- to 1000-ppm range. If any samples span both concentration ranges, prepare
a calibration curve for each range. A set of
three bulbs containing colorimetric reagent
but no CO should serve as a reagent blank
and be taken through the analysis procedure.
Calculate the average absorbance for each
set (3 bulbs) of standards using Equation
10A-1 and Table 10A-1. Construct a graph of
average absorbance for each standard
against its corresponding concentration in
ppm. Draw a smooth curve through the
points. The curve should be linear over the
two concentration ranges discussed In Section 1.3.1.
6. Calculations
Carry out calculations retaining at least
one extra decimal figure beyond that of the
acquired data. Round off figures after final
calculation.
6.1 Nomenclature.
A=Sample absorbance, uncorrected for the
reagent blank.
A,=Absorbance of the reagent blank.
A,=Average sample absorbance per liter,
units/liter.
B.=Moisture content In the bag sample.
C=CO concentration in the stack gas, dry
basis, ppm.
C₂=CO concentration of the bag sample, dry
basis, ppm.
Cg=CO concentration from the calibration
curve, ppm.
F=Volume fraction of CO₂ in the stack.
n=Number of reaction bulbs used per bag
sample.
Pt pressure. mm Hg.
P.=Residual pressure in the sample bulb
after evacuation. mm Hg.
Pw=Vapor pressure of H 20 in the bag (from
Table 10A-2), mm Hg.
V6=Volume of the sample bulb, liters.
V-Volume of reagent added to the sample
bulb, 0.0100 liter.
6.2 Average Sample Absorbance per Liter.
Average the three absorbance values for
each bulb set. Then calculate A, for each set
of gas bulbs using Equation 10A-1. Use A, to
determine the CO concentration from the
calibration curve (C₂)
NOTE: A and A, must be at the same wavelength.
6.3 CO Concentration in the Bag.
Calculate C₆ using Equations 10A-2 and
10A-3. If condensate is visible in the Tedlar
bag, calculate B. using Table 10A-2 and the
temperature and barometric pressure in the
analysis room. If condensate is not visible,
calculate Bw using the temperature and barometric pressure recorded at the sampling
site.
EC16NO91.168
EC16NO91.169
METHOD 10 B 40 CFR PART 60 APPENDIXA
Pt. 60, App. A, Meth. 10B
EC16NO91.170
6.4 CO Concentration in the Stack.
EC16NO91.171
EC01JN92.185
7. BIBLIOGRAPHY
1. Butler. F.E., J.E. Knoll, and M.R.
Midgett. Development and Evaluation of
Methods for Determining Carbon Monoxide
Emissions. Quality Assurance Division. Environmental Monitoring Systems Laboratory, U.S. Environmental Protection Agency, Research Triangle Park, NC 27711. June
1985. 33 p.
2. Ferguson. B.B., R.E. Lester. and W.J.
Mitchell. Field Evaluation of Carbon Monoxide and Hydrogen Sulfide Continuous
Emission Monitors at an Oil Refinery. U.S.
Environmental Protection Agency. Research
Triangle Park, NC. Publication No. EPA-600/
4-82-054. August 1982. 100 p.
3. Lambert, J.L., and R.E. Welns. Induced
Colorimetric Method for Carbon Monoxide.
Analytical Chemistry. 46(7):929-930. June
1974.
4. Levaggi. D.A., and M. Feldstein. The
Colorimetric Determination of Low Concentrations of Carbon Monoxide. Industrial
Hygiene Journal. 25.64-66. January-February
1964.
5. Repp. M. Evaluation of Continuous Monitors for Carbon Monoxide in Stationary
Sources. U.S. Environmental Protection
Agency. Research Triangle Park, NC. Publication No. EPA-600/2-77-063. March 1977. 155
p.
6. Smith, F., D.E. Wagoner, and R.P. Donovan. Guidelines for Development of a Quality
Assurance Program: Volume VIII-Determination of CO Emissions from Stationary
Sources by NDIR Spectrometry. U.S. Envlronmental Protection Agency. Research Triangle Park, NC. Publication No. EPA-850/4-
74-005-h. February 1975. 96 p.
METHOD 10B-DETERMINATION OF CARBON
MONOXIDE EMISSIONS FROM STATIONARY
SOURCES
1. Applicability and Principle
1.1 Applicability. This method applies to
the measurement of carbon monoxide (CO)
emissions at petroleum refineries and from
other sources when specified in an applicable
subpart of the regulations.
Environmental Protection Agency, EPA
Pt. 60, App. A, Meth. 10B
1.2 Principle. An integrated gas sample is
extracted from the sampling point and analyzed for CO. The sample is passed through a
conditioning system to remove interferences
and collected in a Tedlar bag. The CO is separated from the sample by gas chromatography (GC) and catalytically reduced to
methane (CH4) prior to analysis by flame
ionization detection FID. The analytical portion of this method is Identical to applicable
sections in Method 25 detailing CO measurement. The oxidation catalyst required in
Method 25 is not needed for sample analysis.
Complete Method 25 analytical systems are
acceptable alternatives when calibrated for
CO and operated by the Method 25 analytical
procedures.
NOTE: Mention of trade names or commercial products in this method does not constitute the endorsement or recommendation
for use by the Environmental Protection
Agency.
1.3 Interferences. Carbon dioxide (CO₂)
and organics potentially can interfere with
the analysis. Carbon dioxide is primarily removed from the sample by the alkaline permanganate conditioning system; any residual CO₂ and organics are separated from the
CO by GC.
2. Apparatus
2.1 Sampling. Same as in Method 10A, section 2.1.
2.2 Analysis.
2.2.1 Gas Chromatographic (GC) Analyzer.
A semicontinuous GC/FID analyzer capable
of quantifying CO in the sample and containing at least the following major components.
2.2.1.1 Separation Column. A column that
separates CO from CO₂ and organic compounds that may be present. A 1/8-in. OD
stainless-steel column packed with 5.5 ft of
60/80 mesh Carbosieve S-II (available from
Supelco) has been used successfully for this
purpose. The column listed in Addendum 1 of
Method 25 is also acceptable.
2.2.1.2 Reduction Catalyst. Same as in
Method 25, section 2.3.2.
2.2.1.3 Sample Injection System. Same as
in Method 25, section 2.3.4, equipped to accept a sample line from the Tedlar bag.
2.2.1.4 Flame Ionization Detector. Linearlty meeting the specifications In section
2.3.5.1 of Method 25 where the linearity check
is carried out using standard gases containing 20-, 200-, and 1.000-ppm CO. The minimal instrument range shall span 10 to 1,000
ppm CO.
2.2.1.5 Data Recording System. Same as in
Method 25, section 2.3.6.
3. Reagents
3.1 Sampling. Same as in Method 10A, section 3.1.
3.2 Analysis.
3.2.1 Carrier, Fuel, and Combustion Gases.
Same as in Method 25, sections 3.2.1, 3.2.2.
and 3.2.3.
3.2.2 Linearity and Calibration Gases.
Three standard gases with nominal CO concentrations of 20-, 200-, and 1,000-ppm CO in
nitrogen.
3.2.3 Reduction Catalyst Efficiency Check
Calibration Gas. Standard CH₄ gas with a
concentration of 1,000 ppm in air.
4. Procedure
4.1 Sample Bag Leak-checks, Sampling,
and CO₂ Measurement. Same as in Method
10A, sections 4.1. 4.2. and 4.3.
4.2 Preparation for Analysis. Before putting the GC analyzer into routine operation,
conduct the calibration procedures listed In
section 5. Establish an appropriate carrier
flow rate and detector temperature for the
specific instrument used.
4.3 Sample Analysis. Purge the sample
loop with sample, and then inject the sample. Analyze each sample in triplicate, and
calculate the average sample area (A). Determine the bag CO concentration according to
section 6.2.
5. Callbration
5.1 Carrier Gas Blank Check. Analyze
each new tank of carrier gas with the GC analyzer according to section 4.3 to check for
contamination. The corresponding concentration must be less than 5 ppm for the
tank to be acceptable for use.
5.2 Reduction Catalyst Efficiency Check.
Prior to initial use, the reduction catalyst
shall be tested for reduction efficiency. With
the heated reduction catalyst bypassed,
make triplicate Injections of the 1,000-ppm
CH4 gas (section 3.2.3) to calibrate the analyzer. Repeat the procedure using 1.000-ppm
CO (section 3.2.2) with the catalyst in operation. The reduction catalyst operation is
acceptable if the CO response is within 5 percent of the certified gas value.
5.3 Analyzer Linearity Check and Calibration. Perform this test before the system is
first placed into operation. With the reduction catalyst In operation, conduct a linearlty check of the analyzer using the standards specified in section 3.2.2. Make triplicate injections of each calibration gas, and
then calculate the average response factor
(area/ppm) for each gas. as well as the overall mean of the response factor values. The
instrument linearity is acceptable if the average response factor of each calibration gas
is within 2.5 percent of the overall mean
value and if the relative standard deviation
(calculated in section 6.9 of Method 25) for
each set of triplicate injections Is less than
2 percent. Record the overall mean of the response factor values as the calibration response factor (R).
Pt. 60, App. A, Meth. 11
40 CFR Ch. I (7-1-99 Edition)
6. Calculations
Carry out calculations retaining at least
one extra decimal figure beyond that of the
acquired data. Round off results only after
the final calculation.
6.1 Nomenclature.
A=Average sample area.
B.=Moisture content In the bag sample,
fraction.
C=CO concentration in the stack gas, dry
basis, ppm.
Cb=CO concentration in the bag sample,
dry basis, ppm.
F=Volume fraction of CO2 in the stack,
fraction.
Pta-Barometric pressure, mm Hg.
Pw=Vapor pressure H2O in the bag (from
Table 10-2, Method 10A). mm Hg.
R=Mean calibration response factor. area/
ppm.
