Regl. 6302, art. 405(b)(9)-5.2.9.2
Enmienda las Reglas 102 y 405 del Reglamento para el Control de la Contaminación Atmosférica
Length: 5,850 wordsOfficial source
Cite as Reglamento Núm. 6302, Art. 405(b)(9)-5.2.9.2
5.2.16 Container No. 11 (8 N HCI Blank). At
least once during each field test, place 200 ml
of water Into a sample container labeled No.
11. Then carefully add with stirring 25 ml of
8 N HCI. Mix well and seal the container.
5.2.17 Container No. 12 (Sample Filter
Blank). Once during each field test, place
into a petri dish labeled No. 12 three unused
blank filters from the same lot as the sampling filters. Seal the petri dish.
5.3 Sample Preparation. Note the level of
the liquid in each of the containers and determine if any sample was lost during shipment. If a noticeable amount of leakage has
occurred, either void the sample or use
methods, subject to the approval of the Administrator, to correct the final results. A
diagram illustrating sample preparation and
analysis procedures for each of the sample
train components is shown in Figure 29-3.
5.3.1 Container No. 1 (Sample Filter).
5.3.1.1 If particulate emissions are being
determined. first desiccate the filter and filter catch without added heat (do not heat
the filters to speed the drying) and weigh to
a constant weight as described in Section 4.3
of Method 5.
5.3.1.2 Following this procedure, or initially, if particulate emissions are not being
determined in addition to metals analysis,
divide the filter with its filter catch into
portions containing approximately 0.5 B
Pt. 60, App. A, Meth. 29
each. Place the pieces in the analyst's choice
of either Individual microwave pressure relief vessels or Parr Bombs. Add 6 ml of concentrated HNO3 and 4 ml of concentrated HF
to each vessel. For microwave heating.
microwave the samples for approximately 12
to 15 minutes total heating time as follows:
heat for 2 to 3 minutes, then turn off the
microwave for 2 to 3 minutes, then heat for
2 to 3 minutes. etc.. continue this alternation until the 12 to 15 minutes total heating time are completed (this procedure
should comprise approximately 24 to 30 minutes at 600 watts). Microwave heating times
are approximate and are dependent upon the
number of samples being digested simultaneously. Sufficient heating is evidenced by
sorbent reflux within the vessel. For conventional heating, heat the Parr R Bombs at 140
°C (285 °F) for 6 hours. Then cool the samples
to room temperature. and combine with the
acid digested probe rinse as required in Section 5.3.3.
5.3.1.3 If the sampling train Includes an
optional glass cyclone in front of the filter,
prepare and digest the cyclone catch by the
procedures described in section 5.3.1.2 and
then combine the digestate with the digested
filter sample.
5.3.2 Container No. 2 (Acetone Rinse).
Note the level of liquid in the container and
confirm on the analysis sheet whether or not
leakage occurred during transport. If a noticeable amount of leakage has occurred, either void the sample or use methods, subject
to the approval of the Administrator, to correct the final results. Measure the liquid in
this container either volumetrically within 1
ml or gravimetrically within 0.5 g. Transfer
the contents to an acid-cleaned, tared 250-ml
beaker and evaporate to dryness at ambient
temperature and pressure. If particulate
emissions are being determined, desiccate
for 24 hours without added heat, weigh to a
constant weight according to the procedures
described in Section 4.3 of Method 5, and report the results to the nearest 0.1 mg. Redissolve the residue with 10 ml of concentrated
HNO3.
ER25AP96.003
Quantitatively combine the resultant sample, including all liquid and any particulate
matter, with Container No. 3 before beginning Section 5.3.3.
5.3.3 Container No. 3 (Probe Rinse). Verify
that the pH of this sample is 2 or lower. If it
is not, acidify the sample by careful addition
with stirring of concentrated HNO 3 to pH 2.
Use water to rinse the sample into a beaker,
and cover the beaker with a ribbed watch
glass. Reduce the sample volume to approximately 20 ml by heating on a hot plate at a
temperature just below boiling. Digest the
sample in microwave vessels or Parr* Bombs
by quantitatively transferring the sample to
the vessel or bomb, carefully adding the 6 ml
of concentrated HNO 3. 4 ml of concentrated
HF, and then continuing to follow the procedures described in Section 5.3.1.2. Then combine the resultant sample directly with the
acid digested portions of the filter prepared
previously in Section 5.3.1.2. The resultant
combined sample is referred to as "Sample
Fraction 1". Filter the combined sample
using Whatman 541 filter paper. Dilute to 300
ml (or the appropriate volume for the expected metals concentration) with water.
This diluted sample is "Analytical Fraction
1". Measure and record the volume of Analytical Fraction 1 to within 0.1 ml. Quantitatively remove a 50-ml aliquot and label
as "Analytical Fraction 1B". Label the remaining 250-ml portion as "Analytical Fraction 1A". Analytical Fraction IA is used for
ICAP or AAS analysis for all desired metals
except Hg. Analytical Fraction 1B is used for
the determination of front-half Hg.
