Regl. 6302, art. 405(b)(9)-6
however. the leak check is optional.
Length: 2,151 wordsOfficial source
Cite as Reglamento Núm. 6302, Art. 405(b)(9)-6
3.3 Place the probe in the stack, with the
tip of the probe positioned at the sampling
point: purge the sampling line long enough
to allow at least five exchanges. Draw a sample into the analyzer, and immediately analyze it for percent CO2 and percent O2. Determine the percentage of the gas that is N 2 and
CO by subtracting the sum of the percent
CO₂ and percent 0₂ 0 from 100 percent. Calculate the dry molecular weight as indicated
in Section 7.2.
3.4 Repeat the sampling. analysis. and calculation procedures until the dry molecular
weights of any three grab samples differ
from their mean by no more than 0.3 g/g-
mole (0.3 1b/lb-mole). Average these three
molecular weights, and report the results to
the nearest 0.1 g/g-mole (0.1 lb/Ib-mole).
4. SINGLE-POINT, INTEGRATED SAMPLING AND
ANALYTICAL PROCEDURE
4.1 The sampling point in the duct shall be
located as specified in Section 3.1.
4.2 Leak check (optional) the flexible bag
as in Section 2.2.6. Set up the equipment as
shown in Figure 3-2. Just before sampling.
leak check (optional) the train by placing a
vacuum gauge at the condenser inlet. pulling
a vacuum of at least 250 mm Hg (10 In. Hg).
plugging the outlet at the quick disconnect,
and then turning off the pump. The vacuum
should remain stable for at least 0.5 minute.
Evacuate the flexible bag. Connect the
probe, and place it in the stack, with the tip
of the probe positioned at the sampling
point: purge the sampling line. Next. connect
the bag. and make sure that all connections
are tight.
4.3 Sample at a constant rate. The sampling run should be simultaneous with, and
for the same total length of time as, the pollutant emission rate determination. Collection of at least 30 liters (1.00 ft 3) of sample
gas is recommended; however, smaller volumes may be collected, If desired.
4.4 Obtain one integrated flue gas sample
during each pollutant emission rate determination. Within 8 hours after the sample is
taken, analyze it for percent CO₂ and percent
O₂ using either an Orsat analyzer or a Fyrite
type combustion gas analyzer. If an Orsat
analyzer is used. It is recommended that
Orsat leak check described in Section 6, be
performed before this determination; however, the check is optional. Determine the
percentage of the gas that is N₂ and CO by
subtracting the sum of the percent CO₂ and
percent 0 from 100 percent. Calculate the dry
molecular weight as indicated in Section 7.2.
4.5 Repeat the analysis and calculation
procedures until the individual dry molecular weights for any three analyses differ
from their mean by no more than 0.3 g/g-
mole (0.3 lb/lb-mole). Average these three
molecular weights, and report the results to
the nearest 0.1 g/g-mole (0.1 lb/lb-mole).
5. MULTI-POINT, INTEGRATED SAMPLING AND
ANALYTICAL PROCEDURE
5.1 Unless otherwise specified by the Administrator, a minimum of eight traverse
points shall be used for circular stacks having diameters less than 0.61 m (24 in.). a minimum of nine shall be used for rectangular
stacks having equivalent diameters less than
0.61 m (24 in.). and a minimum of 12 traverse
points shall be used for all other cases. The
traverse points shall be located according to
Method 1. The use of fewer points is subject
to approval of the Administrator.
5.2 Follow the procedures outlined in Sections 4.2 through 4.5. except for the following: Traverse all sampling points, and
sample at each point for an equal length of
time. Record sampling data as shown in Figure 3-3.
Time
Traverse pt.
Q. liter/min
% dav."
Average
*% X 100 (Must be s10%)
METHOD 3A 40 CFR PART 60 APPENDIX A
Pt. 60, App. A, Meth. 3A
Figure 3-3. Sampling rate data.
6.
LEAK-CHECK
PROCEDURE
FOR
ANALYZER
Moving an Orsat analyzer frequently
causes It to leak. Therefore. an Orsat analyzer should be thoroughly leak checked on
site before the flue gas sample is introduced
into It. The procedure for leak checking an
Orsat analyzer is as follows:
6.1 Bring the liquid level in each pipette up
to the reference mark on the capillary tubing. and then close the pipette stopcock.
6.2 Raise the leveling bulb sufficiently to
bring the confining liquid meniscus onto the
graduated portion of the burette, and then
close the manifold stopcock.
6.3 Record the meniscus position.
6.4 Observe the menisus in the burette and
the liquid level in the pipette for movement
over the next 4 minutes.