6.2 CO Concentration in the Bag. Calculate C₃ using Equations 10B-1 and 10B-2. If
condensate is visible in the Tedlar bag, calculate Bw using Table 10A-1 of Method 10A
and the temperature and barometric pressure in the analysis room. If condensate is
not visible, calculate B W using the temperature and barometric pressure at the sampling site.
EC18NO91.172
EC16NO91.173
6.3 CO Concentration in the Stack.
C=C₆(1
Eq. 10B-3
1. Butler. F.E, J.E. Knoll, and M.R.
Midgett. Development and Evaluation of
Methods for Determining Carbon Monoxide
Emissions. Quality Assurance Division. Environmental Monitoring Systems Laboratory, U.S. Environmental Protection Agency, Research Triangle Park, NC 27711. June
1985. 33p.
2. Salo, A.E., S. Witz, and R.D. MacPhee.
Determination of Solvent Vapor Concentrations by Total Combustion Analysis: A Comparison of Infrared with Flame Ionization
Detectors. Paper No. 75-33.2. (Presented at
the 68th Annual Meeting of the Air Pollution
Control Association. Boston, MA. June 15.
1975.) 14 p.
3. Salo. A.E., W.L. Oaks, and R.D.
MacPhee. Measuring the Organic Carbon
Content of Source Emissions for Air Pollution Control. Paper No. 74-190. (Presented at
the 67th Annual Meeting of the Air Pollution
7. Bibliography
Control Association. Denver, CO. June 9,
1974.) 25 p.
METHOD 11-DETERMINATION OF HYDROGEN
SULFIDE CONTENT OF FUEL GAS STREAMS IN
PETROLEUM REFINERIES
1. Principle and Applicability
1.1 Principle. Hydrogen sulfide (H2S) is collected from a source in a series of midget
impingers and absorbed In pH 3.0 cadmium
sulfate (CdSO4) solution to form cadmium
sulfide (CdS). The latter compound is then
measured iodometrically. An impinger containing hydrogen peroxide is Included to remove SO₂ as an interfering species. This
method is a revision of the H₂ S method originally published in the FEDERAL REGISTER,
Volume 39, No. 47, dated Friday, March 8,
1974.
1.2 Applicability. This method is applicable for the determination of the hydrogen
sulfide content of fuel gas streams at petroleum refineries.
2. Range and Sensitivity
The lower limit of detection is approximately 8 mg/m3 (6 ppm). The maximum of
the range is 740 mg/m3 (520 ppm).
3. Interferences
Any compound that reduces iodine or oxidizes iodide ion will Interfere in this procedure, provided it is collected In the cadmium
sulfate Impingers. Sulfur dioxide in concentrations of up to 2,600 mg/m 3 is eliminated by the hydrogen peroxide solution.
Thiols precipitate with hydrogen sulfide. In
the absence of H₂S, only co-traces of thiols
are collected. When methane- and ethanethiols at a total level of 300 mg/m3 are
present in addition to H 2S, the results vary
from 2 percent low at an H₂S concentration
of 400 mg/m3 to 14 percent high at an H2S
concentration of 100 mg/m 3. Carbon oxysulfide at a concentration of 20 percent does not
interfere. Certain carbonyl-containing compounds react with iodine and produce recurring end points. However. acetaldehyde and
acetone at concentrations of 1 and 3 percent,
respectively, do not interfere.
Entrained hydrogen peroxide produces a
negative Interference equivalent to 100 percent of that of an equimolar quantity of hydrogen sulfide. Avoid the ejection of hydrogen peroxide into the cadmium sulfate
impingers.
4. Precision and Accuracy
Collaborative testing has shown the within-laboratory coefficient of variation to be
2.2 percent and the overall coefficient of variation to be 5 percent. The method blas was
shown to be 4.8 percent when only H₂S was
present. In the presence of the Interferences
cited in section 3, the blas was positive at
low H₂S concentration and negative at higher concentrations. At 230 mg H₂S/m 3. the
level of the compliance standard. the bias
METHOD 23 40 CFR PART 60 APPENDIXA
EC01JN92.248
METHOD 23-DETERMINATION OF POLY-
CHLORINATED DIBENZO-P-DIOXINS AND POLY-
CHLORINATED DIBENZOFURANS FROM STA-
TIONARY SOURCES
1. Applicability and Principle
1.1 Applicability. This method is applicable to the determination of polychlorinated
dibenzo-p-dioxins (PCDD's) and polychlorinated dibenzofurans (PCDF's) from stationary sources.
1.2 Principle. A sample is withdrawn from
the gas stream isokinetically and collected
in the sample probe, on a glass flber filter,
and on a packed column of adsorbent material. The sample cannot be separated into a
particle vapor fraction. The PCDD's and
PCDF's are extracted from the sample, separated by high resolution gas chromatography. and measured by high resolution
mass spectrometry.
2. Apparatus
2.1 Sampling. A schematic of the sampling train used in this method is shown in
Figure 23-1. Sealing greases may not be used
in assembling the train. The train is identical to that described in section 2.1 of Method 5 of this appendix with the following additions:
EC01JN92.249
2.1.1 Nozzle. The nozzle shall be made of
nickel. nickel-plated stainless steel, quartz,
or borosilicate glass.
2.1.2 Sample Transfer Lines. The sample
transfer lines, If needed, shall be heat traced,
heavy walled TFE (1/2 in. OD with 1/8 in. wall)
with connecting fittings that are capable of
forming leak-free. vacuum-tight connections
without using sealing greases. The line shall
be as short as possible and must be maintained at 120 °C.
2.1.1 Filter Support. Teflon or Tefloncoated wire.
2.1.2 Condenser. Glass, coll type with
compatible fittings. A schematic diagram is
shown in Figure 23-2.
2.1.3 Water Bath. Thermostatically controlled to maintain the gas temperature
exiting the condenser at < 20 °C (68 °F).
2.1.4 Adsorbent Module. Glass container
to hold the solid adsorbent. A shematic diagram is shown in Figure 23-2. Other physical
configurations of the resin trap/condenser assembly are acceptable. The connecting flttings shall form leak-free. vacuum tight
seals. No sealant greases shall be used in the
sampling train. A coarse glass frit is included to retain the adsorbent.
2.2 Sample Recovery.
2.2.1 Fitting Caps. Ground glass, Teflon
tape, or aluminum foil (Section 2.2.6) to cap
off the sample exposed sections of the train.
2.2.2 Wash Bottles. Teflon, 500-ml.
2.2.3 Probe-Liner Probe-Nozzle, and Filter-Holder Brushes. Inert bristle brushes
with precleaned stainless steel or Teflon
handles. The probe brush shall have extensions of stainless steel or Teflon, at least as
long as the probe. The brushes shall be properly sized and shaped to brush out the nozzle,
probe liner. and transfer line, if used.
EC01JN92.250
2.2.4 Filter Storage Container. Sealed filter holder, wide-mouth amber glass jar with
Teflon-lined cap, or glass petri dish.
2.2.5 Balance. Triple beam.
2.2.6 Aluminum Foil. Heavy duty, hexanerinsed.
2.2.7 Metal Storage Container. Air tight
container to store silica gel.
Pt. 60, App. A, Meth. 23
2.2.8 Graduated Cylinder. Glass, 250-ml
with 2-ml graduation.
2.2.9 Glass Sample Storage Container.
Amber glass bottle for sample glassware
washes, 500- or 1000-ml, with leak free Teflonlined caps.
2.3 Analysis.
2.3.1 Sample Container. 125- and 250-ml
flint glass bottles with Teflon-lined caps.
2.3.2 Test Tube. Glass.
2.3.3 Soxhlet Extraction Apparatus. Capable of holding 43 X 123 mm extraction thimbles.
2.3.4 Extraction Thimble. Glass, precleaned cellulosic, or glass fiber.
2.3.5 Pasteur Pipettes. For preparing liquid chromatographic columns.
2.3.6 Reacti-vials. Amber glass, 2-ml,
silanized prior to use.
2.3.7 Rotary Evaporator. Bucht/Brinkman
RF-121 or equivalent.
2.3.8 Nitrogen Evaporative Concentrator.
N-Evap Analytical Evaporator Model III or
equivalent.
2.3.9 Separatory Funnels. Glass, 2-liter.
2.3.10 Gas Chromatograph. Consisting of
the following components:
2.3.10.1 Oven. Capable of maintaining the
separation column at the proper operating
temperature ± °C and performing programmed increases in temperature at rates
of at least 40 °C/min.
2.3.10.2 Temperature Gauge. To monitor
column oven, detector, and exhaust temperatures ±1 °C.
2.3.10.3 Flow System. Gas metering system to measure sample. fuel, combustion
gas, and carrier gas flows.
2.3.10.4 Capillary Columns. A fused silica
column, 60 X 0.25 mm inside diameter (ID).
coated with DB-5 and a fused silica column.
30 m x 0.25 mm ID coated with DB-225. Other
column systems may be used provided that
the user is able to demonstrate using calibration and performance checks that the column system is able to meet the specifications of section 6.1.2.2.
2.3.11 Mass Spectrometer. Capable of routine operation at a resolution of 1:10000 with
a stability of ±5 ppm.
2.3.12 Data System. Compatible with the
mass spectrometer and capable of monitoring at least five groups of 25 ions.
2.3.13 Analytical Balance. To measure
within 0.1 mg.
3. Reagents
3.1 Sampling.
3.1.1 Filters. Glass fiber filters, without
organic binder, exhibiting at least 99.95 percent efficiency (<0.05 percent penetration) on
0.3-micron dioctyl phthalate smoke particles. The filter efficiency test shall be conducted in accordance with ASTM Standard
Method D 2986-71 (Reapproved 1978) (incorporated by reference-see $60.17).
3.1.1.1 Precleaning. All filters shall be
cleaned before their initial use. Place a glass
extraction thimble and 1 g of silica gel and a
plug of glass wool into a Soxhlet apparatus,
charge the apparatus with toluene, and
reflux for a minimum of 3 hours. Remove the
toluene and discard it. but retain the silica
gel. Place no more than 50 filters in the
thimble onto the silica gel bed and top with
the cleaned glass wool. Charge the Soxhlet
with toluene and reflux for 16 hours. After
extraction. allow the Soxhlet to cool, remove the filters. and dry them under a clean
N₂ stream. Store the filters In a glass petri
dish sealed with Teflon tape.