5.3.4 Container No. 4 (Impingers 1-3).
Measure and record the total volume of this
sample to within 0.5 ml and label it "Sample
Fraction 2". Remove a 75- to 100-ml aliquot
for Hg analysis and label the aliquot "Analytical Fraction 2B". Label the remaining portion of Container No. 4 as "Sample Fraction
2A". Sample Fraction 2A defines the volume
of Analytical Fraction 2A prior to digestion.
All of Sample Fraction 2A is digested to
produce "Analytical Fraction 2A". Analytical
Fraction 2A defines the volume of Sample
Fraction 2A after its digestion and the volume of Analytical Fraction 2A is normally
150 ml. Analytical Fraction 2A is analyzed
for all metals except Hg. Verify that the pH
of Sample Fraction 2A is 2 or lower. If necessary, use concentrated HNO3 by careful addition and stirring to lower Sample Fraction
2A to pH 2. Use water to rinse Sample Fraction 2A into a beaker and then cover the
beaker with a ribbed watch glass. Reduce
Sample Fraction 2A to approximately 20 ml
by heating on a hot plate at a temperature
Just below bolling. Then follow either of the
digestion procedures described in Sections
5.3.4.1 or 5.3.4.2.
5.3.4.1 Conventional Digestion Procedure.
Add 30 ml of 50 percent HNO₃ and heat for 30
minutes on a hot plate to just below boiling.
Add 10 ml of 3 percent H₂O₂ and heat for 10
more minutes. Add 50 ml of hot water, and
heat the sample for an additional 20 minutes.
Cool. filter the sample. and dilute to 150 ml
(or the appropriate volume for the expected
metals concentrations) with water. This dilution produces Analytical Fraction 2A.
Measure and record the volume to within 0.1
ml.
5.3.4.2 Microwave Digestion Procedure.
Add 10 ml of 50 percent HNO₃ and heat for 6
minutes total heating time in alternations of
1 to 2 minutes at 600 Watts followed by 1 to
2 minutes with no power, etc., similar to the
procedure described in Section 5.3.1. Allow
the sample to cool. Add 10 ml of 3 percent
H₂O₂ and heat for 2 more minutes. Add 50 ml
of hot water, and heat for an additional 5
minutes. Cool, filter the sample, and dilute
to 150 ml (or the appropriate volume for the
expected metals concentrations) with water.
This dilution produces Analytical Fraction
2A. Measure and record the volume to within
0.1 ml.
(NOTE: All microwave heating times given
are approximate and are dependent upon the
number of samples being digested at a time.
Heating times as given above have been
found acceptable for simultaneous digestion
of up to 12 individual samples. Sufficient
heating is evidenced by solvent reflux within
the vessel.)
5.3.5 Container No. 5A (Impinger 4), Container Nos. 5B and 5C (Impingers 5 and 6).
Keep the samples in Containers Nos. 5A, 5B,
and 5C separate from each other. Measure
and record the volume of 5A to within 0.5 ml.
Label the contents of Container No. 5A to be
Analytical Fraction 3A. To remove any
brown MnO2 precipitate from the contents of
Container No. 5B, filter its contents through
Whatman 40 filter paper into a 500 ml volumetric flask and dilute to volume with
water. Save the filter for digestion of the
brown MnO₂ precipitate. Label the 500 ml flltrate from Container No. 5B to be Analytical
Fraction 3B. Analyze Analytical Fraction 3B
for Hg within 48 hours of the filtration step.
Place the saved filter, which was used to remove the brown MnO₂ precipitate, into an
appropriately sized vented container, which
will allow release of any gases including
chlorine formed when the filter is digested.
In a laboratory hood which will remove any
gas produced by the digestion of the MnO₂,
add 25 ml of 8 N HCI to the filter and allow
to digest for a minimum of 24 hours at room
temperature. Filter the contents of Container No. 5C through a Whatman 40 filter
into a 500-ml volumetric flask. Then filter
the result of the digestion of the brown MnO₂
from Container No. 5B through a Whatman
40 filter into the same 500-ml volumetric
flask, and dilute and mix well to volume
with water. Discard the Whatman 40 filter.
Mark this combined 500-ml dilute HCI solution as Analytical Fraction 3C.
5.3.6 Container No. 6 (Silica Gel). Weigh
the spent silica gel (or silica gel plus impinger) to the nearest 0.5 g using a balance.
