Regl. 6303, art. 6
3, and record all data listed in Figure 18-7. Prepare the GC so that
Length: 470 wordsOfficial source
Cite as Reglamento Núm. 6303, Art. 6
gas can be drawn through the sample valve. Flush the sample loop with gas from
one of the three Tedlar bags containing a calibration mixture, and activate the
valve. Obtain at least two chromatograms for the mixture. The results are
acceptable when the peak areas from two consecutive injections agree to within
5 percent of their average. If they do not agree, run additional samples or
correct the analytical techniques until this requirement is met. Then analyze
the other two calibration mixtures in the same manner. Prepare a calibration
curve as described in the same manner. Prepare a calibration curve as described
in Section 6.3. If the results are acceptable, analyze the other two calibration
gas mixtures in the same manner. Prepare the calibration curve by using the
least squares method.
Analyze the two field audit samples as described in Section 6.5 by connecting
each Tedlar bag containing an audit gas mixture to the sampling valve. Calculate
the results; record and report the data to the audit supervisor. If the results
are acceptable, proceed with the analysis of the source samples.
Analyze the source gas samples by connecting each bag to the sampling valve with
a piece of Teflon tubing identified with that bag. Follow the restrictions on
replicate samples specified for the calibration gases. Record the data in Figure
18-11. If certain items do not apply, use the notation "N.A.". If the bag has
been maintained at an elevated temperature as described in Section 7.1.4,
determine the stack gas water content by Method 4. After all samples have been
analyzed, repeat the analysis of the calibration gas mixtures, and generate a
second calibration curve. Use an average of the two curves to determine the
sample second calibration curve gas concentrations. If the two calibration
curves differ by more than 5 percent from their mean value, then report the final
results by comparison to both calibration curves.
7.1.6 Determination of Bag Water Vapor Content. Measure the ambient temperature
and barometric pressure near the bag. From a water saturation vapor pressure
table, determine and record the water vapor content of the bag as a decimal
figure. (Assume the relative humidity to be 100 percent unless a lesser value
is known.) If the bag has been maintained at an elevated temperature as
described in Section 7.1.4, determine the stack gas water content by Method 4.
7.1.7 Quality Assurance. Immediately prior to the analysis of the stack gas
samples, perform audit analyses as described in Section 6.5. The audit analyses
must agree with the audit concentrations within 10 percent. If the results are
acceptable, proceed with the analyses of the source samples. If they do not
agree within 10 percent, then determine the reason for the discrepancy, and take
corrective action before proceeding.
7.1.8 Emission Calculations. From the average calibration curve described in