3.1.2 Adsorbent Resin. Amberlite XAD-2
resin. Thoroughly cleaned before Initial use.
3.1.2.1 Cleaning Procedure. This procedure
may be carried out in a giant Soxhlet extractor. An all-glass filter thimble containing an
extra-course frit is used for extraction of
XAD-2. The frit is recessed 10-15 mm above a
crenelated ring at the bottom of the thimble
to facilitate drainage. The resin must be
carefully retained in the extractor cup with
a glass wool plug and a stainless steel ring
because it floats on methylene chloride. This
process involves sequential extraction in the
following order.
40 CFR Ch. I (7-1-99 Edition)
Solvent
Procedure
Water
Initial rinse: Place resin in a
beaker, rinse once with
water, and discard. Fill with
water a second time, let
stand overnight, and discard.
Water
Extract with water for 8
hours.
Methanol
Extract for 22 hours.
Methylene Chloride
Extract for 22 hours.
Toluene
Extract for 22 hours.
3.1.2.2 Drying.
3.1.2.2.1 Drying Column. Pyrex pipe, 10.2
cm ID by 0.6 m long, with suitable retainers.
3.1.2.2.2 Procedure. The adsorbent must be
dried with clean inert gas. Liquid nitrogen
from a standard commercial liquid nitrogen
cylinder has proven to be a reliable source of
large volumes of gas free from organic contaminants. Connect the liquid nitrogen cylinder to the column by a length of cleaned
copper tubing, 0.95 cm ID, colled to pass
through a heat source. A convenient heat
source is a water-bath heated from a steam
line. The final nitrogen temperature should
only be warm to the touch and not over 40
°C. Continue flowing nitrogen through the
adsorbent until all the residual solvent is removed. The flow rate should be sufficient to
gently agitate the particles but not so excessive as the cause the particles to fracture.
3.1.2.3 Quality Control Check. The adsorbent must be checked for residual toluene.
3.1.2.3.1 Extraction. Weigh 1.0 g sample of
dried resin into a small vial, add 3 ml of toluene, cap the vial, and shake it well.
Environmental Protection Agency, EPA
3.1.2.3.2 Analysis. Inject a 2 µl sample of
the extract into a gas chromatograph operated under the following conditions:
Column: 6 ft X 1/8 in stainless steel containing 10 percent OV-101 on 100/120
Supelcoport.
Carrier Gas: Helium at a rate of 30 ml/min.
Detector: Flame ionization detector operated at a sensitivity of 4 10-" A/mV.
Injection Port Temperature: 250 °C.
Detector Temperature: 305 °C.
Oven Temperature: 30 °C for 4 min; programmed to rise at 40 °C/min until it
reaches 250 °C: return to 30 °C after 17
minutes.
Compare the results of the analysis to the
results from the reference solution. Prepare
the reference solution by injection 2.5 al of
methylene chloride into 100 ml of toluene.
This corresponds to 100 µg of methylene chloride per 8 of adsorbent. The maximum acceptable concentration is 1000 µg/g of adsorbent. If the adsorbent exceeds this level. dry-
Ing must be continued until the excess methylene chloride is removed.
3.1.2.4 Storage. The adsorbent must be
used within 4 weeks of cleaning. After clean-
Ing, it may be stored in a wide mouth amber
glass container with a Teflon-lined cap or
placed in one of the glass adsorbent modules
tightly sealed with glass stoppers. If
precleaned adsorbent is purchased in sealed
containers, it must be used within 4 weeks
after the seal is broken.
3.1.3 Glass Wool. Cleaned by sequential
immersion In three aliquots of methylene
chloride, dried in a 110 °C oven, and stored in
a methylene chloride-washed glass Jar with a
Teflon-lined screw cap.
3.1.4 Water. Deionized distilled and stored
in a methylene chloride-rinsed glass container with a Teflon-lined screw cap.
3.1.5 Silica Gel. Indicating type. 6 to 16
mesh. If previously used, dry at 175 °C (350
°F) for two hours. New silica gel may be used
as received. Alternately other types of
desiccants (equivalent or better) may be
used, subject to the approval of the Administrator.
3.1.6 Chromic Acid Cleaning Solution.
Dissolve 20 g of sodium dichromate in 15 ml
of water, and then carefully add 400 ml of
concentrated sulfuric acid.
3.2 Sample Recovery.
3.2.2 Acetone. Pesticide quality.
3.2.2 Methylene Chloride.
Pesticide
qualtity.
3.2.3 Toluene. Pesticide quality.
3.3 Analysis.
3.3.1 Potassium Hydroxide. ACS grade. 2-
percent (weight/volume) in water.
3.3.2 Sodium Sulfate. Granulated. reagent
grade. Purify prior to use by rinsing with
methylene chloride and oven drying. Store
the cleaned material in a glass container
with a Teflon-lined screw cap.
3.3.3 Sulfuric Acid. Reagent grade.
3.3.4 Sodium Hydroxide. 1.0 N. Weigh 40 g
of sodium hydroxide into a 1-liter volumetric
flask. Dilute to 1 liter with water.
3.3.5 Hexane. Pesticide grade.
3.3.6 Methylene Chloride. Pesticide grade.
3.3.7 Benzene. Pesticide Grade.
3.3.8 Ethyl Acetate.
3.3.9 Methanol. Pesticide Grade.
3.3.10 Toluene. Pesticide Grade.
3.3.11 Nonane. Pesticide Grade.
3.3.12 Cyclohexane. Pesticide Grade.
3.3.13 Basic Alumina. Activity grade 1.
100-200 mesh. Prior to use, activate the alumina by heating for 16 hours at 130 °C before
use. Store in a desiccator. Pre-activated alumina may be purchased from a supplier and
may be used as received.
3.3.14 Silica Gel. Bio-Sil A, 100-200 mesh.
Prior to use, activate the silica gel by heating for at least 30 minutes at 180 °C. After
cooling, rinse the silica gel sequentially with
methanol and methylene chloride. Heat the
rinsed silica gel at 50 °C for 10 minutes, then
increase the temperature gradually to 180 °C
over 25 minutes and maintain it at this temperature for 90 minutes. Cool at room temperature and store in a glass container with
a Teflon-lined screw cap.
3.3.15 Silica Gel Impregnated with Sulfuric Acid. Combine 100 g of silica gel with 44
g of concentrated sulfuric acid in a screw
capped glass bottle and agitate thoroughly.
Disperse the solids with a stirring rod until
a uniform mixture is obtained. Store the
mixture in a glass container with a Teflonlined screw cap.
3.3.16 Silica Gel Impregnated with Sodium
Hydroxide. Combine 39 g of 1 N sodium hydroxide with 100 g of silica gel in a screw
capped glass bottle and agitate thoroughly.
Disperse solids with a stirring rod until a
unlform mixture is obtained. Store the mixture in glass container with a Teflon-lined
screw cap.
3.3.17 Carbon/Celite. Combine 10.7 g of
AX-21 carbon with 124 g of Celite 545 in a 250-
ml glass bottle with a Teflon-lined screw
cap. Agitate the mixture thoroughly until a
uniform mixture is obtained. Store in the
glass container.
3.3.18 Nitrogen. Ultra high purity.
3.3.19 Hydrogen. Ultra high purity.
3.3.20 Internal Standard Solution. Prepare
a stock standard solution containing the
isotopically labelled PCDD's and PCDF's at
the concentrations shown in Table 1 under
the heading "Internal Standards" in ID ml of
nonane.
3.3.21 Surrogate Standard Solution. Prepare a stock standard solution containing
the isotopically labelled PCDD's and PCDF's
at the concentrations shown In Table 1 under
the heading "Surrogate Standards" in 10 ml
of nonane.
3.3.22 Recovery Standard Solution. Prepare a stock standard solution containing
the isotopically labelled PCDD's and PCDF's
Pt. 60, App. A, Meth. 23
at the concentrations shown in Table 1 under
the heading "Recovery Standards" in 10 ml
of nonane.
4. Procedure
4.1 Sampling. The complexity of this
method is such that, in order to obtain rellable results, testers should be trained and
experienced with the test procedures.
4.1.1 Pretest Preparation.
4.1.1.1 Cleaning Glassware. All glass components of the train upstream of and including the adsorbent module, shall be cleaned as
described in section 3A of the "Manual of
Analytical Methods for the Analysis of Pesticides in Human and Environmental Samples." Special care shall be devoted to the removal of residual silicone grease sealants on
ground glass connections of used glassware.
Any residue shall be removed by soaking the
glassware for several hours in a chromic acid
cleaning solution prior to cleaning as described above.
4.1.1.2 Adsorbent Trap. The traps must be
loaded in a clean area to avoid contamination. They may not be loaded in the field.
Fill a trap with 20 to 40 g of XAD-2. Follow
the XAD-2 with glass wool and tightly cap
both ends of the trap. Add 100 µl of the surrogate standard solution (section 3.3.21) to
each trap.
4.1.1.3 Sample Train. It is suggested that
all components be maintained according to
the procedure described in APTD-0576.
4.1.1.4 Silica Gel. Weigh several 200 to 300
g portions of silica gel in an air tight container to the nearest 0.5 g. Record the total
weight of the silica gel plus container, on
each container. As an alternative. the silica
gel may be weighed directly in its impinger
or sampling holder Just prior to sampling.
4.1.1.5 Filter. Check each filter against
light for irregularities and flaws or pinhole
leaks. Pack the filters flat in a clean glass
container.
4.1.2 Preliminary Determinations. Same
as section 4.1.2 of Method 5.
4.1.3 Preparation of Collection Train.
4.1.3.1 During preparation and assembly of
the sampling train, keep all train openings
where contamination can enter, sealed until
just prior to assembly or until sampling is
about to begin.