5.4 Sample Analysis. For each sampling
train sample run, seven individual analytical
samples are generated; two for all desired
Environmental Protection Agency, EPA
metals except Hg. and five for Hg. A schematic identifying each sample container and
the prescribed analytical preparation and
analysis scheme is shown in Figure 29-3. The
first two analytical samples, labeled Analytical Fractions 1A and 1B, consist of the digested samples from the front-half of the
train. Analytical Fraction 1A is for ICAP,
ICP-MS or AAS analysis as described in Sections 5.4.1 and 5.4.2. respectively. Analytical
Fraction 1B is for front-half Hg analysis as
described in Section 5.4.3. The contents of
the back-half of the train are used to prepare
the third through seventh analytical samples. The third and fourth analytical samples, labeled Analytical Fractions 2A and 2B,
contain the samples from the moisture removal impinger No. 1. If used, and HNO3 H2O₂
impingers Nos. 2 and 3. Analytical Fraction
2A is for ICAP, ICP-MS or AAS analysis for
target metals, except Hg. Analytical Fraction 2B is for analysis for Hg. The fifth
through seventh analytical samples, labeled
Analytical Fractions 3A, 3B, and 3C, consist
of the impinger contents and rinses from the
empty impinger No. 4 and the H 2SO/KMnO4
Impingers Nos. 5 and 6. These analytical
samples are for analysis for Hg as described
in Section 5.4.3. The total back-half Hg catch
is determined from the sum of Analytical
Fractions 2B, 3A, 3B, and 3C. Analytical
Fractions 1A and 2A can be combined proportionally prior to analysis.
5.4.1 ICAP and ICP-MS Analysis. Analyze
Analytical Fractions 1A and 2A by ICAP
using Method 6010 or Method 200.7 (40 CFR
part 136, appendix C). Calibrate the ICAP.
and set up an analysis program as described
in Method 6010 or Method 200.7. Follow the
quality control procedures described in Section 7.3.1. Recommended wavelengths for
analysis are as follows:
Pt. 60, App. A, Meth. 29
Wave-
Element
length
(nm)
Aluminum
308.215
Antimony
206.833
Arsenic
193.696
Barlum
455.403
Beryllium
313.042
Wave-
Element
length
(nm)
Cadmium
226.502
Chromium
267.716
Cobalt
228.616
Copper
324.754
Iron
259.940
Lead
220.353
Manganese
257.610
Nickel
231.604
Phosphorous
214.914
Selenium
196.026
Silver
328.068
Thailium
190.864
Zinc
213.858
These wavelengths represent the best combination of specificity and potential detection limit. Other wavelengths may be substituted if they can provide the needed specificity and detection limit. and are treated
with the same corrective techniques for
spectral Interference. Initially, analyze all
samples for the target metals (except Hg)
plus Fe and Al. If Fe and Al are present. the
sample might have to be diluted so that each
of these elements is at a concentration of
less than 50 ppm so as to reduce their spectral interferences on As. Cd. Cr. and Pb. Perform ICP-MS analysis by following Method
6020 in EPA Publication SW-846 Third Edition (November 1986) including updates I, II.
IIA. and пв. as incorporated by reference in
$60.17(i).
(NOTE: When analyzing samples In a HF
matrix, an alumina torch should be used;
since all front-half samples will contain HF,
use an alumina torch.)
5.4.2. AAS by Direct Aspiration and/or
GFAAS. If analysis of metals in Analytical
Fractions 1A and 2A by using GFAAS or direct aspiration AAS is needed, use Table 29-
2 to determine which techniques and procedures to apply for each target metal. Use
Table 29-2. if necessary. to determine techniques for minimization of Interferences.
Calibrate the Instrument according to Section 6.3 and follow the quality control procedures specified in Section 7.3.2.
TABLE 29-2.-APPLICABLE TECHNIQUES, METHODS AND MINIMIZATION OF INTERFERENCE FOR AAS
ANALYSIS
Metal
SW-8461
Technique
Wavelength
Interferences
method No.
(nm)
Cause
Minimization
Fe
Aspiration
7380
248.3
Contamination
Great care taken to avoid
contamination.
Pb
Aspiration
7420
283.3
217.0 nm alternate
Background correction required.
Pb
Furnace
7421
283.3
Poor recoveries
Matrix modifier, add 10 of
of phosphorus acid to 1
ml of prepared sample
in sampler cup.
Mn
Aspiration
7460
279.5
403.1 nm alternate
Background correction required.
TABLE 29-2.-APPLICABLE TECHNIQUES, METHODS AND MINIMIZATION OF INTERFERENCE FOR AAS
ANALYSIS-Continued
Technique
SW-8461
Interferences
Metal
Wavelength
method No.
(nm)
Cause
Minimization
NI
Aspiration
7520
232.0
352.4 nm alternate Fe,
Background correction re-
Co, and Cr.
quired.
Matrix matching or nitrousoxide/acetylene flame.
Nonlinear response
sample dilution or use
352.3 nm line.