6.5 For the Orsat analyzer to pass the leak
check, two conditions must be met:
6.5.1 The liquid level in each pipette must
not fall below the botton of the capillary
tubing during this 4-minute interval.
6.5.2 The menisus in the burette must not
change by more than 0.2 ml during this 4-
minute interval.
6.6 If the anlyzer fails the leak-check procedure. check all rubber connections and
stopcocks to determine whether they might
be the cause of the leak. Disassemble, clean,
and regrease leaking stopcocks. Replace
leaking rubber connections. After the analyzer is reassembled, repeat the lead-check
procedure.
7. CALCULATIONS
7.1 Nomenclature
Ma = Dry molecular weight, g/g-mole (1b/1b-
mole).
%CO₂ = Percent CO2 by volume, dry basis.
%O₂ = Percent O₂ by volume. dry basis.
%CO = Percent CO by volume, dry basis.
%N₂ - Percent N₂ by volume, dry basis.
0.280 - Molecular weight of N₂ or CO. divided
by 100.
0.320 - Molecular wight of O2 divided by 100.
0.440 = Molecular weight of CO2 divided by
100.
7.2 Dry Molecular Weight. Use Equation 3-
1 to calculate the dry molecular weight of
the stack gas.
Md - 0.440(%CO₂) + 0.320 (%0 2) + 0.280(%N₂ +
%CO)
Eq. 3-1
NOTE. The above equation does not consider argon in air (about 0.9 percent, molecular weight of 39.9). A negative error of
about 0.4 percent is introduced. The tester
may choose to include argon in the analysis
using procedures subject to approval of the
Administrator.
8. BIBLIOGRAPHY
ORSATl. Altshuller, A.P. Storage of Gases and
Vapors in Plastic Bags. International Journal of Air and Water Pollution. 6:75-81. 1963.
2. Conner, William D. and J.S. Nader. Air
Sampling with Plastic Bags. Journal of the
American Industrial Hyglene Association.
25.292-297. 1964.
3. Burrell Manual for Gas Analysts, Seventh edition. Burrell Corporation, 2223 Fifth
Avenue, Pittsburgh, PA. 15219. 1951.
4. Mitchell. W.J. and M.R. Midgett. Field
Reliability of the Orsat Analyzer. Journal of
Air Pollution Control Association. 26:491-495.
May 1976.
5. Shigehara, R.T., R. M. Neulicht, and
W.S. Smith. Validating Orsat Analysis Data
from Fossil Fuel-Fired Units. Stack Sampling News. 4(2):21-26. August 1976.
METHOD 3A-DETERMINATION OF OXYGEN AND
CARBON DIOXIDE CONCENTRATIONS IN EMIS-
SIONS FROM STATIONARY SOURCES (INSTRU-
MENTAL ANALYZER PROCEDURE)
1. Applicability and Principle
1.1 Applicability. This method is applicable to the determination of oxygen (O₂)
and carbon dioxide (CO₂) concentrations in
emissions from stationary sources only when
specified within the regulations.
1.2 Principle. A sample is continuously
extracted from the effluent stream: a portion
of the sample stream is conveyed to an instrumental analyzer(s) for determination of
O₂ and CO2 concentration(s). Performance
specifications and test procedures are provided to ensure reliable data.
2. Range and Sensitivity
Same as Method 6C, Sections 2.1 and 2.2,
except that the span of the monitoring system shall be selected such that the average
O2 or CO2 concentration is not less than 20
percent of the span.
3. Definitions
3.1 Measurement System. The total equipment required for the determination of the
O₂ or CO₂ concentration. The measurement
system consists of the same major subsystems as defined in Method 6C, Sections
3.1.1. 3.1.2. and 3.1.3.
3.2 Span, Calibration Gas, Analyzer Callbration Error, Sampling System Bias, Zero
Drift, Calibration Drift, Response Time. and
Calibration Curve. Same as Method 6C. Sections 3.2 through 3.8, and 3.10.
3.3 Interference Response. The output response of the measurement system to a component in the sample gas, other than the gas
component being measured.
4. Measurement System Performance Specifications
Same as Method 6C. Sections 4.1 through
4.4.
5. Apparatus and Reagents
Environmental Protection Agency, EPA
Pt. 60, App. A, Meth. 3A
5.1 Measurement System. Any measurement system for O₂ or CO₂ that meets the
specifications of this method. A schematic of
an acceptable measurement system is shown
in Figure 6C-1 of Method 6C. The essential
components of the measurement system are
described below:
5.1.1 Sample Probe. A leak-free probe, of
sufficient length to traverse the sample
points.