NOTE: Do not use sealant grease in assembling the train.
4.1.3.2 Place approximately 100 ml of
water in the second and third impingers,
leave the first and fourth impingers empty.
and transfer approximately 200 to 300 g of
preweighed silica gel from its container to
the fifth impinger.
4.1.3.3 Place the silica gel container in a
clean place for later use in the sample recovery. Alternatively. the weight of the silica
gel plus impinger may be determined to the
nearest 0.5 g and recorded.
4.1.3.4 Assemble the train as shown in Figure 23-1.
4.1.3.5 Turn on the adsorbent module and
condenser coil recirculating pump and begin
monitoring the adsorbent module gas entry
temperature. Ensure proper sorbent temperature gas entry temperature before proceeding and before sampling is initiated. It is
extremely important that the XAD-2 adsorbent resin temperature never exceed 50 °C because thermal decomposition will occur.
During testing. the XAD-2 temperature must
not exceed 20 °C for efficient capture of the
PCDD's and PCDF's.
4.1.4 Leak-Check Procedure. Same as
Method 5, section 4.1.4.
4.1.5 Sample Train Operation. Same as
Method 5, section 4.1.5.
4.2 Sample Recovery. Proper cleanup procedure begins as soon as the probe is removed from the stack at the end of the sampling period. Seal the nozzle end of the sampling probe with Teflon tape or aluminum
foll.
When the probe can be safely handled, wipe
off all external particulate matter near the
tip of the probe. Remove the probe from the
train and close off both ends with aluminum
foll. Seal off the Inlet to the train with Teflon tape, a ground glass cap. or aluminum
foll.
Transfer the probe and impinger assembly
to the cleanup area. This area shall be clean
and enclosed so that the chances of losing or
contaminating the sample are minimized.
Smoking. which could contaminate the sample, shall not be allowed in the cleanup area.
Inspect the train prior to and during disassembly and note any abnormal conditions,
e.g., broken filters, colored Impinger liquid,
etc. Treat the samples as follows:
4.2.1 Container No. 1. Either seal the filter
holder or carefully remove the filter from
the filter holder and place it in its identified
container. Use a pair of cleaned tweezers to
handle the filter. If It is necessary to fold the
filter. do so such that the particulate cake is
inside the fold. Carefully transfer to the container any particulate matter and filter fibers which adhere to the filter holder gasket.
by using a dry inert bristle brush and a
sharp-edged blade. Seal the container.
4.2.2 Adsorbent Module. Remove the module from the train, tightly cap both ends,
label it. cover with aluminum foil. and store
it on ice for transport to the laboratory.
4.2.3 Container No. 2. Quantitatively recover material deposited in the nozzle, probe
transfer lines, the front half of the filter
holder, and the cyclone, if used. first. by
brushing while rinsing three times each with
acetone and then. by rinsing the probe three
times with methylene chloride. Collect all
the rinses in Container No. 2.
Environmental Protection Agency, EPA
Rinse the back half of the filter holder
three times with acetone. Rinse the connecting line between the filter and the condenser three times with acetone. Soak the
connecting line with three separate portions
of methylene chloride for 5 minutes each. If
using a separate condenser and adsorbent
trap, rinse the condenser in the same manner
as the connecting line. Collect all the rinses
in Container No. 2 and mark the level of the
liquid on the container.
4.2.4 Container No. 3. Repeat the methylene chloride-rinsing described in Section
4.2.3 using toluene as the rinse solvent. Collect the rinses in Container No. 3 and mark
the level of the liquid on the container.
4.2.5 Impinger Water. Measure the liquid
in the first three impingers to within +1 ml
by using a graduated cylinder or by weighing
it to within ±0.5 g by using a balance. Record
the volume or weight of liquid present. This
information is required to calculate the
moisture content of the effluent gas.
Discard the liquid after measuring and recording the volume or weight.
4.2.7 Silica Gel. Note the color of the indicating silica gel to determine if It has been
completely spent and make a mention of its
condition. Transfer the silica gel from the
fifth impinger to its original container and
seal.
5. Analysis
All glassware shall be cleaned as described
in section 3A of the "Manual of Analytical
Methods for the Analysis of Pesticides in
Human and Environmental Samples." All
samples must be extracted within 30 days of
collection and analyzed within 45 days of extraction.
5.1 Sample Extraction.
5.1.1 Extraction System. Place an extraction thimble (section 2.3.4). 1 g of silica gel,
and a plug of glass wool into the Soxhlet apparatus, charge the apparatus with toluene,
and reflux for a minimum of 3 hours. Remove
the toluene and discard it, but retain the
silica gel. Remove the extraction thimble
from the extraction system and place it in a
glass beaker to catch the solvent rinses.
5.1.2 Container No. 1 (Filter). Transfer the
contents directly to the glass thimble of the
extraction system and extract them simultaneously with the XAD-2 resin.
5.1.3 Adsorbent Cartridge. Suspend the adsorbent module directly over the extraction
thimble in the beaker (See section 5.1.1). The
glass frit of the module should be in the up
position. Using a Teflon squeeze bottle containing toluene, flush the XAD-2 into the
thimble onto the bed of cleaned silica gel.
Thoroughly rinse the glass module catching
the rinsings in the beaker containing the
thimble. If the resin is wet, effective extraction can be accomplished by loosely packing
the resin in the thimble. Add the XAD-2
glass wool plug into the thimble.
5.1.4 Container No. 2 (Acetone and Methylene Chloride). Concentrate the sample to a
volume of about 1-5 ml using the rotary
evaporator apparatus, at a temperature of
less than 37 °C. Rinse the sample container
three times with small portions of methylene chloride and add these to the concentrated solution and concentrate further
to near dryness. This residue contains particulate matter removed In the rinse of the
train probe and nozzle. Add the concentrate
to the filter and the XAD-2 resin in the
Soxhlet apparatus described in section 5.1.1.
5.1.5 Extraction. Add 100 µ1 of the internal
standard solution (Section 3.3.20) to the extraction thimble containing the contents of
the adsorbent cartridge, the contents of Container No. 1. and the concentrate from section 5.1.4. Cover the contents of the extraction thimble with the cleaned glass wool
plug to prevent the XAD-2 resin from floating into the solvent reservoir of the extractor. Place the thimble in the extractor, and
add the toluene contained in the beaker to
the solvent reservoir. Pour additional toluene to fill the reservoir approximately 2/3
full. Add Teflon boiling chips and assemble
the apparatus. Adjust the heat source to
cause the extractor to cycle three times per
hour. Extract the sample for 16 hours. After
extraction. allow the Soxhlet to cool. Transfer the toluene extract and three 10-ml rinses
to the rotary evaporator. Concentrate the
extract to approximately 10 ml. At this point
the analyst may choose to split the sample
in half. If so, split the sample, store one half
for future use, and analyze the other according to the procedures in sections 5.2 and 5.3.
In either case, use a nitrogen evaporative
concentrator to reduce the volume of the
sample being analyzed to near dryness. Dissolve the residue in 5 ml of hexane.
5.1.6 Container No. 3 (Toluene Rinse). Add
100 µl of the Internal Standard solution (section 3.3.2) to the contents of the container.
Concentrate the sample to a volume of about
1-5 ml using the rotary evaporator apparatus
at a temperature of less than 37 °C. Rinse the
sample container apparatus at a temperature
of less than 37 °C. Rinse the sample container
three times with small portions of toluene
and add these to the concentrated solution
and concentrate further to near dryness.
Analyze the extract separately according to
the procedures in sections 5.2 and 5.3, but
concentrate the solution In a rotary evaporator apparatus rather than a nitrogen evaporative concentrator.
5.2 Sample Cleanup and Fractionation.
5.2.1 Silica Gel Column. Pack one end of a
glass column, 20 mm X 230 mm, with glass
wool. Add in sequence. 1 g silica gel. 2 g of
sodium hydroxide Impregnated silica gel, 1 g
silica gel. 4 g of acid-modified silica gel. and
I g of silica gel. Wash the column with 30 ml
of hexane and discard it. Add the sample extract, dissolved in 5 ml of hexane to the column with two additional 5-ml rinses. Elute
the column with an additional 90 ml of
hexane and retain the entire eluate. Concentrate this solution to a volume of about 1
ml using the nitrogen evaporative concentrator (section 2.3.7).
5.2.2 Basic Alumina Column. Shorten a 25-
ml disposable Pasteur pipette to about 16 ml.
Pack the lower section with glass wool and
12 g of basic alumina. Transfer the concentrated extract from the silica gel column
to the top of the basic alumina column and
elute the column sequentially with 120 ml of
0.5 percent methylene chloride in hexane followed by 120 ml of 35 percent methylene
chloride in hexane. Discard the first 120 ml of
eluate. Collect the second 120 ml of eluate
and concentrate it to about 0.5 ml using the
nitrogen evaporative concentrator.
5.2.3 AX-21 Carbon/Celite 545 Column. Remove the botton 0.5 in. from the tip of a 9-ml
disposable Pasteur plpette. Insert a glass
fiber filter disk in the top of the pipette 2.5
cm from the constriction. Add sufficient carbon/celite mixture to form a 2 cm column.
Top with a glass wool plug. In some cases
AX-21 carbon fines may wash through the
glass wool plug and enter the sample. This
may be prevented by adding a celite plug to
the exit end of the column. Rinse the column
in sequence with 2 ml of 50 percent benzene
in ethyl acetate, 1 ml of 50 percent methylene chloride in cyclohexane, and 2 ml of
hexane. Discard these rinses. Transfer the
concentrate in 1 ml of hexane from the basic
alumina column to the carbon/celite column
along with 1 ml of hexane rinse. Elute the
column sequentially with 2 ml of 50 percent
methylene chloride in hexane and 2 ml of 50
percent benzene in ethyl acetate and discard
these eluates. Invert the column and elute in
the reverse direction with 13 ml of toluene.
Collect this eluate. Concentrate the eluate
in a rotary evaporator at 50 °C to about 1 ml.