Se
Furnace
7740
196.0
Volatility
Spike samples and reference materials and
add nickel nitrate to
minimize volatilization.
Adsorption & scatter
Background correction is
required and Zeeman
background correction
can be useful.
Ag
Aspiration
7760
328.1
Adsorption & Scatter AgCl
Background correction is
insoluble.
required. Avoid Hydrochioric acid unless silver
is in solution as a chioride complex Sample
and standards montored for aspiration rate.
TI
Aspiration
7840
276.8
Background correction is
required. Hydrochloric
acid should not be used.
TI
Furnace
7841
276.8
Hydrochloric acid or chio-
Background correction is
ride.
required.
Verify that losses are not
occurring for volatization
by spiked samples or
standard addition; Palladium is B suitable matrix
modifier.
Zn
Aspiration
7950
213.9
High Si, Cu, & P Contami-
Strontium removes Cu and
nation.
phosphate, Great care
taken to avoid contamination.
Sb
Aspiration
7040
217.6
1000 mg/ml Pb NI, Cu, or
Use secondary waveacid.
lengths of 231.1.nm;
match sample & standands acid concentration
or use nitrous
oxidefacetylene flame.
Sb
Furnace
7041
217.6
High Pb
Secondary Wavelength or
Zeeman correction.
As
Furnace
7060
193.7
Arsenic volatilization
Spiked samples and add
Aluminum
nickel nitrate solution to
digestates prior to anaiysis.
Use Zeeman background
correction.
Ba
Aspiration 7080
7080
553.6
Calcium
High hollow cathode our-
Barium ionization
rent and narrow band
sel.
2 ml of KCI per 100 ml of
sample.
Be
Aspiration
7090
234.9
500 ppm AI High Mg and
Add 0.1% fluoride.
Si.
Use method of standard
additions.
Be
Furnace
7091
234.9
Be in optical path
Optimize parameters to
minimize effects.
Cd
Aspiration
7130
228.8
Absorption and light scat-
Background correction is
tering.
required.
Cd
Furnace
7131
228.8
As above
As above.
Excess Chloride
Ammonium phosphate
Pipet tips
used as 8 matrix modifier.
Use cadmiun-free tips.
Environmental Protection Agency, EPA
TABLE 29-2.-APPLICABLE TECHNIQUES, METHODS AND MINIMIZATION OF INTERFERENCE FOR AAS
ANALYSIS-Continued
Metal
Technique
SW-8461
Wavelength
Interferences
method No.
(nm)
Cause
Minimization
Cr
Aspiration
7190
357.9
Akali metal
KCI ionization suppressant
in samples and standards-Consult mfgs literature.
Co
Furnace
7201
240.7
Excess chloride
Use Method of Standard
Additions.
Cr
Furnace
7191
357.9
200 mg/L Ca and P
All calcium nitrate for a
known constant effect
and to eliminate effect
of phosphate.
Cu
Aspiration
7210
324.7
Absorption & scatter
Consult manufacturer's
manual.
1 Refer to EPA publication SW-846 Third Edition (November 1986) including updates I, II, IIA, and IIB, as incorporated by reference in $60.17(i).
5.4.3 CVAAS Hg analysis. Analyze Analytical Fractions 1B, 2B, 3A, 3B. and 3C separately for Hg using CVAAS following the
method outlined in Method 7470 in EPA Publication SW-846 Third Edition (November
1986) including updates I. II, IIA and UB, as
Incorporated by reference in $60.17(i) or in
Standard Methods for the Examination of Water
and Wastewater, 16th Edition, (1985). Method
303F, as incorporated by reference in $60.17.
or, optionally using NOTE No. 2 in this section. Set up the calibration curve (zero to
1000 ng) as described in Method 7470 or similar to Method 303F using 300-ml BOD bottles
instead of Erlenmeyers. Perform the following for each Hg analysis. From each
original sample, select and record an aliquot
in the size range from 1 ml to 10 ml. If no
prior knowledge of the expected amount of
Hg in the sample exists. a 5 ml aliquot Is
suggested for the first dilution to 100 ml (see
NOTE No. 1 in this Section). The total
amount of Hg in the aliquot shall be less
than I µ g and within the range (zero to 1000
ng) of the callbration curve. Place the sample aliquot into a separate 300-ml BOD bottle, and add enough water to make a total
volume of 100 ml. Next add to it sequentially
the sample digestion solutions and perform
the sample preparation described In the procedures of Method 7470 or Method 303F. (See
NOTE No. 2 in this Section). If the maximum
readings are off-scale (because Hg in the allquot exceeded the calibration range; Including the situation where only a 1-ml aliquot
of the original sample was digested). then dilute the original sample (or a portion of it)
with 0.15 percent HNO (1.5 ml concentrated
HNO3 per liter aqueous solution) so that
when a 1- to 10-ml aliquot of the "0.15 HNO3
percent dilution of the original sample" is
digested and analyzed by the procedures described above, it will yield an analysis within the range of the calibration curve.