5.1.2 Sample Line. Tubing, to transport
the sample gas from the probe to the moisture removal system. A heated sample line is
not required for systems that measure the O₂
or CO₂ concentration on a dry basis, or transport dry gases.
5.1.3 Sample Transport Line, Calibration
Value Assembly. Moisture Removal System,
Particulate Filter, Sample Pump. Sample
Flow Rate Control, Sample Gas Manifold,
and Data Recorder. Same as Method 6C, Sections 5.1.3 through 5.1.9. and 5.1.11, except
that the requirements to use stainless steel.
Teflon, and nonreactive glass filters do not
apply.
5.1.4 Gas Analyzer. An analyzer to determine continuously the O₂ or CO2 concentration in the sample gas stream. The analyzer
shall meet the applicable performance specifications of Section 4. A means of controlling
the analyzer flow rate and a device for determining proper sample flow rate (e.g., precision rotameter, pressure gauge downstream
of all flow controls, etc.) shall be provided at
the analyzer. The requirements for measuring and controlling the analyzer flow rate
are not applicable if data are presented that
demonstrate the analyzer is insensitive to
flow variations over the range encountered
during the test.
5.2 Calibration Gases. The calibration
gases for CO₂ analyzers shall be CO2 in N₂ or
CO₂ in air. Alternatively. CO₂/SO₂ O₂/SO₂ or
O₂/CO₂/SO₂ gas mixtures in N₂ may be used.
Three calibration gases, as specified Section
5.3.1 through 5.3.3 of Method 6C. shall be
used. For O₂ monitors that cannot analyze
zero gas, a calibration gas concentration
equivalent to less than 10 percent of the span
may be used in place of zero gas.
6. Measurement System Performance Test Procedures
Perform the following procedures before
measurement of emissions (Section 7).
6.1 Calibration
Concentration
Verification. Follow Section 6.1 of Method
6C, except If calibration gas analysis is required, use Method 3 and change the acceptance criteria for agreement among Method 3
results to 5 percent (or 0.2 percent by volume, whichever is greater).
6.2 Interference Response. Conduct an interference response test of the analyzer prior
to Its initial use in the field. Thereafter, recheck the measurement system If changes
are made in the instrumentation that could
alter the interference response (e.g., changes
in the type of gas detector). Conduct the interference response in accordance with Section 5.4 of Method 20.
6.3 Measurement System Preparation.
Analyzer Calibration Error, and Sampling
System Bias Check. Follow Sections 6.2
through 6.4 of Method 6C.
7. Emission Test Procedure
7.1 Selection of Sampling Site and Sampling Points. Select a measurement site and
sampling points using the same criteria that
are applicable to tests performed using
Method 3.
7.2 Sample Collection. Position the sampling probe at the first measurement point,
and begin sampling at the same rate as used
during the sampling system bias check.
Maintain constant rate sampling (i.e., +10
percent) during the entire run. The sampling
time per run shall be the same as for tests
conducted using Method 3 plus twice the system response time. For each run, use only
those measurements obtained after twice the
response time of the measurement system
has elapsed to determine the average effluent concentration.
7.3 Zero and Calibration Drift Test. Follow Section 7.4 of Method 6C.
8. Quality Control Procedures
The following quality control procedures
are recommended when the results of this
method are used for an emission rate correction factor, or excess air determination. The
tester should select one of the following options for validating measurement results:
8.1 If both O₂ and CO₂ are measured using
Method 3A, the procedures described In Section 4.4 of Method 3 should be followed to
validate the O₂ and CO2 measurement results.
8.2 If only O₂ is measured using Method
3A. measurements of the sample stream CO2
concentration should be obtained at the sample by-pass vent discharge using an Orsat or
Fyrite analyzer. or equivalent. Duplicate
samples should be obtained concurrent with
at least one run. Average the duplicate Orsat
or Fyrite analysis results for each run. Use
the average CO2 values for comparison with
the O₂ measurements in accordance with the
procedures described in Section 4.4 of Method 3.
8.3 If only CO2 is measured using Method
3A, concurrent measurements of the sample
stream CO2 concentration should be obtained
using an Orsat or Fyrite analyzer as described in Section 8.2. For each run, differences greater than 0.5 percent between the
Method 3A results and the average of the duplicate Fyrite analysis should be investigated.
9. Emission Calculation
For all CO₂ analyzers, and for O₂ analyzers
that can be calibrated with zero gas, follow
Pt. 60, App. A, Meth. 3B