Transfer the concentrate to a Reacti-vial
using a toluene rinse and concentrate to a
volume of 200 µl using a stream of N2. Store
extracts at room temperature. shielded from
light. until the analysis is performed.
5.3 Analysis. Analyze the sample with a
gas chromatograph coupled to a mass spectrometer (GC/MS) using the instrumental parameters in sections 5.3.1 and 5.3.2. Immediately prior to analysis. add a 20 µl aliquot
of the Recovery Standard solution from
Table 1 to each sample. A 2 μl aliquot of the
extract is injected Into the GC. Sample extracts are first analyzed using the DB-5 capillary column to determine the concentration of each isomer of PCDD's and PCDF's
(tetra-through octa-). If tetra-chlorinated
dibenzofurans are detected in this analysis.
then analyze another aliquot of the sample
In a separate run, using the DB-225 column
to measure the 2,3,7.8 tetra-chloro
dibenzofuran Isomer. Other column systems
may be used. provided that the user is able
to demonstrate using calibration and performance checks that the column system is
able to meet the specifications of section
6.1.2.2.
5.3.1 Gas Chromatograph Operating Conditions.
5.3.1.1 Injector. Configured for capillary
column, splitless, 250 °C.
5.3.1.2 Carrier Gas. Helium, 1-2 ml/min.
5.3.1.3 Oven. Initially at 150 °C. Raise by
at least 40 °C/min to 190 °C and then at 3 °C/
min up to 300 °C.
5.3.2 High Resolution Mass Spectrometer.
5.3.2.1 Resolution. 10000 m/e.
5.3.2.2 Ionization Mode. Electron impact.
5.3.2.3 Source Temperature 250 °C.
5.3.2.4 Monitoring Mode. Selected ion
monitoring. A list of the various ions to be
monitored is summarized in Table 3.
5.3.2.5 Identification Criteria. The following identification criteria shall be used
for the characterization of polychlorinated
dibenzodioxins and dibenzofurans.
1. The integrated ion-abundance ratio (M/
M+2 or M+2/M+4) shall be within 15 percent
of the theoretical value. The acceptable ionabundance ratio ranges for the identification
of chlorine-containing compounds are given
in Table 4.
2. The retention time for the analytes
must be within 3 seconds of the corresponding 13C-labeled internal standard,
surrogate or alternate standard.
3. The monitored ions, shown in Table 3 for
a given analyte. shall reach their maximum
within 2 seconds of each other.
4. The identification of specific Isomers
that do not have corresponding 13C-labeled
standards is done by comparison of the relative retention time (RRT) of the analyte to
the nearest Internal standard retention time
with reference (i.e., within 0.005 RRT units)
to the comparable RRT's found in the continuing calibration.
5. The signal to noise ratio for all monitored ions must be greater than 2.5.
6. The confirmation of 2, 3, 7. 8-TCDD and
2, 3. 7. 8-TCDF shall satisfy all of the above
identification criteria.
7. For the identification of PCDF's, no signal may be found in the corresponding
PCDPE channels.
5.3.2.6 Quantification. The peak areas for
the two lons monitored for each analyte are
summed to yield the total response for each
analyte. Each Internal standard is used to
quantify the Indigenous PCDD's or PCDF's
in its homologous series. For example, the
13C₁₂-2,3,7,8-tetra chlorinated dibenzodioxin
is used to calculate the concentrations of all
other tetra chlorinated Isomers. Recoveries
of the tetra- and penta- Internal standards
are calculated using the
Recoveries of the hexa- through octa- Internal standards are calculated using 13C12
Environmental Protection Agency, EPA
1,2,3,7,8,9-HxCDD. Recoveries of the surrogate standards are calculated using the corresponding homolog from the internal standard.
6. Calibration
Same as Method 5 with the following additions.
6.1 GC/MS System.
6.1.1 Initial Calibration. Calibrate the GC/
MS system using the set of five standards
shown in Table 2. The relative standard deviation for the mean response factor from each
of the unlabeled analytes (Table 2) and of the
Internal, surrogate, and alternate standards
shall be less than or equal to the values in
Table 5. The signal to noise ratio for the GC
signal present in every selected ion current
profile shall be greater than or equal to 2.5.
The ion abundance ratios shall be within the
control limits in Table 4.
6.1.2 Daily Performance Check.
6.1.2.1 Calibration Check. Inject on µl of
solution Number 3 from Table 2. Calculate
the relative response factor (RRF) for each
compound and compare each RRF to the corresponding mean RRF obtained during the
initial calibration. The analyzer performance is acceptable If the measured RRF's for
the labeled and unlabeled compounds for the
daily run are within the limits of the mean
values shown in Table 5. In addition, the ionabundance ratios shall be within the allowable control limits shown in Table 4.
6.1.2.2 Column Separation Check. Inject a
solution of a mixture of PCDD's and PCDF's
that documents resolution between 2,3,7,8-
TCDD and other TCDD isomers. Resolution
is defined as a valley between peaks that is
less than 25 percent of the lower of the two
peaks. Identify and record the retention time
windows for each homologous series.
Perform a similar resolution check on the
confirmation column to document the resolution between 2,3,7.8 TCDF and other TCDF
isomers.
6.2 Lock Channels. Set mass spectrometer
lock channels as specified in Table 3. Monitor the quality control check channels specified in Table 3 to verify instrument stability
during the analysis.
7. Quality Control
7.1 Sampling Train Collection Efficiency
Check. Add 100 µl of the surrogate standards
in Table 1 to the absorbent cartridge of each
train before collecting the field samples.
7.2 Internal Standard Percent Recoveries.
A group of nine carbon labeled PCDD's and
PCDF's representing, the tetra-through
octachlorinated homologues, is added to
every sample prior to extraction. The role of
the internal standards is to quantify the native PCDD's and PCDF's present in the sample as well as to determine the overall method efficiency. Recoveries of the internal
standards must be between 40 to 130 percent
for the tetra-through hexachlorinated compounds while the range is 25 to 130 percent
for the higher hepta- and octachlorinated
homologues.
7.3 Surrogate Recoveries. The five surrogate compounds In Table 2 are added to the
resin in the adsorbent sampling cartridge before the sample is collected. The surrogate
recoveries are measured relative to the internal standards and are a measure of collection efficiency. They are not used to measure
native PCDD's and PCDF's. All recoveries
shall be between 70 and 130 percent. Poor recoveries for all the surrogates may be an indication of breakthrough in the sampling
train. If the recovery of all standards is
below 70 percent. the sampling runs must be
repeated. As an alternative, the sampling
runs do not have to be repeated if the final
results are divided by the fraction of surrogate recovery. Poor recoveries of isolated
surrogate compounds should not be grounds
for rejecting an entire set of the samples.
7.4 Toluene QA Rinse. Report the results
of the toluene QA rinse separately from the
total sample catch. Do not add it to the total
sample.
8. Quality Assurance
8.1 Applicability. When the method is
used to analyze samples to demonstrate compliance with a source emission regulation. an
audit sample must be analyzed, subject to
availability.
8.2 Audit Procedure. Analyze an audit
sample with each set of compliance samples.
The audit sample contains tetra through
octa isomers of PCDD and PCDF. Concurrently, analyze the audit sample and a set of
compliance samples in the same manner to
evaluate the technique of the analyst and
the standards preparation. The same analyst, analytical reagents, and analytical system shall be used both for the compliance
samples and the EPA audit sample.
8.3 Audit Sample Availability. Audit samples will be supplied only to enforcement
agencies for compliance tests. The availability of audit samples may be obtained by
writing: Source Test Audit Coordinator (MD-
77B). Quality Assurance Division. Atmospheric Research and Exposure Assessment
Laboratory. U.S. Environmental Protection
Agency, Research Triangle Park, NC 27711,
or by calling the Source Test Audit Coordinator (STAC) at (919) 541-7834. The request
for the audit sample must be made at least
30 days prior to the scheduled compliance
sample analysis.
8.4 Audit Results. Calculate the audit
sample concentration according to the calculation procedure described in the audit instructions included with the audit sample.
Fill in the audit sample concentration and
the analyst's name on the audit response
form included with the audit Instructions.
Pt. 60, App. A, Meth. 23
40 CFR Ch. I (7-1-99 Edition)
Send one copy to the EPA Regional Office or
the appropriate enforcement agency and a
second copy to the STAC. The EPA Regional
office or the appropriate enforcement agency
will report the results of the audit to the
laboratory being audited. Include this response with the results of the compliance
samples in relevant reports to the EPA Regional Office or the appropriate enforcement
agency.
9. Calculations
Same as Method 5. section 6 with the following additions.
9.1 Nomenclature.
A=Integrated ion current of the noise at
the retention time of the analyte.
A*α=Integrated ion current of the two lons
characteristic of the internal standard 1
in the calibration standard.
A lon current of the two ions
characteristic of compound i In the Jth
calibration standard.
A"aj=Integrated ion current of the two ions
characteristic of the internal standard 1
in the Jth callbration standard.
A ion current of the two ions
characteristic of surrogate compound 1 in
the calibration standard.
A,=Integrated ion current of the two ions
characteristic of compound 1 in the sample.
A'₁=Integrated ion current of the two ions
characteristic of internal standard i in
the sample.
An=Integrated ion current of the two lons
characteristic of the recovery standard.
As=Integrated ion current of the two lons
characteristic of surrogate compound 1 in
the sample.
C=Concentration of PCDD or PCDF 1 in
the sample, pg/M 3.
CT=Total concentration of PCDD's or
PCDF's in the sample, pg/M 3.
ma=Mass of compound 1 in the calibration
standard injected into the analyzer, pg.
mm-Mass of recovery standard in the callbration standard injected into the analyzer. pg.
m=Mass of surrogate compound in the
calibration standard, pg.
RRF;=Relative response factor.
RRF₂=Recovery standard response factor.
RRF,=Surrogate compound response factor.
9.2 Average Relative Response Factor.
EC16NO91.219
EC16NO91.220
9.3 Concentration of the PCDD's and
PCDF's.