NOTE No. 1 TO SECTION 5.4.3. When Hg levels
in the sample fractions are below the instack detection limit given in Table 29-1. select a 10 ml aliquot for digestion and analysis as described.
NOTE No. 2 TO SECTION 5.4.3. Optionally. Hg
can be analyzed by using the CVAAS analyt-
Ical procedures given by some instrument
manufacturer's directions. These include
callbration and quality control procedures
for the Leeman Model PS200, the Perkin
Elmer FIAS systems, and similar models, if
available. of other instrument manufacturers. For digestion and analyses by these instruments, perform the following two steps:
(1) Digest the sample aliquot through the
addition of the aqueous hydroxylamine hydrochloride/sodium chloride solution the
same as described in this Section 5.4.3.: (The
Leeman, Perkin Elmer, and similar instruments
described in this note add automatically the
necessary stannous chloride solution during the
automated analysis of Hg.) and
(2) Upon completion of the digestion described in paragraph (I), of this note, analyze
the sample according to the instrument
manufacturer's directions. This approach allows multiple (Including duplicate) automated analyses of a digested sample aliquot.
6. Calibration
Maintain a laboratory log of all calibrations.
6.1 Sampling Train Calibration. Calibrate
the sampling train components according to
the Indicated sections of Method 5: Probe
Nozzle (Section 5.1); Pitot Tube (Section 5.2):
Metering System (Section 5.3): Probe Heater
(Section 5.4): Temperature Gauges (Section
5.5); Leake-Check of the Metering System
(Section 5.6): and Barometer (Section 5.7).
6.2 Industively Coupled Argon Plasma
Spectrometer Calibration. Prepare standards
as outlined In Section 4.5. Profile and calibrate the instrument according to the manufacturer's recommended procedures using
those standards. Check the calibration once
per hour. If the Instrument does not reproduce the standard concentrations within 10
percent, perform the complete calibration
procedures. Perform ICP-MS analysis by following Method 6020 in EPA Publication SW-
846 Third Edition (November 1986) including
updates I, II, ПА and IIB, as incorporated by
reference in § 60.17(i).
6.3 Atomic Absorption Spectrometer-Direct Aspiration AAS, GFAAS, and CVAAS
analyses. Prepare the standards as outlined
in Section 4.5 and use them to calibrate the
spectrometer. Calibration procedures are
also outlined in the EPA methods referred to
in Table 29-2 and in Method 7470 in EPA Publication SW-846 Third Edition (November
1986) including updates I. II. IIA and IIB. as
incorporated by reference in $60.17(i) or in
Standard Methods for the Examination of Water
and Wastewater, 16th Edition, (1985). Method
303F (for Hg) as incorporated by reference in
$60.17. Run each standard curve in duplicate
and use the mean values to calculate the
calibration line. Recalibrate the instrument
approximately once every 10 to 12 samples.
7. Quality Control
7.1 Field Reagent Blanks, If analyzed.
Perform the digestion and analysis of the
blanks in Container Nos. 7 through 12 that
were produced in Sections 5.2.11 through
5.2.17, respectively. For Hg field reagent
blanks. use a 10 ml aliquot for digestion and
analysis.
7.1.1 Digest and analyze one of the filters
from Container No. 12 per Section 5.3.1. 100
ml from Container No. 7 per Section 5.3.2.
and 100 ml from Container No. 8A per Section 5.3.3. This step produces blanks for Analytical Fractions 1A and 1B.
7.1.2 Combine 100 ml of Container No. 8A
with 200 ml from Container No. 9, and digest
and analyze the resultant volume per Section 5.3.4. This step produces blanks for Analytical Fractions 2A and 2B.
7.1.3 Digest and analyze a 100-ml portion
of Container No. 8A to produce a blank for
Analytical Fraction 3A.
7.1.4 Combine 100 ml from Container No.
10 with 33 ml from Container No. 8B to
produce a blank for Analytical Fraction 3B.
Filter the resultant 133 ml as described for
Container No. 5B in Section 5.3.5, except do
not dilute the 133ml. Analyze this blank for
Hg within 48 hrs. of the filtration step. and
use 400 ml as the blank volume when calculating the blank mass value. Use the actual
volumes of the other analytical blanks when
calculating their mass values.
7.1.5 Digest the filter that was used to remove any brown MnO₂ precipitate from the
blank for Analytical Fraction 3B by the
same procedure as described in Section 5.3.5
for the similar sample filter. Filter the
digestate and the contents of Container No.
11 through Whatman 40 paper into a 500-ml
volumetric flask, and dilute to volume with
water. These steps produce a blank for Analytical Fraction 3C.
7.1.6 Analyze the blanks for Analytical
Fraction Blanks 1A and 2A per Section 5.4.1
and/or Section 5.4.2. Analyze the blanks for
Analytical Fractions 1B. 2B, 3A. 3B. and 3C
per Section 5.4.3. Analysis of the blank for
Analytical Fraction 1A produces the fronthalf reagent blank correction values for the
desired metals except for Hg; Analysis of the
blank for Analytical Fraction IB produces
the front-half reagent blank correction value
for Hg. Analysis of the blank for Analytical
Fraction 2A produces the back-half reagent
blank correction values for all of the desired
metals except for Hg. while separate analyses of the blanks for Analytical Fractions
2B. 3A, 3B, and 3C produce the back-half reagent blank correction value for Hg.
7.2 Quality Control Samples. Analyze the
following quality control samples.
7.2.1 ICAP and ICP-MS Analysis. Follow
the respective quality control descriptions inalyze each
sample by the Method of Standard Additions.
Analyze a quality control sample to check
the accuracy of the calibration standards. If
the results are not within 20 percent. repeat
the calibration.
7.2.3 CVAAS Analysis for Hg. Analyze all
samples in duplicate. Analyze a quality control sample to check the accuracy of the
calibration standards (if not within 15 percent, repeat calibration). Perform a matrix
spike on one sample (if not within 25 percent,
analyze all samples by the Method of Standard Additions). Additional Information on
quality control can be obtained from Method
Environmental Protection Agency, EPA
7470 of EPA Publication SW-846 Third Edition (November 1986) Including updates I. II,
IIA and IIB, as incorporated by reference in
$60.17(i) or in Standard Methods for the Examination of Water and Wastewater, 16th Edition,
(1985). Method 303F as incorporated by reference in $60.17.
8. Calculations
8.1 Dry Gas Volume. Using the data from
this test, calculate Vₘₓ std). the dry gas sample
volume at standard conditions as outlined in
Section 6.3 of Method 5.
8.2 Volume of Water Vapor and Moisture
Content. Using the total volume of condensate collected during the source sampling,
calculate the volume of water vapor Vw (atd)
and the moisture content Bws of the stack
gas. Use Equations 5-2 and 5-3 of Method 5.
8.3 Stack Gas Velocity. Using the data
from this test and Equation 2-9 of Method 2.
calculate the average stack gas velocity.
8.4 Metals (Except Hg) in Source Sample.
8.4.1 Analytical Fraction 1A, Front-Half,
Metals (except Hg). Calculate separately the
amount of each metal collected in Sample
Fraction 1 of the sampling train using the
following equation:
Ma=Cal Fd Vaoin,1
Eq. 29-1
where:
Ma=Total mass of each metal (except Hg)
collected in the front half of the sampling train (Sample Fraction 1), µ g.
C₁=Concentration of metal in Analytical
Fraction 1A as read from the standard
curve. µg/ml.
F&=Dilution factor (F d = the inverse of the
fractional portion of the concentrated
sample in the solution actually used in
the instrument to produce the reading
Cal. For example, If a 2 ml aliquot of Analytical Fraction 1A is diluted to 10 ml to
place it in the calibration range, Fd - 5).
Vaoin,1=Total volume of digested sample solution (Analytical Fraction 1). ml.
8.4.1.1 If Analytical Fractions 1A and 2A
are combined, use proportional aliquots.
Then make appropriate changes in Equations
29-1 through 29-3 to reflect this approach.
8.4.2 Analytical Fraction 2A, Back-Half.
Metals (except Hg). Calculate separately the
amount of each metal collected in Fraction
2 of the sampling train using the following
equation.
where:
Mes-Total mass of each metal (except Hg)
collected in the back-half of the sampling train (Sample Fraction 2), H g.
C,-Concentration of metal in Analytical
Fraction 2A as read from the standard
curve. (µg/ml).
Fr=Aliquot factor, volume of Sample Fraction 2 divided by volume of Sample Fraction 2A (see Section 5.3.4.)
M6h=Ca F, V.
Eq. 29-2
Va=Total volume of digested sample solution
(Analytical Fraction 2A), ml (see Section
5.3.4.1 or 5.3.4.2, as applicable).
8.4.3 Total Train. Metals (except Hg). Calculate the total amount of each of the quantified metals collected in the sampling train
as follows:
M,=(M₂ - Mas) + (Mah - Monb)
Eq. 29-3
where:
M,=Total mass of each metal (separately
stated for each metal) collected in the
sampling train. µg.
Mₘ₆=Blank correction value for mass of
metal detected in front-half field reagent
blank, µg.
Math=Blank correction value for mass of
metal detected In back-half field reagent
blank. µg.