9.4 Recovery Standard Response Factor.
EC16NO91.221
9.5 Recovery of Internal Standards (R*).
EC16NO91.222
9.6 Surrogate Compound Response Factor.
EC16NO91.223
9.7 Recovery of Surrogate Compounds
(R,).
EC16NO91.224
9.8 Minimum Detectable Limit (MDL).
EC16NO91.225
9.9 Total Concentration of PCDD's and
PCDF's in the Sample.
EC16NO91.226
Any PCDD's or PCDF's that are reported
as nondetected (below the MDL) shall be
counted as zero for the purpose of calculating the total concentration of PCDD's and
PCDF's in the sample.
10. Bibliography
1. American Society of Mechanical Engineers. Sampling for the Determination of
Chlorinated Organic Compounds in Stack
Emissions. Prepared for U.S. Department of
Energy and U.S. Environmental Protection
Agency. Washington DC. December 1984. 25 p.
2. American Society of Mechanical Englneers. Analytical Procedures to Assay Stack
Effluent Samples and Residual Combustion
Products for Polychlorinated Dibenzo-p-
Dioxins (PCDD) and Polychlorinated Dibenzofurans (PCDF). Prepared for the U.S. Department of Energy and U.S. Environmental
Protection Agency. Washington, DC. December 1984. 23 p.
3. Thompson, J. R. (ed.). Analysis of Pesticide Residues in Human and Environmental Samples. U.S. Environmental Protection Agency. Research Triangle Park, NC.
1974.
Environmental Protection Agency, EPA
Pt. 60, App. A, Meth. 23
4. Triangle Laboratories. Case Study: Analysis of Samples for the Presence of Tetra
Through Octachloro-p-Dibenzodioxins and
Dibenzofurans. Research Triangle Park. NC.
1988. 26 p.
5. U.S. Environmental Protection Agency.
Method 8290-The Analysis of Polychlorinated Dibenzo-p-dioxin and Polychlorinated Dibenzofurans by High-Resolution Gas Chromotography/High-Resolution
Mass Spectrometry. In: Test Methods for
Evaluating Solid Waste. Washington, DC.
SW-846.
TABLE 1-COMPOSITION OF THE SAMPLE FOR-
TIFICATION AND RECOVERY STANDARDS SOLU-
TIONS
Analyte
Concentration (pg/ul)
Internal Standards:
13 C₁₂-2,3,7,8-TCDD
100
13 C₁₂-1,2,3,7,8-PeCDD
100
13 C₁₂-1,2,3,6,7,8-HxCDD
100
13 C₁₂-1,2,3,4,6,7,8-HpCDD
100
13 C12-OCDD
100
13 C₁₂-2,3,7,8-TCDF
100
13 C₁₂-1,2,3,7,8-PeCDF
100
13
100
13 Ct-1,2,3,4,6,7,8-HpCDF
100
Surrogate Standards:
37 Cl-2,3.7,8-TCDD
100
TABLE 1-COMPOSITION OF THE SAMPLE FOR-
TIFICATION AND RECOVERY STANDARDS SOLU-
TIONS-Continued
Concentra-
Analyte
tion (pg/µl)
13 C₁₂-1,2,3,4,7,8-HxCDD
100
13 C₁₂-2,3,4,7,8-PeCDF
100
13 Cr₂1,2,3,4,7,8-HxCDF
100
13 C₁₂₁,2,3,4,7,8,9-HpCDF
100
Recovery Standards:
13 C₁₂₁,2,3,4-TCDD
500
13 Ci-1,2,3,7,8,9-HxCDD
500
TABLE 2-COMPOSITION OF THE INITIAL
CALIBRATION SOLUTIONS
TABLE 3-ELEMENTAL COMPOSITIONS AND EXACT MASSES OF THE IONS MONITORED BY HIGH
RESOLUTION MASS SPECTROMETRY FOR PCDD's AND PCDF's
Concentrations (pg/uL)
Compound
Solution No.
1
2
3
4
5
Alternate Standard:
13 C₁₂-1,2,3,7,8,9-
HxCDF
2.5
5
25
250
500
Recovery Standards:
13 Ctr1,2,3,4-TCDD
100
100
100
100
100
13 C₁₂₁,2,3,7,8,9-
HxCDD
100
100
100
100
100
Descriptor
No.
Accurate mass
lon type
Elemental composition
Analyte
2
292.9825
LOCK
CF,,
PFK
303.9016
M
C12H2CLO
TCDF
305.8987
M+2
TCDF
315.9419
M
TCDF (S)
317.9389
M+2
TCDF (S)
319.8965
M
C12H45CIO₂
TCDD
321.8936
M+2
TCDD
327.8847
M
C12H27CLO₂
TCDD (S)
330.9792
QC
CF13
PFK
331.9368
M
TCDD (S)
333.9339
M+2
TCDD (S)
339.8597
M+2
PECDF
341.8567
M+4
PeCDF
351.9000
M+2
PeCDF (S)
353.8970
M+4
PeCDF (S)
355.8546
M+2
PeCDD
357.8516
M+4
PeCDD
367.8949
M+2
PeCDD (S)
369.8919
M+4
"CuH3"Ch" Cl₂O₂
PeCDD (S)
375.8364
M+2
HxCDPE
409.7974
M+2
HpCPDE
3
373.8208
M+2
HxCDF
375.8178
M+4
HxCDF
383.8639
M
HxCDF (S)
385.8810
M+2
HxCDF (S)
389.8157
M+2
HxCDD
391.8127
M+4
HxCDD
392.9760
LOCK
CF15
PFK
401.8559
M+2
HxCDD (S)
403.8529
M+4
HxCDD (S)
445.7555
M+4
OCDPE
430.9729
QC
CF,,
PFK
4
407.7818
M+2
C
HpCDF
409.7789
M+4
HpCDF
Pt. 60, App. A, Meth. 24
TABLE 3-ELEMENTAL COMPOSITIONS AND EXACT MASSES OF THE IONS MONITORED BY HIGH
RESOLUTION MASS SPECTROMETRY FOR PCDD's AND PCDF's-Continued
Descriptor
No.
Accurate mass
Ion type
Elemental composition
Analyte
417.8253
M
13C12H35Cl70
HpCDF (S)
419.8220
M+2
HpCDF (S)
423.7766
M+2
HpCDD
425.7737
M+4
HpCDD
435.8169
M+2
HpCDD (S)
437.8140
M+4
HpCDD (S)
479.7165
M+4
NCPDE
430.9729
LOCK
CF17
PFK
441.7428
M+2
C123Ch3TCIO
OCDF
443.7399
M+4
OCDF
457.7377
M+2
OCDD
459.7348
M+4
OCDD
469.7779
M+2
OCDD (S)
471.7750
M+4
OCDD (S)
513.6775
M+4
DCDPE
442.9728
QC
C1057
PFK
(a) The following nuclidic masses were used:
H = 1.007825
C - 12.000000
13C = 13.003355
F = 18.9984
O = 15.994915
35CI = 34.968853
37CI = 36.965903
S = Labeled Standard
QC = Ion selected for monitoring instrument stability during the GC/MS analysis.
TABLE 4-ACCEPTABLE RANGES FOR ION-
ABUNDANCE RATIOS OF PCDD's AND PCDF's
No. of
Theo-
Control limits
chiorine
Ion type
retical
atoms
ratio
Lower
Upper
4
M/M+2
0.77
0.65
0.89
5
M+2/M+4
1.55
1.32
1.78
6
M+2/M+4
1.24
1.05
1.43
6=
M/M+2
0.51
0.43
0.59
7b
M/M+2
0.44
0.37
0.51
7
M+2/M+4
1.04
0.88
1.20
8
M+2/M+4
0.89
0.76
1.02
*Used only for 13C-HxCDF.
Used only for 13C-HpCDF.
TABLE 5-MINIMUM REQUIREMENTS FOR INITIAL
AND DAILY CALIBRATION RESPONSE FACTORS
Relative response factors
Compound
Initial calibra-
Daily calibration RSD
tion % difference
Unlabeled
Analytes:
2,3,7,8-TCDD
25
25
2,3,7,8-TCDF
25
25
1,2,3,7,8-PeCDD
25
25
1,2,3,7,8-PeCDF
25
25
2,3,4,7,8-PeCDF
25
25
1,2,4,5,7,8-HxCDD
26
25
1,2,3,6,7,8-HxCDD
25
25
1,2,3,7,8,9-HxCDD
25
25
1,2,3,4,7,8-HxCDF
25
25
1,2,3,6,7,8-HxCDF
25
25
1,2,3,7,8,9-HxCDF
25
25
2,3,4,6,7,8-HxCDF
25
25
1,2,3,4,6,7,8-HpCDD
25
25
1,2,3,4,6,7,8-HpCDF
25
25
OCDD
25
25
TABLE 5-MINIMUM REQUIREMENTS FOR INITIAL
AND DAILY CALIBRATION RESPONSE FAC-
TORS-Continued
Relative response factors
Compound
Initial calibra-
Daily calibration RSD
tion % difference
OCDF
30
30
Internal
Standards:
13C₁ᵣ⁻2,3,7,8-TCDD
25
25
³C₁-1,2,3,7,8-PeCDD
30
30
25
25
13C₁₂-1,2,3,4,6,7,8-
HpCDD
30
30
13C₁₂-OCDD
30
30
13C₁₂-2,3,7,8-TCDF
30
30
13C₁₃⁻1,2,3,7,8-PeCDF
30
30
30
30
13C₁₂-1,2,3,4,6,7,8-
HpCDF
30
30
Surrogate
Standards:
"CL-2,3,7,8-TCDD
25
25
:
25
25
25
25
25
25
13C₁₁⁻1,2,3,4,7,8,9-
HpCDF
25
25
Alternate
Standard:
25
25
METHOD 24-DETERMINATION OF VOLATILE
MATTER CONTENT. WATER CONTENT. DEN-
SITY, VOLUME SOLIDS, AND WEIGHT SOLIDS
OF SURFACE COATINGS
1. Applicability and Principle
METHOD 26 40 CFR PART 60 APPENDIXA
Pt. 60, App. A, Meth. 26
40 CFR Ch. I (7-1-99 Edition)
6. Operational Checks and Calibration
Maintain a record of performance of each
item.