8.4.3.1 If the measured blank value for the
front half (M mb) is in the range 0.0 to "A" µg
[where "A" µg equals the value determined
by multiplying 1.4 µg/in.² times the actual
area in in.2 of the sample filter]. use M fhb to
correct the emission sample value (Ma ): if
Mmb exceeds "A" µg. use the greater of I or
II:
I. "A" µg.
II. the lesser of (a) M thbs or (b) 5 percent of
Mm.
If the measured blank value for the blackhalf (Mbab) is in the range 0.0 to 1 µg, use Mbhb
to correct the emission sample value (M bh): If
Mbbb) exceeds 1 µg, use the greater of I or II:
I. 1 µg.
II. the lesser of (a) Mbhb or (b) 5 percent of
Mbh.
8.5 Hg in Source Sample.
8.5.1 Analytical Fraction 1B: Front-Half
Hg. Calculate the amount of Hg collected in
the front-half, Sample Fraction 1, of the
sampling train by using Equation 29-4:
ER25AP96.005
where:
Hgn=Total mass of Hg collected in the fronthalf of the sampling train (Sample Fraction 1), µg.
Qa=Quantity of Hg, µg, TOTAL in the ALI-
QUOT of Analytical Fraction 1B selected
for digestion and analysis.
8.5.1.1 For example, if a 10 ml aliquot of
Analytical Fraction 1B is taken and digested
and analyzed (according to Section 5.4.3 and
its NOTES Nos. 1 and 2), then calculate and
use the total amount of Hg in the 10 ml allquot for Qm.
Vacin,1=Total volume of Analytical Fraction 1,
ml.
VπB=Volume of aliquot of Analytical Fraction IB analyzed, ml.
8.5.1.2 For example, if a 1 ml aliquot of
Analytical Fraction IB was diluted to 50 ml
Pt. 60, App. A, Meth. 29
with 0.15 percent HNO3 as described in Section 5.4.3 to bring it into the proper analytical range, and then 1 ml of that 50-ml wa digested according to Section 5.4.3 and analyzed, VnB would be 0.02 ml.
8.5.2 Analytical Fractions 2B, 3A, 3B. and
3C: Back Half Hg.
8.5.2.1 Calculate the amount of Hg collected in Sample Fraction 2 by using Equation 29-5:
40 CFR Ch. I (7-1-99 Edition)
ER25AP96.006
where:
Hgsh=Total mass of Hg collected in Sample
Fraction 2, µg.
Q of Hg, µg. TOTAL in the ALI-
QUOT of Analytical Fraction 2B selected
for digestion and analysis.
8.5.2.1.1 For example, if a 10 ml aliquot of
Analytical Fraction 2B is taken and digested
and analyzed (according to Section 5.4.3 and
its NOTES Nos. I and 2), then calculate and
use the total amount of Hg in the 10 ml aliquot for Quitz-
Vaoin,2=Total volume of Sample Fraction 2,
ml.
VπB=Volume of Analytical Fraction 2B analyzed, ml.
8.5.2.1.2 For example, If 1 ml of Analytical
Fraction 2B was diluted to 10 ml with 0.15
percent HNO3 as described in Section 5.4.3 to
bring it into the proper analytical range, and
then 5 ml of that 10-ml was analyzed, V C2B
would be 0.5 ml.
8.5.2.2 Calculate each of the back-half Hg
values for Analytical Fractions 3A, 3B, and
3C by using Equation 29-6:
ER25AP96.007
where:
mass of Hg collected separately in Fraction 3A, 3B, or 3C, µg.
Quantacy=Quantity of Hg, µg. TOTAL. separately, in the ALIQUOT of Analytical Fraction 3A, 3B. and 3C selected for digestion
and analysis, (see previous notes in Sections 8.5.1 and 8.5.2 describing the quantity "Q" and calculate similarly).
VO(A,B,C)=Volume, separately, of Analytical
Fraction 3A. 3B, or 3C analyzed, ml (see
previous notes in Sections 8.5.1 and 8.5.2,
describing the quantity "V" and calculate similarly).
Vaoia,KA,B,C)=Total volume, separately. of Analytical Fraction 3A. 3B. or 3C, ml.
8.5.2.3 Calculate the total amount of Hg
collected in the back-half of the sampling
train by using Equation 29-7:
Eq. 29-7
where:
Hgah=Total mass of Hg collected in the backhalf of the sampling train. µg.
8.5.3 Total Train Hg Catch. Calculate the
total amount of Hg collected in the sampling
train by using Equation 29-8:
Eq. 29-8
where:
Hg=Total mass of Hg collected In the sampling train, µg.
Hgms=Blank correction value for mass of Hg
detected in front-half field reagent
blank. µg.
Hgass-Blank correction value for mass of Hg
detected in back-half field reagent
blanks, µg.
8.5.4 If the total of the measured blank
values (Hgmh+Hghh) is in the range of 0.0 to
0.6 µg. then use the total to correct the sample value (Hga+Hgan): if it exceeds 0.6 µg, use
the greater of I. or II:
II. the lesser of (a) (Hg mo+Hgbbb). or (b) 5
percent of the sample value (Hge +Hgbb).
8.6 Individual Metal Concentrations in
Stack Gas. Calculate the concentration of
each metal in the stack gas (dry basis, adjusted to standard conditions) by using
Equation 29-9:
I. 0.6 µg.
ER25AP96.008
C=Concentration of a metal In the stack
gas, mg/dscm.
K4=10⁻³ mg/µg.
M,=Total mass of that metal collected in the
sampling train. µg: (substitute Hg, for M,
for the Hg calculation).
Vm(etd)=Volume of gas sample as measured by
the dry gas meter, corrected to dry
standard conditions, dscm.
8.7 Isokinetic Variation and Acceptable
Results. Same as Method 5, Sections 6.11 and
6.12, respectively.
Environmental Protection Agency, EPA
3. Bibliography
1. Method 303F in Standard Methods for the
Examination of Water Wastewater, 16th Edition, 1985. Available from the American Public Health Association, 1015 18th Street NW.,
Washington, DC 20036.
2. EPA Methods 6010, 6020, 7000, 7041, 7060,
7131, 7421, 7470, 7740, and 7841, Test Methods for
Evaluating Solid Waste: Physical/Chemical
Methods. SW-846, Third Edition, September
1986, with updates I. II, ПА and IIB. Office of
Solid Waste and Emergency Response. U.S.
Environmental Protection Agency, Washington, DC 20460.
3. EPA Method 200.7. Code of Federal Regulations, Title 40. Part 136, Appendix C. July 1.
1987.
Pt. 60, App. A, Meth. 29
4. EPA Methods 1 through 5, Code of Federal
Regulations, Title 40, Part 60, Appendix A.
July 1, 1991.
5. EPA Method 101A, Code of Federal Regulations, Title 40, Part 61, Appendix B. July 1.
1991.
[36 FR 24877. Dec. 23, 1971]
EDITORIAL NOTE: For FEDERAL REGISTER cltations affecting part 60, appendix A see the
List of CFR Sections in the Finding Aids section of this volume.
EFFECTIVE DATE NOTE: At 64 FR 26490. May
14, 1999, Appendix A to part 60 was amended
by adding Methods 2F, 2G, and 2H and corrected at 64 FR 37196, July 9, 1999, and 64 FR
38241, July 15, 1999, effective July 13. 1999.
40 CFR PART 60 APPENDIX B
APPENDIX B-PERFORMANCE SPECIFICATIONS
Performance Specification 1-Specifications and test procedures for opacity continuous emission monitoring
systems in stationary sources
Performance Specification 2-Specifications and test procedures for SO₂ and NO. continuous emission monitoring systems in stationary sources
Performance Specification 3-Specifications and test procedures for O₂ and CO₂ continuous emission monitoring systems in stationary sources
Performance Specification 4-Specifications and test procedures for carbon monoxide continuous emission
monitoring systems in stationary sources
Performance Specification 4A-Specifications and test
procedures for carbon monoxide continuous emission
monitoring systems in stationary sources
Performance Specification 5-Specifications and test procedures for TRS continuous emission monitoring
systems in stationary sources
Performance Specification 6-Specifications and test procedures for continuous emission rate monitoring systems in stationary sources
Performance Specification 7-Specifications and test procedures for hydrogen suifide continuous emission
monitoring systems in stationary sources
PERFORMANCE SPECIFICATION 1-SPECIFICATIONS AND
TEST PROCEDURES FOR OPACITY CONTINUOUS EMIS-
SION MONITORING SYSTEMS IN STATIONARY
SOURCES
1. Applicability and Principle
1.1 Applicability. This specification contains requirements for the design. performance, and installation of instruments for opacity continuous emission monitoring systerms (CEMS's) and data computation procedures for evaluating the acceptability of a CEMS. Certain design requirements and test procedures established in this specification may not apply to all instrument designs. In such
instances, equivalent design requirements and test procedures may be used with prior approval of the Administrator.
Performance Specification 1 (PS 1) applies to opacity
monitors installed after March 30, 1983. Opacity monitors
installed before March 30, 1983, are required to comply
with the provisions and requirements of PS 1 except for
the following:
(a) Section 4. "Installation Specifications."
(b) Sections 5.1.4, 5.1.6, 5.1.7, and 5.1.8 of Section 5.
"Design and Performance Specifications."
(c) Section 6.4 of Section 6 "Design Specifications
Verification Procedure."
An opacity monitor installed before March 30, 1983,
need not be tested to demonstrate compliance with PS 1
unless required by regulatory action other than the promulgation of PS 1. If an existing monitor is replaced with
a new monitor, PS I shall apply except that the new monitor may be located at the old measurement location regardless of whether the location meets the requirements of