6.1 Use the procedures in Section 6.1.1 to
calibrate the headspace analyzer and FID
and check for linearity before the system is
first placed in operation, after any shutdown
longer than 6 months, and after any modification of the system.
6.1.1 Calibration and Linearity. Use the
procedures in Section 6.2.1 of Method 18 of
Part 60, Appendix A. to prepare the standards and calibrate the flowmeters, using propane as the standard gas. Fill the calibration
standard vials halfway (+5 percent) with deionized water. Purge and fill the airspace
with calibration standard. Prepare a minimum of three calibration standards in triplicate at concentrations that will bracket
the applicable cutoff. For a cutoff of 5.2 kPa.
prepare nominal concentrations of 30,000,
50,000. and 70,000 ppm as propane. For a cutoff of 27.6 kPa, prepare nominal concentrations of 200,000. 300,000, and 400,000 ppm as
propane.
6.1.1.1 Use the procedures in Section 5.2.3
to measure the FID response of each standard. Use a linear regression analysis to calculate the values for the slope (k) and the y-
Intercept (b). Use the procedures in Sections
7.2 and 7.3 to test the calibration and the linearity.
6.1.2 Daily FID Calibration Check. Check
the calibration at the beginning and at the
end of the daily runs by using the following
procedures. Prepare two calibration standards at the nominal cutoff concentration
using the procedures in Section 6.1.1. Place
one at the beginning and one at the end of
the daily run. Measure the FID response of
the daily callbration standard and use the
values for k and b from the most recent calibration to calculate the concentration of the
daily standard. Use an equation similar to
25E-2 to calculate the percent difference between the daily standard and C R. If the difference is within 5 percent. then the previous
values for k and b may be used. Otherwise,
use the procedures in Section 6.1.1 to recalibrate the FID.
7. Calculations
7.1 Nomenclature.
A at Measurement of the area under the response curve, counts.
b = y-intercept of the linear regression line.
C. = Measured vapor phase organic concentration of sample. ppm as propane.
Cms - Average measured vapor phase organic
concentration of standard. ppm as propane.
Cm - Measured vapor phase organic concentration of standard, ppm as propane.
C. = Calculated standard concentration, ppm
as propane.
k = Slope of the linear regression line.
Phar = Atmospheric pressure at analysis conditions, mm Hg (in. Hg).
P* = Organic vapor pressure in the sample,
kPa (psi).
B = 1.333 X 10-7 kPa/[(mm Hg)(ppm)]. (4.91 X
10-7 psi/[(in. Hg)(ppm)])
7.2 Linearity. Use the following equation
to calculate the measured standard concentration for each standard vial.
Cm=kA+b Eq. 25E-1
7.2.1 Calculate the average measured
standard concentration (Cma) for each set of
triplicate standards and use the following
equation to calculate the percent difference
(PD) between Cms and Ca.
ER06DE94.001
The instrument linearity is acceptable if
the percent difference is within five for each
standard.
7.3 Relative Standard Deviation (RSD).
Use the following equation to calculate the
RSD for each triplicate set of standards.
ER06DE94.002
The calibration is acceptable if the RSD is
within five for each standard concentration.
7.4 Concentration of organics in the
headspace. Use the following equation to calculate the concentration of vapor phase
organics in each sample.
C.=kA+b Eq. 25E-4
7.5 Vapor Pressure of Organics in the
Headspace Sample. Use the following equation to calculate the vapor pressure of
organics in the sample.
P* = ß Phar C. Eq. 25E-5
METHOD 26-DETERMINATION OF HYDROGEN
CHLORIDE EMISSIONS FROM STATIONARY
SOURCES
J. Applicability, Principle, Interferences,
Precision, Blas, and Stability
1.1 Applicability. This method is applicable for determining emissions of hydrogen
halides (HX) [hydrogen chloride (HCI), hydrogen bromide (HBr). and hydrogen fluoride
(HF)] and halogens (X2) [chlorine (CI 2) and
bromine (Br₂)] from stationary sources.
Sources, such as those controlled by wet
scrubbers, that emit acid particulate matter
must be sampled using Method 26A.
NOTE: Mention of trade names or specific
products does not constitute endorsement by
the Environmental Protection Agency.]
1.2 Principle. An integrated sample is extracted from the source and passed through
Environmental Protection Agency, EPA
a prepurged heated probe and filter into dilute sulfuric acid and dilute sodium hydrox-
Ide solutions which collect the gaseous hydrogen halides and halogens, respectively.
The filter collects other particulate matter
including halide salts. The hydrogen halides
are solubilized in the acidic solution and
form chloride (CI- ), bromide (Br⁻). and fluoride (F⁻) ions. The halogens have a very low
solubility in the acidic solution and pass
through to the alkaline solution where they
are hydrolyzed to form a proton (H-), the halide ion, and the hypohalous acid (HCIO or
HBrO). Sodium thiosulfate is added In excess
to the alkaline solution to assure reaction
with the hypohalous acid to form a second
halide ion such that 2 halide ions are formed
for each molecule of halogen gas. The halide
lons in the separate solutions are measured
by ion chromatography (IC).
1.3 Interferences. Volatile materials, such
as chlorine dioxide (C1O2) and ammonium
chloride (NH4CI), which produce halide ions
upon dissolution during sampling are potential Interferents. Interferents for the halide
measurements are the halogen gases which
disproportionate to a hydrogen halide and a
hydrohalous acid upon dissolution in water.
However, the use of acidic rather than neutral or basic solutions for collection of the
hydrogen halides greatly reduces the dissolution of any halogens passing through this SOlution. The simultaneous presence of HBr
and CL2 may cause a positive blas in the
HCL result with a corresponding negative
bias in the Cl₂ result as well as affecting the
HBr/Br₂ split. High concentrations of nitrogen oxides (NOx) may produce sufficient nitrate (NO₃ -) to interfere with measurements
of very low Br - levels.
1.4 Precision and Bias. The within-laboratory relative standard deviations are 6.2 and
3.2 percent at HCI concentrations of 3.9 and
15.3 ppm, respectively. The method does not
exhibit a bias to Cl₂ when sampling at concentrations less than 50 ppm.
1.5 Sample Stability. The collected Clsamples can be stored for up to 4 weeks.
1.6 Detection Limit. The analytical detection limit for Cl- is 0.1 µg/ml. Detection lim-
Its for the other analyses should be similar.
2. Apparatus
2.1 Sampling. The sampling train is
shown in Figure 26-1. and component parts
are discussed below.
2.1.1 Probe. Borsilicate glass, approximately 3/8-in. (9-mm) I.D. with a heating system to prevent moisture condensation. A
Teflon-glass filter in a mat configuration
shall be installed behind the probe to remove
particulate matter from the gas stream (see
section 2.1.5). A glass wool plug should not be
used to remove particulate matter since a
negative blas in the data could result.
2.1.2 Three-Way Stopcock. A borosilicate
glass three-way stopcock with a heating system to prevent molsturecondensation. The
heated stopcock should connect to the outlet
of the heated filter and the inlet of the first
impinger. The heating system shall be capable of preventing condensation up to the
inlet of the first Impinger. Silicone grease
may be used, if necessary. to prevent leakage.
ER22AP94.011
2.1.3 Impingers. Four 30-ml midget
impingers with leak-free glass connectors.
Silicone grease may be used, If necessary, to
prevent leakage. For sampling at high moisture sources or for sampling times greater
than 1 hour, a midget impinger with a shortened stem (such that the gas sample does not
bubble through the collected condensate)
should be used in front of the first impinger.
2.1.4 Drying Tube or Impinger. Tube or
impinger, of Mae West design. filled with 6-
to 16-mesh indicating type silica gel, or
equivalent, to dry the gas sample and to protect the dry gas meter and pump. If the sillca gel has been used previously, dry at 175 °C
(350 °F) for 2 hours. New silica gel may be
used as received. Alternatively, other types
of desiccants (equivalent or better) may be
used.
2.1.5 Filter. When the stack gas temperature exceeds 210 °C (410 °F) and the HCI concentration is greater than 20 ppm, a quartzfiber filter may be used.
2.1.6 Filter Holder and Support. The filter
holder should be made of Teflon or quartz.
The filter support shall be made of Teflon.
All-Teflon filter holders and supports are
available from Savillex Corp., 5325 Hwy 101.
Minnetonka, MN 55345.
2.1.7 Sample Line. Leak-free, with compatible fittings to connect the last Impinger
to the needle valve.
2.1.8 Rate Meter. Rotameter, or equivalent, capable of measuring flow rate to within 2 percent of the selected flow rate of 2 11-
ters/min.
2.1.9 Purge Pump, Purge Line, Drying
Tube, Needle Valve, and Rate Meter. Pump
capable of purging the sampling probe at 2 11-
ters/min. with drying tube, filled with silica
gel or equivalent, to protect pump, and a
rate meter capable of measuring 0 to 5 liters/
min.
2.1.10 Stopcock Grease, Valve, Pump. Volume Meter, Barometer. and Vacuum Gauge.
Same as in Method 6. Sections 2.1.4, 2.1.7,
2.1.8, 2.1.10, 2.1.11, and 2.1.12.
2.1.11 Temperature Measuring Devices.
Temperature measuring device to monitor
the temperature of the probe and a thermometer or other temperature measuring
device to monitor the temperature of the
sampling system from the outlet of the probe
to the Inlet of the first Impinger.
2.1.12 Ice Water Bath. To minimize loss of
absorbing solution.
2.2 Sample Recovery.
2.2.1 Wash Bottles. Polyethylene or glass,
500-ml or larger, two.
2.2.2 Storage Bottles. 100- or 250-ml. highdensity polyethylene bottles with Teflon ®
screw cap liners to store Impinger samples.
2.3 Sample Preparation and Analysis. The
materials required for volumetric dilution
and chromatographic analysis of samples are
described below.
2.3.1 Volumetric Flasks. Class A, 100-ml
size.
2.3.2 Volumetric Pipets. Class A. assortment. To dilute samples Into the calibration
range of the instrument.
Environmental Protection Agency, EPA
2.3.3 Ion Chromatograph. Suppressed or
nonsuppressed. with a conductivity detector
and electronic Integrator operating in the
peak area mode. Other detectors, strip chart
recorders, and peak height measurements
may be used.
3. Reagents
Unless otherwise indicated, all reagents
must conform to the specifications established by the Committee on Analytical Reagents of the American Chemical Society
(ACS reagent grade). When such specifications are not available. the best available
grade shall be used.
3.1 Sampling.
3.1.1 Water. Delonized, distilled water
that conforms to ASTM Specification D 1193-
77, Type 3.
3.1.2 Acidic Absorbing solution, 0.1 N Sulfuric Acid (H₂SO₄). To prepare 100 ml of the
absorbing solution for the front impinger
pair, slowly add 0.28 ml of concentrated
H2SO4 to about 90 ml of water while stirring,
and adjust the final volume to 100 ml using
additional water. Shake well to mix the solution.
3.1.3 Alkaline Absorbing Solution. 0.1 N
Sodium Hydroxide (NaOH). To prepare 100 ml
of the scrubber solution for the back pair of
impingers, dissolve 0.40 g of solid NaOH in
about 90 ml of water, and adjust the final SOlution volume to 100 ml using additional
water. Shake well to mix the solution.
3.1.4 Sodium Thiosulfate (Na₂S₂O₃.5H₂O)
3.2 Sample Preparation and Analysis.
3.2.1 Water. Same as in Section 3.1.1.
3.2.2 Absorbing Solution Blanks. A separate blank solution of each absorbing reagent should be prepared for analysis with the
field samples. Dilute 30 ml of each absorbing
solution to approximately the same final
volume as the field samples using the blank
sample of rinse water.
3.2.3 Halide Salt Stock Standard Solutions. Prepare concentrated stock solutions
from reagent grade sodium chloride (NaCl).
sodium bromide (NaBr). and sodium fluoride
(NaF). Each must be dried at 110 °C for two
or more hours and then cooled to room temperature in a desiccator immediately before
weighing. Accurately weigh 1.6 to 1.7 g of the
dried NaCl to within 0.1 mg. dissolve in
water, and dilute to 1 liter. Calculate the
exact Cl- concentration using Equation 26-1.
µg Cl⁻/ml = g of NaCl x 103 x 35.453/58.44
Eq. 26-1
In a similar manner, accurately weigh and
solubilize 1.2 to 1.3 g of dried NaBr and 2.2 to
2.3 g of NaF to make 1-liter solutions. Use
Equations 26-2 and 26-3 to calculate the Br -
and F- concentrations.
µg Br-/ml = g of NaBr X 103 X 79.904/102.90
Eq. 26-2
µg F-/ml = g of NaF x 103 X 18.998/41.99
Eq. 26-3
Alternately, solutions containing a nominal
certified concentration of 1000 mg/l NaCl are
commercially available as convenient stock
solutions from which standards can be made
by appropriate volumetric dilution. Refrigerate the stock standard solutions and store
no longer than one month.
3.2.4 Chromatographic Eluent. Effective
eluents for nonsuppressed IC using a resinor silica-based weak ion exchange column
are a 4 mM potassium hydrogen phthalate
solution, adjusted to pH 4.0 using a saturated
sodium borate solution, and a 4 mM 4-hydroxy benzoate solution, adjusted to pH 8.6
using 1 N NaOH. An effective eluent for suppressed ion chromatography is a solution
containing 3 mM sodium bicarbonate and 2.4
mM sodium carbonate. Other dilute solutions buffered to a similar pH and containing
no interfering ions may be used. When using
suppressed ion chromatography. If the
"water dip" resulting from sample injection
interferes with the chloride peak, use a 2 mM
NaOH/2.4 mM sodium bicarbonate eluent.
4. Procedure
4.1 Sampling.
4.1.1 Preparation of Collection Train. Prepare the sampling train as follows: Pour 15
ml of the acidic absorbing solution into each
one of the first pair of impingers, and 15 ml
of the alkaline absorbing solution Into each
one of the second pair of impingers. Connect
the impingers in series with the knockout
Impinger first, If used, followed by the two
impingers containing the acidic absorbing
solution and the two impingers containing
the alkaline absorbing solution. Place a
fresh charge of silica gel, or equivalent, in
the drying tube or impinger at the end of the
impinger train.
4.1.2 Adjust the probe temperature and
the temperature of the filter and the stopcock. i.e., the heated area in Figure 26-1 to a
temperature sufficient to prevent water condensation. This temperature should be at
least 20°C above the source temperature, but
not greater than 120 °C. The temperature
should be monitored throughout a sampling
run to ensure that the desired temperature is
maintained.
4.1.3 Leak-Check Procedure. A leak-check
prior to the sampling run is optional; however, a leak-check after the sampling run is
mandatory. The leak-check procedure is as
follows: Temporarily attach a suitable (e.g.,
0-40 cc/min) rotameter to the outlet of the
dry gas meter and place a vacuum gauge at
or near the probe inlet. Plug the probe inlet,
pull a vacuum of at least 250 mm Hg (10 in.
Hg). and note the flow rate as Indicated by
the rotameter. A leakage rate not in excess
of 2 percent of the average sampling rate is
acceptable. (NOTE: Carefully release the
probe inlet plug before turning off the
Pt. 60, App. A, Meth. 26
pump.) It is suggested (not mandatory) that
the pump be leak-checked separately, either
prior to or after the sampling run. If done
prior to the sampling run, the pump leakcheck shall precede the leak-check of the
sampling train described immediately above;
if done after the sampling run, the pump
leak-check shall follow the train leak-check.
To leak-check the pump, proceed as follows:
Disconnect the drying tube from the probeimpinger assembly. Place a vacuum gauge at
the inlet to either the drying tube or pump,
pull a vacuum of 250 mm (10 in.) Hg, plug or
pinch off the outlet of the flowmeter, and
then turn off the pump. The vacuum should
remain stable for at least 30 sec. Other leakcheck procedures may be used, subject to the
approval of the Administrator, U.S. Environmental Protection Agency.
4.1.4 Purge Procedure. Immediately before sampling, connect the purge line to the
stopcock. and turn the stopcock to permit
the purge pump to purge the probe (see Figure 1A of Figure 26-1). Turn on the purge
pump. and adjust the purge rate to 2 liters/
min. Purge for at least 5 minutes before sampling.
4.1.5 Sample Collection. Turn on the sampling pump. pull a slight vacuum of approximately 25 mm Hg (1 in. Hg) on the impinger
train. and turn the stopcock to permit stack
gas to be pulled through the impinger train
(see Figure IC of Figure 26-1). Adjust the
sampling rate to 2 liters/min, as indicated by
the rate meter, and maintain this rate to
within 10 percent during the entire sampling
run. Take readings of the dry gas meter volume and temperature, rate meter, and vacuum gauge at least once every 5 minutes during the run. A sampling time of 1 hour is recommended. Shorter sampling times may introduce a significant negative blas in the
HCI concentration. At the conclusion of the
sampling run, remove the train from the
stack, cool, and perform a leak-check as described in section 4.1.2.
4.2 Sample Recovery. Disconnect the
impingers after sampling. Quantitatively
transfer the contents of the acid impingers
and the knockout impinger, if used. to a
leak-free storage bottle. Add the water
rinses of each of these impingers and connecting glassware to the storage bottle. Repeat this procedure for the alkaline
impingers and connecting glassware using a
separate storage bottle. Add 25 mg sodium
thiosulfate per the product of ppm of halogen
anticipated to be in the stack gas times the
dscm stack gas sampled. [Note: This amount
of sodium thiosulfate includes a safety factor of approximately 5 to assure complete reaction with the hypohalous acid to form a
second C1- ion in the alkaline solution.]
Save portions of the absorbing reagents (0.1
N H2SO4 and 0.1 N NaOH) equivalent to the
amount used in the sampling train (these are
the absorbing solution blanks described in
Section 3.2.2); dilute to the approximate volume of the corresponding samples using
rinse water directly from the wash bottle
being used. Add the same amount of sodium
thiosulfate solution to the 0.1 N NaOH absorbing solution blank. Also, save a portion
of the rinse water used to rinse the sampling
train. Place each in a separate, prelabeled
storage bottle. The sample storage bottles
should be sealed, shaken to mix, and labeled.
Mark the fluid level.
4.3 Sample Preparation for Analysis. Note
the liquid levels in the storage bottles and
confirm on the analysis sheet whether or not
leakage occurred during transport. If a noticeable leakage has occurred, either void
the sample or use methods, subject to the approval of the Administrator, to correct the
final results. Quantitatively transfer the
sample solutions to 100-ml volumetric flasks,
and dilute to 100 ml with water.
4.4 Sample Analysis.
4.4.1 The IC conditions will depend upon
analytical column type and whether suppressed or nonsuppressed IC is used. An example chromatogram from a nonsuppressed
system using a 150-mm Hamilton PRP-X100
anion column, a 2 ml/min flow rate of 4 mM
4-hydroxy benzoate solution adjusted to a pH
of 8.6 using 1 N NaOH, a 50-µ 1 sample loop,
and a conductivity detector set on 1.0 µS full
scale is shown in Figure 26-2.
4.4.2 Before sample analysis, establish a
stable baseline. Next, Inject a sample of
water, and determine If any CI- Br-, or F-
appears In the chromatogram. If any of these
ions are present, repeat the load/injection
procedure until they are no longer present.
Analysis of the acid and alkaline absorbing
solution samples requires separate standard
calibration curves: prepare each according to